RayStation 4.5, 4.7, 4.9, 5, 6, 7, 8A, 8B, 9A, 9B, 10A… recallZ-1597-2021 · Raysearch Laboratories AB · initiated Feb 22, 2021
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Raysearch Laboratories AB or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
For some LINAC types, merging clinical beams with beams of approximate dose may lead to the approximate dose erroneously labeled as clinical dose. Merge beams can be used manually, in scripting, or as part of the Automated breast planning feature. In some cases, when merging two beams where the first beam has clinical dose and the second beam has approximate dose, the dose of the resulting beam will be labeled as Clinical, although dose for some of the beam segments is still calculated with the SVD dose engine and it should be labeled Approximate: Mixed dose. The difference between approximate and final Clinical dose is in most cases small, but there can be body sites such as lung where the difference can be significant.
FDA root cause category: Software design
What the company is doing
RaySearch issued Field Safety Notice, Medical Device Correction #74977 letter on 22 February 2021 via Email. Letter states reason for recall, health risk and action to take: ACTIONS TO BE TAKEN BY THE USER " Manual Merge beams in the 3D-CRT module Scenario: When merging a beam B1 with other beams (e.g., B2 and B3) in the 3D-CRT module. Action: If the target beam B1 has a clinical dose before the merge, make sure that all the beamsto be merged (B2 and B3) also have a clinical dose computed with the same dose engine. " Scripting when using the action MergeBeamSegments Scenario: When running a script containing the action MergeBeamSegments as below: beam_set.MergeBeamSegments(TargetBeamName="B1", MergeBeamNames=["B2", "B3"]) Action: If the target beam B1 has a clinical dose before running the script, make sure that the beams to be merged (B2 and B3) also have a clinical dose computed with the same dose engine. " Automated breast planning Scenario: When creating a treatment plan in the Automated breast planning module, the generated breast plan has two beams: A1 and A2. Actions: 1. Change module to Plan optimization. 2. Invalidate the dose for both the A1 and A2 beams by changing a beam property that can easily be restored (e.g., change the couch angle from 0 to 1). 3. Set the beam property back to the intended value (e.g., set the couch angle back to 0 for both beams). 4. Compute final dose. " Educate planning staff and all users about this workaround. " Inspect your product and identify all installed units with the above software version number(s). " Confirm that you have read and understood this notice by replying to the notification email. SOLUTION This issue will be resolved in the next version of RayStation/RayPlan, scheduled for market release in May 2021 (subject to market clearance in some markets). If customers wish to continue using versions of RayStation/RayPlan affected by this notice, all users must maintain awareness of this notice.
Recall details
- Recall number
- Z-1597-2021
- Status
- Terminated
- Initiated by the company
- Feb 22, 2021
- Recalling firm
- Raysearch Laboratories AB
- Quantity
- 116 units
- Product code
- MUJ
- FDA submissions
- K200569
- Distribution
- US Nationwide distribution.
Product description
RayStation 4.5, 4.7, 4.9, 5, 6, 7, 8A, 8B, 9A, 9B, 10A, 10B, RayPlan 1, 2, 7, 8A, 8B, 9A, 10A, 10B including some of the Service Packs. Standalone software treatment planning system
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Build number: 4.5.1.14 , 4.7.2.5 , 4.7.3.13 , 4.7.4.4 , 4.7.5.4 , 4.7.6.7 , 4.9.0.42 , 5.0.1.11 , 5.0.2.35 , 5.0.3.17 , 6.0.0.24 , 6.1.1.2 , 6.2.0.7 , 6.3.0.6 , 7.0.0.19 , 8.0.0.61 , 8.0.1.10 , 8.1.0.47 , 8.1.1.8 , 8.1.2.5 , 9.0.0.113 , 9.1.0.933 , 9.2.0.483 , 10.0.0.1154 , 10.0.1.52 or 10.1.0.613 UDI: RayStation 4.0 (4.0.0.14) to RayStation 5 Service Pack 2 (5.0.2.35) N/A RayStation 5 Service Pack 3 (5.0.3.17) UDI 07350002010020 RayStation 6/RayPlan 2 (6.0.0.24) UDI 07350002010013 RayStation 6/RayPlan 2 Service Pack 1 (6.1.1.2) UDI 07350002010082 RayStation 6/RayPlan 2 Service Pack 2 (6.2.0.7) UDI 07350002010075 RayStation 6/RayPlan 2 Service Pack 3 (6.3.0.6) UDI 07350002010242 RayStation/RayPlan 7 (7.0.0.19) UDI 07350002010068 RayStation/RayPlan 8A (8.0.0.61) UDI 07350002010112 RayStation/RayPlan 8A Service Pack 1 (8.0.1.10) UDI 07350002010136 RayStation/RayPlan 8B (8.1.0.47) UDI 07350002010129 RayStation/RayPlan 8B Service Pack 1 (8.1.1.8) UDI 07350002010204 RayStation/RayPlan 8B Service Pack 2 (8.1.2.5) UDI 07350002010235 RayStation/RayPlan 9A (9.0.0.113) UDI 07350002010174 RayStation/RayPlan 9B (9.1.0.933) UDI 07350002010266 RayStation/RayPlan 9B Service Pack 1 (9.2.0.483) UDI 07350002010297 RayStation/RayPlan 10A (10.0.0.1154) UDI 07350002010303 RayStation/RayPlan 10A Service Pack 1 (10.0.1.52) UDI 07350002010365 RayStation/RayPlan 10B (10.1.0.613) UDI 07350002010310
Devices named in this recall
15 devices in the FDA device registry whose identifier appears in the recall's code information.
- RayStation 9Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment12 recalls0 reports
- RayStation 8.0.1Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment11 recalls0 reports
- RayStation 9B SP1Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment11 recalls0 reports
- RayStation 10.1Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment10 recalls1 report
- RayStation 8.0Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment10 recalls0 reports
- RayStation 8.1.1Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment9 recalls0 reports
- RayStation 10.0Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment9 recalls0 reports
- RayStation 10.0.1Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment9 recalls0 reports
- RayStation 7Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment8 recalls0 reports
- RayStation 8.1Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment8 recalls0 reports
- RayStation 8.1.2Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment8 recalls0 reports
- RayStation 6.2Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment7 recalls0 reports
- RayStation 6.1.1Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment7 recalls0 reports
- RayStation 6.0Raysearch Laboratories AB (Publ) · System,Planning,Radiation Therapy Treatment6 recalls0 reports
- RayStation 5.0.3Raysearch Laboratories AB (Publ) · System,Planning,Radiation Therapy Treatment5 recalls0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page