Skip to content

GETINGE 86-SERIES S-8668 (REF S-86682001)Recalls & reported problems

Getinge Disinfection AB · WASHER DISINFECTOR Catalog number was S-86681001. Catalog number updated to S-86682001 in November 2018. · Class 2

All models of GETINGE 86-SERIES (5)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Mar 2, 2018Terminated

    There is potential for a cart with disinfected goods to unload without the trolley being docked. Disinfected goods could end up falling to the floor causing risk to nearby users and damage to goods and equipment.

    Z-1379-2018 · Root cause: Software design · Recall details · FDA record ↗

Other Disinfector, Medical Devices devices

All Disinfector, Medical Devices devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page