The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07460691951464), received between Mar 28, 2024 and Feb 6, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

10 other recalls cover devices cleared under the same FDA submission (K142749); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 4, 2023Open, Classified

    Their is the potential for flushing difficulties during setup of the pressure monitoring kit with pressure transducer.

    Z-0413-2024 · Root cause: Device Design · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component2
  • Inaccurate Delivery2
  • Inaccurate Flow Rate2
  • Material Separation2

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 6, 2025Inaccurate Flow Rate; Inaccurate Delivery; Inaccurate Flow Rate; Inaccurate Delivery

    AS REPORTED; AFTER OBSERVING A DAMPED WAVEFORM DURING USE IN PATIENT OF THIS DISPOSABLE PRESSURE TRANSDUCER (REPORT REF 2015691-2025-00807), IT WAS DETECTED A PROBLEM IN THE SALINE FLOW RATE. IT LED TO THE SALINE BAG OF 500 ML GETTING EMPTY AND THE SOLUTION ENTERING THE PATIENT IN LESS THAN 24 HOURS. ANOTHER TRANSDUCER (REPORT REF 2015691-2025-00808) AND SALINE BAG WERE CHANGED BUT THE SAME ISSUE OCCURRED, LEADING TO OTHER 500 ML OF SALINE ENTERING THE PATIENT. A TOTAL OF 1000 ML OF SALINE ENTERED THE PATIENT IN 48 HOURS. THERE WAS A LEVEL INCREASE TO 150 MEQ/L NA+ IN THE ARTERIAL BLOOD GAS. THE DEVICE WAS REPLACED WITH A NEW UNIT AND THE ISSUE WAS SOLVED. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME. THERE WAS NO ALLEGATION OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21321847 ↗
  • MalfunctionFeb 6, 2025Inaccurate Flow Rate; Inaccurate Delivery; Inaccurate Flow Rate; Inaccurate Delivery

    AS REPORTED; AFTER OBSERVING A DAMPED WAVEFORM DURING USE IN PATIENT OF THIS DISPOSABLE PRESSURE TRANSDUCER (REPORT REF 2015691-2025-00807), IT WAS DETECTED A PROBLEM IN THE SALINE FLOW RATE. IT LED TO THE SALINE BAG OF 500 ML GETTING EMPTY AND THE SOLUTION ENTERING THE PATIENT IN LESS THAN 24 HOURS. ANOTHER TRANSDUCER (REPORT REF 2015691-2025-00808) AND SALINE BAG WERE CHANGED BUT THE SAME ISSUE OCCURRED, LEADING TO OTHER 500 ML OF SALINE ENTERING THE PATIENT. A TOTAL OF 1000 ML OF SALINE ENTERED THE PATIENT IN 48 HOURS. THERE WAS A LEVEL INCREASE TO 150 MEQ/L NA+ IN THE ARTERIAL BLOOD GAS. THE DEVICE WAS REPLACED WITH A NEW UNIT AND THE ISSUE WAS SOLVED. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME. THERE WAS NO ALLEGATION OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21321840 ↗
  • MalfunctionMar 28, 2024Material Separation; Detachment of Device or Device Component; Material Separation; Detachment of Device or Device Component

    AS REPORTED, DURING USE IN PATIENT IN A LIFE-THREATENING EMERGENCY, THIS DISPOSABLE PRESSURE TRANSDUCER (DPT) (2024-06126-01) UNSCREWED WHEN CONNECTED TO THE ARTERIAL ACCESS. IT WAS NECESSARY TO REDO THE ENTIRE ASSEMBLY OF THE PRESSURE LINE BY MANUALLY CONTINUING THE PURGE ON THE ARTERIAL LINE IN ORDER TO PREVENT IT FROM CLOGGING. TO SOLVE THE ISSUE, A NEW DEVICE WAS OPENED (2024-06126-02), HOWEVER, IT PRESENTED THE SAME ISSUE AS TWO OF THE SCREW THREADS WERE INSUFFICIENTLY SCREWED. ALL SCREW THREADS WERE CHECKED BEFORE NEW CONNECTION. PATIENT DEMOGRAPHICS ASKED. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICES WERE NOT AVAILABLE FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18990248 ↗
  • MalfunctionMar 28, 2024Material Separation; Detachment of Device or Device Component; Material Separation; Detachment of Device or Device Component

    AS REPORTED, DURING USE IN PATIENT IN A LIFE-THREATENING EMERGENCY, THIS DISPOSABLE PRESSURE TRANSDUCER (DPT) (MEDWATCH # 45420) UNSCREWED WHEN CONNECTED TO THE ARTERIAL ACCESS. IT WAS NECESSARY TO REDO THE ENTIRE ASSEMBLY OF THE PRESSURE LINE BY MANUALLY CONTINUING THE PURGE ON THE ARTERIAL LINE IN ORDER TO PREVENT IT FROM CLOGGING. TO SOLVE THE ISSUE, A NEW DEVICE WAS OPENED (MEDWATCH # 45423), HOWEVER, IT PRESENTED THE SAME ISSUE AS TWO OF THE SCREW THREADS WERE INSUFFICIENTLY SCREWED. ALL SCREW THREADS WERE CHECKED BEFORE NEW CONNECTION. PATIENT DEMOGRAPHICS ASKED. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICES WERE NOT AVAILABLE FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18990115 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K142749. One submission often covers many models, so check the product and lot details in each record.

  • Z-0414-2024 · Oct 4, 2023 · Open, Classified
    TruWave with Vamp System Pressure monitoring set, REF: PXAVMP(3cc)/VAMP 68 in (173cm) PXAVMP3: x3 (3cc)/VAMP PXVK0364:(3cc)/VAMP 96in (241cm); PXVK0853: x3 (3cc)/VAMP 76in (193cm)
  • Z-1962-2021 · May 26, 2021 · Terminated
    Edwards REF: VJ0990, Blood Management System, Rx Only, Single Use, CE, Sterile EO UDI: (01)57460691951520
  • Z-1964-2021 · May 26, 2021 · Terminated
    Edwards TruWave (30 cc)/VAMP Jr Kit, REF: VMP306PX, Pressure Monitoring Set, Rx Only, CE, UDI: (01)57460691950370
  • Z-1961-2021 · May 26, 2021 · Terminated
    Edwards TruWave x 2 (3 cc)/VAMP Jr/54 in (136) Kit, REF: T443952B, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 57460691930174 ***N
  • Z-1966-2021 · May 26, 2021 · Terminated
    Edwards TruWave (3 cc)/VAMP Jr, REF: VMP406PX, Pressure Monitoring Set, Rx Only, CE UDI: (01)07460691958746
  • Z-1960-2021 · May 26, 2021 · Terminated
    Edwards TruWave (30 cc)/VAMP Jr/52 in (132 cm) Kit, REF: PXVJ0711, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 07460691954489
  • Z-1963-2021 · May 26, 2021 · Terminated
    Edwards VAMP Jr. 6 in (15 cm) REF: VMP306, Blood Management System, Rx Only, Single Use, CE, Sterile EO UDI: (01)57460691944782
  • Z-1959-2021 · May 26, 2021 · Terminated
    Edwards TruWave (3 cc)/VAMP Jr 48 in (120 cm) Kit, REF: PXVJ356, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 57460691954507
  • Z-1965-2021 · May 26, 2021 · Terminated
    Edwards TruWave 3 cc/VAMP Jr Kit, REF: VMP426PX, Pressure Monitoring Set, Rx Only, CE UDI: (01)57460691950547
  • Z-2776-2015 · Jul 31, 2015 · Terminated
    Edwards Pressure Monitoring Kit with TruWave disposable pressure transducer. A sterile, single-use kit that monitors pressures when attached to pressure monitoring catheters. For

Other TRANSDUCER, PRESSURE, CATHETER TIP devices

All TRANSDUCER, PRESSURE, CATHETER TIP devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.