RayStation 6.1.1Recalls & reported problems

Raysearch Laboratories AB (Publ) · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

4 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
7

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

7 FDA recall records name this device's identifier, 4 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Jan 16, 2025Open, Classified

    Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.

    Z-1192-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Feb 11, 2022Open, Classified

    Incorrect source to skin or surface distance (SSD) calculation may result in dose error when using RayStation system.

    Z-0850-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  3. Dec 30, 2021Open, Classified

    If a new primary image set is selected while the cine loop is running, the primary image set will be displayed as both primary and secondary image set in all side-by-side views. This will also be true for any new patient or case opened while the cine loop is running

    Z-0732-2022 · Root cause: Software design · Recall details · FDA record ↗

  4. Sep 22, 2021Open, Classified

    An issue where the combined density in a dose grid voxel partially covered by the External ROI and also partially covered by an ROI of type Bolus, Support or Fixation may be unexpected.The density in the voxel can be both under- and overestimated

    Z-0849-2022 · Root cause: Device Design · Recall details · FDA record ↗

  5. Sep 9, 2021Terminated

    If the dynamic jaw mode is used and an optimization is continued after changing the dose grid or modifying the target ROI, the positions of the jaws may change unexpectedly for some control points. User must be aware to avoid unnecessary jaw openings during treatment planning

    Z-0848-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  6. Feb 22, 2021Terminated

    For some LINAC types, merging clinical beams with beams of approximate dose may lead to the approximate dose erroneously labeled as clinical dose. Merge beams can be used manually, in scripting, or as part of the Automated breast planning feature. In some cases, when merging two beams where the first beam has clinical dose and the second beam has approximate dose, the dose of the resulting beam will be labeled as Clinical, although dose for some of the beam segments is still calculated with the SVD dose engine and it should be labeled Approximate: Mixed dose. The difference between approximate and final Clinical dose is in most cases small, but there can be body sites such as lung where the difference can be significant.

    Z-1597-2021 · Root cause: Software design · Recall details · FDA record ↗

  7. May 18, 2020Terminated

    Software error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms that could result in a local underestimation of expected dose.

    Z-2352-2020 · Root cause: Software design · Recall details · FDA record ↗

Other System, Planning, Radiation Therapy Treatment devices

All System, Planning, Radiation Therapy Treatment devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page