TruWave, VAMP T001744MRecalls & reported problems
Edwards Lifesciences LLC · PRESSURE MONITORING SET · Class 2
All models of TRUWAVE, VAMP (19)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 07460691959507), received between Apr 23, 2025 and Apr 23, 2025 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Obstruction of Flow” (6 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Oct 4, 2023Open, Classified
Their is the potential for flushing difficulties during setup of the pressure monitoring kit with pressure transducer.
Z-0414-2024 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Obstruction of Flow6
Patient problems
- Embolism/Embolus6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 23, 2025Obstruction of Flow; Obstruction of Flow
AS REPORTED, DURING USE OF THIS PRESSURE MONITORING SET IN THE PREVIOUS 1-2 WEEKS, DAMPED WAVEFORM WAS OBSERVED, SO DEVICE WAS CHECKED AND CONFIRMED THERE WERE NO KINKS AND SALINE BAG WAS FULLY PRESSURIZED. AFTER REMOVING THE LINE, BLOOD CLOT APPEARED. THE DEVICE WAS NOT AVAILABLE FOR EVALUATION.
Patient problems: Embolism/Embolus
MAUDE report 21891018 ↗MalfunctionApr 23, 2025Obstruction of Flow; Obstruction of Flow
AS REPORTED, DURING USE OF THIS PRESSURE MONITORING SET IN THE PREVIOUS 1-2 WEEKS, DAMPED WAVEFORM WAS OBSERVED, SO DEVICE WAS CHECKED AND CONFIRMED THERE WERE NO KINKS AND SALINE BAG WAS FULLY PRESSURIZED. AFTER REMOVING THE LINE, BLOOD CLOT APPEARED. THE DEVICE WAS NOT AVAILABLE FOR EVALUATION.
Patient problems: Embolism/Embolus
MAUDE report 21890926 ↗InjuryApr 23, 2025Obstruction of Flow; Obstruction of Flow
AS REPORTED, DURING USE IN PATIENT OF THIS PRESSURE MONITORING SET, THERE WERE BLOOD CLOTS IN BOTH ARMS AND IT WAS NOT POSSIBLE TO PLACE ANOTHER ARTERIAL LINE. THE DEVICE WAS NOT AVAILABLE FOR EVALUATION.
Patient problems: Embolism/Embolus
MAUDE report 21890866 ↗MalfunctionApr 23, 2025Obstruction of Flow; Obstruction of Flow
AS REPORTED, DURING USE OF THIS PRESSURE MONITORING SET IN THE PREVIOUS 1-2 WEEKS, DAMPED WAVEFORM WAS OBSERVED, SO DEVICE WAS CHECKED AND CONFIRMED THERE WERE NO KINKS AND SALINE BAG WAS FULLY PRESSURIZED. AFTER REMOVING THE LINE, BLOOD CLOT APPEARED. THE DEVICE WAS NOT AVAILABLE FOR EVALUATION.
Patient problems: Embolism/Embolus
MAUDE report 21890471 ↗MalfunctionApr 23, 2025Obstruction of Flow; Obstruction of Flow
AS REPORTED, DURING USE OF THIS PRESSURE MONITORING SET IN THE PREVIOUS 1-2 WEEKS, DAMPED WAVEFORM WAS OBSERVED, SO DEVICE WAS CHECKED AND CONFIRMED THERE WERE NO KINKS AND SALINE BAG WAS FULLY PRESSURIZED. AFTER REMOVING THE LINE, BLOOD CLOT APPEARED. THE DEVICE WAS NOT AVAILABLE FOR EVALUATION.
Patient problems: Embolism/Embolus
MAUDE report 21890397 ↗MalfunctionApr 23, 2025Obstruction of Flow; Obstruction of Flow
AS REPORTED, DURING USE OF THIS PRESSURE MONITORING SET IN THE PREVIOUS 1-2 WEEKS, DAMPED WAVEFORM WAS OBSERVED, SO DEVICE WAS CHECKED AND CONFIRMED THERE WERE NO KINKS AND SALINE BAG WAS FULLY PRESSURIZED. AFTER REMOVING THE LINE, BLOOD CLOT APPEARED. THE DEVICE WAS NOT AVAILABLE FOR EVALUATION.
Patient problems: Embolism/Embolus
MAUDE report 21890271 ↗
Other Transducer, Pressure, Catheter Tip devices
- TRUWAVE PX260Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall29 reports
- TRUWAVE, VAMP PLUS PXVP2284AT3Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall9 reports
- Comet H7493932430Boston Scientific Corporation · TRANSDUCER, PRESSURE, CATHETER TIP56 reports
- TRUWAVE PXMK2041Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall4 reports
- TRUWAVE PX284Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall3 reports
- TRUWAVE, VAMP PLUS PXVP23X3Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall2 reports
- TRUWAVE, VAMP PLUS PXVP2260Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall2 reports
- TRUWAVE, VAMP PLUS PXVP2272Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page