TRUWAVE PX284ANRecalls & reported problems
Edwards Lifesciences LLC · PRESSURE MONITORING SET · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07460691950245), received between May 1, 2023 and Jan 8, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Crack” (1 report).
11 other recalls cover devices cleared under the same FDA submission (K142749, K171996); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Crack1
- Fluid/Blood Leak1
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 8, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THE THREADS ON SYRINGES & THREADS ON TRANSDUCER ARE NOT LINING UP, CAUSING SYRINGE TIP TO BREAK OFF IN THE TRANSDUCER. WHEN COMING ON SHIFT, THE CUSTOMER NOTED THAT THE SYRINGE HOOKED UP TO STOPCOCK ON THE CVP TRANSDUCER WAS LEAKING EVEN THOUGH IT APPEARED CONNECTED. WHEN INSPECTING FURTHER, THE CUSTOMER NOTED THE TIP BROKE OFF INTO THE STOP COCK. THIS MEANT THE TRANSDUCER WASN'T ZEROED TO AIR DURING THE PREVIOUS SHIFT. WHEN NEW TRANSDUCER AND SYRINGE WERE HOOKED UP (SYRINGE FROM DIFFERENT LOT BUT TRANSDUCER FROM SAME LOT), IT IMMEDIATELY BROKE OFF AGAIN BEFORE SYRINGE WAS EVEN TWISTED ON ALL THE WAY. NEW TRANSDUCER GRABBED FROM NEW LOT AND SYRINGE SCREWED ON WITHOUT ANY ISSUES. CHARGE NOTIFIED OF EQUIPMENT ISSUE. NO ALLEGATION OF PATIENT INJURY, DEVICE AVAILABLE FOR RETURN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21104901 ↗MalfunctionMay 1, 2023Crack; Fluid/Blood Leak
EARLY 60¿S MALE WITH HISTORY OF CORONARY ARTERY DISEASE AND RECENT SHORTNESS OF BREATH. 7 DAYS POST OP FROM CORONARY ARTERY BYPASS GRAFT X 1 AND MONITORED IN CVICU. A NEW PACKAGE OF TRUWAVE 3CC PRESSURE TUBING OPENED TO CHANGE ARTERIAL LINE. WHEN THE TUBING WAS PRIMED, TUBING WAS WET AND THERE WAS A PUDDLE ON THE FLOOR. A CRACK WAS VISUALIZED IN THE TUBING AND THE TUBING WAS REMOVED, WITHOUT KNOWN HARM TO PATIENT. MANUFACTURER RESPONSE FOR TRANSDUCER, PRESSURE, CATHETER TIP, TRUWAVE (PER SITE REPORTER) WILL OBTAIN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16843322 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K142749, K171996. One submission often covers many models, so check the product and lot details in each record.
- Z-0413-2024 · Oct 4, 2023 · Open, ClassifiedTruWave Pressure monitoring set, REF: PX212: cc/12 in (30cm), PX260: (3 cc/60in (150cm), PX272: 3 cc/72in (180cm), PX284: 3 cc/84 in (210 cm), PX2X2: x2 (3cc), PX3X3284C: X3 (3cc),
- Z-0414-2024 · Oct 4, 2023 · Open, ClassifiedTruWave with Vamp System Pressure monitoring set, REF: PXAVMP(3cc)/VAMP 68 in (173cm) PXAVMP3: x3 (3cc)/VAMP PXVK0364:(3cc)/VAMP 96in (241cm); PXVK0853: x3 (3cc)/VAMP 76in (193cm)
- Z-1963-2021 · May 26, 2021 · TerminatedEdwards VAMP Jr. 6 in (15 cm) REF: VMP306, Blood Management System, Rx Only, Single Use, CE, Sterile EO UDI: (01)57460691944782
- Z-1959-2021 · May 26, 2021 · TerminatedEdwards TruWave (3 cc)/VAMP Jr 48 in (120 cm) Kit, REF: PXVJ356, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 57460691954507
- Z-1964-2021 · May 26, 2021 · TerminatedEdwards TruWave (30 cc)/VAMP Jr Kit, REF: VMP306PX, Pressure Monitoring Set, Rx Only, CE, UDI: (01)57460691950370
- Z-1965-2021 · May 26, 2021 · TerminatedEdwards TruWave 3 cc/VAMP Jr Kit, REF: VMP426PX, Pressure Monitoring Set, Rx Only, CE UDI: (01)57460691950547
- Z-1961-2021 · May 26, 2021 · TerminatedEdwards TruWave x 2 (3 cc)/VAMP Jr/54 in (136) Kit, REF: T443952B, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 57460691930174 ***N
- Z-1960-2021 · May 26, 2021 · TerminatedEdwards TruWave (30 cc)/VAMP Jr/52 in (132 cm) Kit, REF: PXVJ0711, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 07460691954489
- Z-1966-2021 · May 26, 2021 · TerminatedEdwards TruWave (3 cc)/VAMP Jr, REF: VMP406PX, Pressure Monitoring Set, Rx Only, CE UDI: (01)07460691958746
- Z-1962-2021 · May 26, 2021 · TerminatedEdwards REF: VJ0990, Blood Management System, Rx Only, Single Use, CE, Sterile EO UDI: (01)57460691951520
- Z-2776-2015 · Jul 31, 2015 · TerminatedEdwards Pressure Monitoring Kit with TruWave disposable pressure transducer. A sterile, single-use kit that monitors pressures when attached to pressure monitoring catheters. For
Other TRANSDUCER, PRESSURE, CATHETER TIP devices
- TRUWAVE PX260Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall29 reports
- TRUWAVE, VAMP PLUS PXVP2284AT3Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall9 reports
- Comet H7493932430Boston Scientific Corporation · TRANSDUCER, PRESSURE, CATHETER TIP56 reports
- TruWave, VAMP T001744MEdwards Lifesciences LLC · Transducer, Pressure, Catheter Tip1 recall6 reports
- TRUWAVE PXMK2041Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall4 reports
- TRUWAVE PX284Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall3 reports
- TRUWAVE, VAMP PLUS PXVP23X3Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall2 reports
- TRUWAVE, VAMP PLUS PXVP2260Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall2 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.