GETINGE 9100-SERIES 9120ERecalls & reported problems

Getinge Disinfection AB · WASHER DISINFECTOR 9120E · Class 2

All models of GETINGE 9100-SERIES (6)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07340153710023), received between Dec 4, 2017 and Mar 2, 2020 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Apr 1, 2022Open, Classified

    Inadequate documentation verifying whether device installation has been executed in accordance with the manufacturer's requirements. This can lead to installation failure modes and hazardous situations.

    Z-1078-2022 · Root cause: Other · Recall details · FDA record ↗

Most reported problems

Device problems

  • Improper or Incorrect Procedure or Method1
  • Leak/Splash1
  • Misassembled1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 2, 2020Leak/Splash; Misassembled; Leak/Splash; Misassembled

    MANUFACTURER REFERENCE NUMBER (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 9774512 ↗
  • MalfunctionDec 4, 2017Use of Device Problem; Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    ON (B)(6) WE BECOME AWARE OF AN INCIDENT WHERE ONE OF OUR MEDICAL DEVICE WAS INVOLVED. PER INFORMATION PROVIDED BY THE CUSTOMER IT WAS ESTABLISHED THAT DURING PREVENTIVE MAINTENANCE PERFORMED ON THE 9100-SERIES WASHER DISINFECTOR THE TECHNICIAN WENT INSIDE THE WASHER CHAMBER TO CHECK AND CLEAN THE FILTERS OF THE CHAMBER AND DID NOT LOCK DOOR WITH KEY SWITCH. IN THE SAME TIME OPERATOR OF THE DEVICE STARTED THE CYCLE. TECHNICIAN WHO WAS INSIDE THE CHAMBER STOPPED MACHINE IMMEDIATELY AND AS A RESULT HE WAS SPLASHED WITH COLD WATER. NO INJURY HAS BEEN REPORTED, HOWEVER WE DECIDED TO REPORT THIS CASE IN ABUNDANCE OF CAUTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 7078692 ↗

Other Disinfector, Medical Devices devices

All Disinfector, Medical Devices devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page