TRUWAVE PX200Recalls & reported problems

Edwards Lifesciences LLC · PRESSURE MONITORING SET · Class 2

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07460691950528), received between Feb 18, 2026 and Mar 3, 2026 — classified as 2 injuries.

The most frequently reported specific device problem is “Inaccurate Delivery” (2 reports).

11 other recalls cover devices cleared under the same FDA submission (K142749, K171996); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Inaccurate Delivery2
  • Inaccurate Flow Rate1

Patient problems

  • Hypervolemia2
  • Unspecified Respiratory Problem2
  • Weight Changes2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 3, 2026Inaccurate Delivery

    AS REPORTED, DURING USE IN A NEONATAL PATIENT, THIS PRESSURE MONITORING KIT ADMINISTRATED TOO MUCH FLUID BECAUSE OF HIGHER FLOW RATE: FROM 0.2 TO 0.3 CC/HOUR. THE VOLUME OF SALINE ADMINISTRATED OVER 48 HOURS WAS ABOUT 120ML. ALL THE CONNECTIONS WERE CORRECTLY MADE. THEN, IT WAS DECIDED TO REMOVE THE UMBILICAL CATHETER. THE PATIENT HAD AN ELECTROLYTE IMBALANCE, GAINED TOO MUCH WEIGHT, AND SUFFERED A RESPIRATORY DEGRADATION. UNKNOWN MEDICATION WAS PROVIDED FOR IONIC REBALANCING. THE DEVICE WAS AVAILABLE FOR EVALUATION.

    Patient problems: Hypervolemia; Unspecified Respiratory Problem; Weight Changes

    MAUDE report 24494587 ↗
  • InjuryFeb 18, 2026Inaccurate Flow Rate; Inaccurate Delivery

    AS REPORTED, SEVERAL TIMES DURING 7 DAYS OF USE IN A PREMATURE BABY, THIS DISPOSABLE PRESSURE TRANSDUCER GOT DEPRESSURIZED CAUSING BACK FLOW AND A FLAT BLOOD PRESSURE CURVE. THE SYSTEM WAS FLUSHED AND PRESSURIZED EVERY TIME THE PROBLEM OCCURRED A PART FROM THE DAILY SYSTEM CHECKS. THE SYSTEM PRESSURE HAD TO BE ADJUSTED SEVERAL TIMES A DAY, WHICH IS UNUSUAL. ONE OF THE TIMES, THE SALINE BAG WAS NOTICED EMPTY. THIS MEANS THAT THE CHILD SHOULD HAVE RECEIVED 336 ML OVER 7 DAYS AND THAT THE ENTIRE 1-LITER BAG HAS BEEN CONSUMED. THE ENTIRE SYSTEM WAS REPLACED. THE CONSEQUENCES FOR THE PATIENT WERE ELECTROLYTE IMBALANCES, FAILURE TO INSERT A CENTRAL VENOUS CATHETER DUE TO EXCESSIVE LYMPHATIC FLUID, RESPIRATORY DIFFICULTIES WITH DELAYED EXTUBATING AND GAINED UP TO 310G BY DAY 4. THE PATIENT WAS TREATED WITH ANTIDIURETICS (LASILIX) AND HE IS OKAY.

    Patient problems: Hypervolemia; Unspecified Respiratory Problem; Weight Changes

    MAUDE report 24375460 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K142749, K171996. One submission often covers many models, so check the product and lot details in each record.

  • Z-0414-2024 · Oct 4, 2023 · Open, Classified
    TruWave with Vamp System Pressure monitoring set, REF: PXAVMP(3cc)/VAMP 68 in (173cm) PXAVMP3: x3 (3cc)/VAMP PXVK0364:(3cc)/VAMP 96in (241cm); PXVK0853: x3 (3cc)/VAMP 76in (193cm)
  • Z-0413-2024 · Oct 4, 2023 · Open, Classified
    TruWave Pressure monitoring set, REF: PX212: cc/12 in (30cm), PX260: (3 cc/60in (150cm), PX272: 3 cc/72in (180cm), PX284: 3 cc/84 in (210 cm), PX2X2: x2 (3cc), PX3X3284C: X3 (3cc),
  • Z-1966-2021 · May 26, 2021 · Terminated
    Edwards TruWave (3 cc)/VAMP Jr, REF: VMP406PX, Pressure Monitoring Set, Rx Only, CE UDI: (01)07460691958746
  • Z-1964-2021 · May 26, 2021 · Terminated
    Edwards TruWave (30 cc)/VAMP Jr Kit, REF: VMP306PX, Pressure Monitoring Set, Rx Only, CE, UDI: (01)57460691950370
  • Z-1961-2021 · May 26, 2021 · Terminated
    Edwards TruWave x 2 (3 cc)/VAMP Jr/54 in (136) Kit, REF: T443952B, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 57460691930174 ***N
  • Z-1959-2021 · May 26, 2021 · Terminated
    Edwards TruWave (3 cc)/VAMP Jr 48 in (120 cm) Kit, REF: PXVJ356, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 57460691954507
  • Z-1962-2021 · May 26, 2021 · Terminated
    Edwards REF: VJ0990, Blood Management System, Rx Only, Single Use, CE, Sterile EO UDI: (01)57460691951520
  • Z-1965-2021 · May 26, 2021 · Terminated
    Edwards TruWave 3 cc/VAMP Jr Kit, REF: VMP426PX, Pressure Monitoring Set, Rx Only, CE UDI: (01)57460691950547
  • Z-1963-2021 · May 26, 2021 · Terminated
    Edwards VAMP Jr. 6 in (15 cm) REF: VMP306, Blood Management System, Rx Only, Single Use, CE, Sterile EO UDI: (01)57460691944782
  • Z-1960-2021 · May 26, 2021 · Terminated
    Edwards TruWave (30 cc)/VAMP Jr/52 in (132 cm) Kit, REF: PXVJ0711, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 07460691954489
  • Z-2776-2015 · Jul 31, 2015 · Terminated
    Edwards Pressure Monitoring Kit with TruWave disposable pressure transducer. A sterile, single-use kit that monitors pressures when attached to pressure monitoring catheters. For

Other TRANSDUCER, PRESSURE, CATHETER TIP devices

All TRANSDUCER, PRESSURE, CATHETER TIP devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.