Skip to content

TRUWAVE, VAMP PLUS PXVP2260Recalls & reported problems

Edwards Lifesciences LLC · PRESSURE MONITORING SET · Class 2

All models of TRUWAVE, VAMP PLUS (17)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07460691950108), received between Mar 27, 2025 and Feb 18, 2026 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Air/Gas in Device” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

11 other recalls cover devices cleared under the same FDA submission (K142749, K181684, K896819); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 4, 2023Open, Classified

    Their is the potential for flushing difficulties during setup of the pressure monitoring kit with pressure transducer.

    Z-0414-2024 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Air/Gas in Device1

Patient problems

  • Burn(s)1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 18, 2026Air/Gas in Device

    AS REPORTED, AFTER ONE HOUR CONNECTED, WHEN DRAWING BLOOD, THIS DISPOSABLE PRESSURE TRANSDUCER WITH VAMP, LEAKED SALINE AND APPROXIMATELY 3-5 INCHES OF AIR ENTERED IN TO THE SYSTEM FROM THE TUBING CONNECTOR TO THE TREE-WAY KEY NEXT TO THE VAMP. THE BLOOD REMAINED IN THE SYSTEM; THERE WAS NO BLOOD LEAKING. THE DEVICE WAS REPLACED WITH A NEW UNIT. THERE WAS NO ALLEGATION OF PATIENT INJURY. PATIENT DEMOGRAPHICS COULD NOT BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24375268 ↗
  • InjuryMar 27, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED, A PATIENT RECEIVED A MAGNETIC RESONANCE IMAGING (MRI) THERMAL BURN WHILE CONNECTED TO THIS PRESSURE MONITORING SYSTEM. THE TRANSDUCER WAS LOCATED AT THE PATIENT'S HUMERUS AND CONNECTED TO AN IRADIMED ARTERIAL LINE MONITORING POD. THE PATIENT REQUIRED WOUND CONSULTS AND MEDICATED CREAM FOR THE THERMAL INJURY ON THE RIGHT HUMERUS. NO SURGICAL INTERVENTION WAS REQUIRED. THE CUSTOMER CONFIRMED THAT THERE WAS NO PRODUCT NON-CONFORMANCE WITH THE PRESSURE MONITORING KIT AND THEY WILL PROVIDE EDUCATION TO THEIR HOSPITAL STAFF REGARDING MRI SAFETY.

    Patient problems: Burn(s)

    MAUDE report 21715515 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K142749, K181684, K896819. One submission often covers many models, so check the product and lot details in each record.

  • Z-0413-2024 · Oct 4, 2023 · Open, Classified
    TruWave Pressure monitoring set, REF: PX212: cc/12 in (30cm), PX260: (3 cc/60in (150cm), PX272: 3 cc/72in (180cm), PX284: 3 cc/84 in (210 cm), PX2X2: x2 (3cc), PX3X3284C: X3 (3cc),
  • Z-1960-2021 · May 26, 2021 · Terminated
    Edwards TruWave (30 cc)/VAMP Jr/52 in (132 cm) Kit, REF: PXVJ0711, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 07460691954489
  • Z-1965-2021 · May 26, 2021 · Terminated
    Edwards TruWave 3 cc/VAMP Jr Kit, REF: VMP426PX, Pressure Monitoring Set, Rx Only, CE UDI: (01)57460691950547
  • Z-1963-2021 · May 26, 2021 · Terminated
    Edwards VAMP Jr. 6 in (15 cm) REF: VMP306, Blood Management System, Rx Only, Single Use, CE, Sterile EO UDI: (01)57460691944782
  • Z-1961-2021 · May 26, 2021 · Terminated
    Edwards TruWave x 2 (3 cc)/VAMP Jr/54 in (136) Kit, REF: T443952B, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 57460691930174 ***N
  • Z-1966-2021 · May 26, 2021 · Terminated
    Edwards TruWave (3 cc)/VAMP Jr, REF: VMP406PX, Pressure Monitoring Set, Rx Only, CE UDI: (01)07460691958746
  • Z-1959-2021 · May 26, 2021 · Terminated
    Edwards TruWave (3 cc)/VAMP Jr 48 in (120 cm) Kit, REF: PXVJ356, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 57460691954507
  • Z-1962-2021 · May 26, 2021 · Terminated
    Edwards REF: VJ0990, Blood Management System, Rx Only, Single Use, CE, Sterile EO UDI: (01)57460691951520
  • Z-1964-2021 · May 26, 2021 · Terminated
    Edwards TruWave (30 cc)/VAMP Jr Kit, REF: VMP306PX, Pressure Monitoring Set, Rx Only, CE, UDI: (01)57460691950370
  • Z-2776-2015 · Jul 31, 2015 · Terminated
    Edwards Pressure Monitoring Kit with TruWave disposable pressure transducer. A sterile, single-use kit that monitors pressures when attached to pressure monitoring catheters. For
  • Z-1146-2013 · Jan 30, 2013 · Terminated
    Pressure Monitoring sets with VAMP Plus closed blood sampling system model # T001691A, PXVP0566, PXVP2260, PXVP0826, PXVP0827, PXVP2284AT3, PXVP23X3AT3, T398807D, T100605B, T001762

Other TRANSDUCER, PRESSURE, CATHETER TIP devices

All TRANSDUCER, PRESSURE, CATHETER TIP devices →

Sources and updates