TRUWAVE, VAMP JR. VMP306PXRecalls & reported problems
Edwards Lifesciences LLC · PRESSURE MONITORING SET · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07460691950375), received between May 24, 2018 and Jan 16, 2026 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Air/Gas in Device” (1 report).
12 other recalls cover devices cleared under the same FDA submission (K142749, K896819); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Air/Gas in Device1
- Device Inoperable1
- Leak/Splash1
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 16, 2026Air/Gas in Device
IT WAS REPORTED THAT DURING USE OF THE PRESSURE MONITORING SET THE TEAM NOTICED THAT THE LINE STOPPED READING. UPON CHECKING THE LINE THE TEAM DISCOVERED BUBBLES IN THE LINE AND ATTEMPTED TO CLEAR THEM BUT KEPT PULLING AIR AND COULD NOT IDENTIFY WHERE THE AIR WAS COMING FROM. THEY FOUND THAT THE AIR WAS IN THE SYRINGE CONNECTED TO THE LINE TO TAKE A BLOOD SAMPLE. THEY DID LOOK AT THE FLUIDS HANDING AND THERE WERE WAS NOT ANY AIR IN THE BAG OR THE LINE. THIS WAS USED IN CONJUNCTION WITH THE ARGYLE UMBILICAL VESSEL CATHETER 3.5 FRENCH. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24097418 ↗MalfunctionJun 6, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A VMP306PX TRUWAVE 30CC VAMP JR FROM LOT 64615627, HAD SNAPPED IN THE MIDDLE OF THE DEVICE. IT HAD NOT BEEN USED ON A PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17075988 ↗InjuryJul 1, 2021Leak/Splash; Leak/Splash
IT WAS REPORTED THAT WHILE USING THIS VAMP JR., THERE WAS BLOOD BACKING UP IN THE TUBING AND THERE WAS LEAKAGE FROM THE CONNECTION POINT ON THE TRANSDUCER. THERE WAS NO PATIENT HARM, HOWEVER, THERE WAS 1ML OF BLOOD LOSS. THE PATIENT WAS AN INFANT AT (B)(6) GEST AGE, (B)(6) GRAMS, (B)(6) DAY OLD. THERE WAS NO PROCEDURE BEING CONDUCTED, HAD JUST HUNG IV FLUIDS. NURSE WAS AT BEDSIDE AT TIME AND REPORTED FLUID LOSS WAS MINIMAL. BABY DID RECEIVE A TRANSFUSION AFTER EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12100867 ↗MalfunctionMay 24, 2018Device Inoperable
VAMP JR TRANSDUCER WOULD NOT READ ARTERIAL BLOOD PRESSURE OR PULSE RATE. NEW TRANSDUCER WAS OBTAINED AND WORKED APPROPRIATELY. DEVICE IS IN MY POSSESSION HOWEVER PACKING WAS NOT RETAINED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7540245 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K142749, K896819. One submission often covers many models, so check the product and lot details in each record.
- Z-0413-2024 · Oct 4, 2023 · Open, ClassifiedTruWave Pressure monitoring set, REF: PX212: cc/12 in (30cm), PX260: (3 cc/60in (150cm), PX272: 3 cc/72in (180cm), PX284: 3 cc/84 in (210 cm), PX2X2: x2 (3cc), PX3X3284C: X3 (3cc),
- Z-0414-2024 · Oct 4, 2023 · Open, ClassifiedTruWave with Vamp System Pressure monitoring set, REF: PXAVMP(3cc)/VAMP 68 in (173cm) PXAVMP3: x3 (3cc)/VAMP PXVK0364:(3cc)/VAMP 96in (241cm); PXVK0853: x3 (3cc)/VAMP 76in (193cm)
- Z-1965-2021 · May 26, 2021 · TerminatedEdwards TruWave 3 cc/VAMP Jr Kit, REF: VMP426PX, Pressure Monitoring Set, Rx Only, CE UDI: (01)57460691950547
- Z-1962-2021 · May 26, 2021 · TerminatedEdwards REF: VJ0990, Blood Management System, Rx Only, Single Use, CE, Sterile EO UDI: (01)57460691951520
- Z-1959-2021 · May 26, 2021 · TerminatedEdwards TruWave (3 cc)/VAMP Jr 48 in (120 cm) Kit, REF: PXVJ356, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 57460691954507
- Z-1964-2021 · May 26, 2021 · TerminatedEdwards TruWave (30 cc)/VAMP Jr Kit, REF: VMP306PX, Pressure Monitoring Set, Rx Only, CE, UDI: (01)57460691950370
- Z-1966-2021 · May 26, 2021 · TerminatedEdwards TruWave (3 cc)/VAMP Jr, REF: VMP406PX, Pressure Monitoring Set, Rx Only, CE UDI: (01)07460691958746
- Z-1963-2021 · May 26, 2021 · TerminatedEdwards VAMP Jr. 6 in (15 cm) REF: VMP306, Blood Management System, Rx Only, Single Use, CE, Sterile EO UDI: (01)57460691944782
- Z-1961-2021 · May 26, 2021 · TerminatedEdwards TruWave x 2 (3 cc)/VAMP Jr/54 in (136) Kit, REF: T443952B, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 57460691930174 ***N
- Z-1960-2021 · May 26, 2021 · TerminatedEdwards TruWave (30 cc)/VAMP Jr/52 in (132 cm) Kit, REF: PXVJ0711, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 07460691954489
- Z-2776-2015 · Jul 31, 2015 · TerminatedEdwards Pressure Monitoring Kit with TruWave disposable pressure transducer. A sterile, single-use kit that monitors pressures when attached to pressure monitoring catheters. For
- Z-1146-2013 · Jan 30, 2013 · TerminatedPressure Monitoring sets with VAMP Plus closed blood sampling system model # T001691A, PXVP0566, PXVP2260, PXVP0826, PXVP0827, PXVP2284AT3, PXVP23X3AT3, T398807D, T100605B, T001762
Other TRANSDUCER, PRESSURE, CATHETER TIP devices
- TRUWAVE PX260Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall29 reports
- TRUWAVE, VAMP PLUS PXVP2284AT3Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall9 reports
- TruWave, VAMP T001744MEdwards Lifesciences LLC · Transducer, Pressure, Catheter Tip1 recall6 reports
- Comet H7493932430Boston Scientific Corporation · TRANSDUCER, PRESSURE, CATHETER TIP56 reports
- TRUWAVE PXMK2041Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall4 reports
- TRUWAVE PX284Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall3 reports
- TRUWAVE, VAMP PLUS PXVP2260Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall2 reports
- TRUWAVE, VAMP PLUS PXVP23X3Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page