TomoHelical and TomoDirect plans in RayStation 6, 7, 8A… recallZ-0848-2022 · Raysearch Laboratories AB · initiated Sep 9, 2021

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Raysearch Laboratories AB or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

If the dynamic jaw mode is used and an optimization is continued after changing the dose grid or modifying the target ROI, the positions of the jaws may change unexpectedly for some control points. User must be aware to avoid unnecessary jaw openings during treatment planning

FDA root cause category: Under Investigation by firm

What the company is doing

RaySearch issued Field Safety Notice, Medical Device Correction #83773 notice via email on Sept. 9, 2021. Letter states reason for recall, health risk and action to take: ACTIONS TO BE TAKEN BY THE USER " Avoid changing the dose grid after starting an optimization of TomoHelical and TomoDirect plans with dynamic jaw mode. " Always reset the optimization after editing the dose grid for TomoHelical and TomoDirect plans with dynamic jaw mode. " Always reset the optimization after modifying the target ROI. " It is recommended to use the BEV to inspect the jaw positions superior and inferior to the target dynamic jaw mode. " Educate planning staff and all users about this workaround. " Inspect your product and identify all installed units with the above software version number(s). " Confirm you have read and understood this notice by replying to the notification email. SOLUTION This issue will be resolved in the next version of RayStation, scheduled for market release in October 2021 (subject to market clearance in some markets). If customers wish to continue using versions of RayStation affected by this notice, all users must maintain awareness of this notice. Alternatively, customers can choose to upgrade to the new version once it becomes available for clinical use. For regulatory information, please contact [email protected].

Recall details

Recall number
Z-0848-2022
Status
Terminated
Initiated by the company
Sep 9, 2021
Recalling firm
Raysearch Laboratories AB
Quantity
51 (US) Licenses
Product code
MUJ
FDA submissions
K210645
Distribution
US Nationwide distribution in the states of AL, CA, FL, GA, IA, IL, MO, MS, NC, OK, OR, SC, TN, WA, WI.

Product description

TomoHelical and TomoDirect plans in RayStation 6, 7, 8A, 8B, 9A, 9B, 10A, 10B and 11A. Build Numbers: 6.1.1.2, 6.2.0.7, 6.3.0.6, 7.0.0.19, 8.0.0.61, 8.0.1.10, 8.1.0.47, 8.1.1.8, 8.1.2.5, 9.0.0.113, 9.1.0.933, 9.2.0.483, 10.0.0.1154, 10.0.1.52, 10.1.0.613, 11.0.0.951 or 11.0.1.29.

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Name (build number) UDI-DI: RayStation 6/RayPlan 2 Service Pack 1 (6.1.0.26) 07350002010037 RayStation 6/RayPlan 2 Service Pack 1 v.6.1.1 (6.1.1.2) 07350002010082 RayStation 6/RayPlan 2 Service Pack 2 (6.2.0.7) 07350002010075 RayStation 6/RayPlan 2 Service Pack 3 (6.3.0.6) 07350002010242 RayStation/RayPlan 7 (7.0.0.19) 07350002010068 RayStation/RayPlan 8A (8.0.0.61) 07350002010112 RayStation/RayPlan 8A Service Pack 1 (8.0.1.10) 07350002010136 RayStation/RayPlan 8B (8.1.0.47) 07350002010129 RayStation/RayPlan 8B Service Pack 1 (8.1.1.8) 07350002010204 RayStation/RayPlan 8B Service Pack 2 (8.1.2.5) 07350002010235 RayStation 9A (9.0.0.113) 07350002010174 RayStation 9B (9.1.0.933) 07350002010266 RayStation Service Pack 1 (9.2.0.483) 07350002010297 RayStation 10A (10.0.0.1154) 07350002010303 RayStation 10A Service Pack 1 (10.0.1.52) 07350002010365 RayStation 10B (10.1.0.613) 07350002010310 RayStation 11A (11.0.0.951) 07350002010389 RayStation 11A Service Pack 1 (11.0.1.29) 07350002010433

Devices named in this recall

16 devices in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page