TRUWAVE PX200Recalls & reported problems
Edwards Lifesciences LLC · PRESSURE MONITORING SET · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07460691950528), received between Feb 18, 2026 and Mar 3, 2026 — classified as 2 injuries.
The most frequently reported specific device problem is “Inaccurate Delivery” (2 reports).
11 other recalls cover devices cleared under the same FDA submission (K142749, K171996); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Inaccurate Delivery2
- Inaccurate Flow Rate1
Patient problems
- Hypervolemia2
- Unspecified Respiratory Problem2
- Weight Changes2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMar 3, 2026Inaccurate Delivery
AS REPORTED, DURING USE IN A NEONATAL PATIENT, THIS PRESSURE MONITORING KIT ADMINISTRATED TOO MUCH FLUID BECAUSE OF HIGHER FLOW RATE: FROM 0.2 TO 0.3 CC/HOUR. THE VOLUME OF SALINE ADMINISTRATED OVER 48 HOURS WAS ABOUT 120ML. ALL THE CONNECTIONS WERE CORRECTLY MADE. THEN, IT WAS DECIDED TO REMOVE THE UMBILICAL CATHETER. THE PATIENT HAD AN ELECTROLYTE IMBALANCE, GAINED TOO MUCH WEIGHT, AND SUFFERED A RESPIRATORY DEGRADATION. UNKNOWN MEDICATION WAS PROVIDED FOR IONIC REBALANCING. THE DEVICE WAS AVAILABLE FOR EVALUATION.
Patient problems: Hypervolemia; Unspecified Respiratory Problem; Weight Changes
MAUDE report 24494587 ↗InjuryFeb 18, 2026Inaccurate Flow Rate; Inaccurate Delivery
AS REPORTED, SEVERAL TIMES DURING 7 DAYS OF USE IN A PREMATURE BABY, THIS DISPOSABLE PRESSURE TRANSDUCER GOT DEPRESSURIZED CAUSING BACK FLOW AND A FLAT BLOOD PRESSURE CURVE. THE SYSTEM WAS FLUSHED AND PRESSURIZED EVERY TIME THE PROBLEM OCCURRED A PART FROM THE DAILY SYSTEM CHECKS. THE SYSTEM PRESSURE HAD TO BE ADJUSTED SEVERAL TIMES A DAY, WHICH IS UNUSUAL. ONE OF THE TIMES, THE SALINE BAG WAS NOTICED EMPTY. THIS MEANS THAT THE CHILD SHOULD HAVE RECEIVED 336 ML OVER 7 DAYS AND THAT THE ENTIRE 1-LITER BAG HAS BEEN CONSUMED. THE ENTIRE SYSTEM WAS REPLACED. THE CONSEQUENCES FOR THE PATIENT WERE ELECTROLYTE IMBALANCES, FAILURE TO INSERT A CENTRAL VENOUS CATHETER DUE TO EXCESSIVE LYMPHATIC FLUID, RESPIRATORY DIFFICULTIES WITH DELAYED EXTUBATING AND GAINED UP TO 310G BY DAY 4. THE PATIENT WAS TREATED WITH ANTIDIURETICS (LASILIX) AND HE IS OKAY.
Patient problems: Hypervolemia; Unspecified Respiratory Problem; Weight Changes
MAUDE report 24375460 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K142749, K171996. One submission often covers many models, so check the product and lot details in each record.
- Z-0414-2024 · Oct 4, 2023 · Open, ClassifiedTruWave with Vamp System Pressure monitoring set, REF: PXAVMP(3cc)/VAMP 68 in (173cm) PXAVMP3: x3 (3cc)/VAMP PXVK0364:(3cc)/VAMP 96in (241cm); PXVK0853: x3 (3cc)/VAMP 76in (193cm)
- Z-0413-2024 · Oct 4, 2023 · Open, ClassifiedTruWave Pressure monitoring set, REF: PX212: cc/12 in (30cm), PX260: (3 cc/60in (150cm), PX272: 3 cc/72in (180cm), PX284: 3 cc/84 in (210 cm), PX2X2: x2 (3cc), PX3X3284C: X3 (3cc),
- Z-1966-2021 · May 26, 2021 · TerminatedEdwards TruWave (3 cc)/VAMP Jr, REF: VMP406PX, Pressure Monitoring Set, Rx Only, CE UDI: (01)07460691958746
- Z-1964-2021 · May 26, 2021 · TerminatedEdwards TruWave (30 cc)/VAMP Jr Kit, REF: VMP306PX, Pressure Monitoring Set, Rx Only, CE, UDI: (01)57460691950370
- Z-1961-2021 · May 26, 2021 · TerminatedEdwards TruWave x 2 (3 cc)/VAMP Jr/54 in (136) Kit, REF: T443952B, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 57460691930174 ***N
- Z-1959-2021 · May 26, 2021 · TerminatedEdwards TruWave (3 cc)/VAMP Jr 48 in (120 cm) Kit, REF: PXVJ356, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 57460691954507
- Z-1962-2021 · May 26, 2021 · TerminatedEdwards REF: VJ0990, Blood Management System, Rx Only, Single Use, CE, Sterile EO UDI: (01)57460691951520
- Z-1965-2021 · May 26, 2021 · TerminatedEdwards TruWave 3 cc/VAMP Jr Kit, REF: VMP426PX, Pressure Monitoring Set, Rx Only, CE UDI: (01)57460691950547
- Z-1963-2021 · May 26, 2021 · TerminatedEdwards VAMP Jr. 6 in (15 cm) REF: VMP306, Blood Management System, Rx Only, Single Use, CE, Sterile EO UDI: (01)57460691944782
- Z-1960-2021 · May 26, 2021 · TerminatedEdwards TruWave (30 cc)/VAMP Jr/52 in (132 cm) Kit, REF: PXVJ0711, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 07460691954489
- Z-2776-2015 · Jul 31, 2015 · TerminatedEdwards Pressure Monitoring Kit with TruWave disposable pressure transducer. A sterile, single-use kit that monitors pressures when attached to pressure monitoring catheters. For
Other TRANSDUCER, PRESSURE, CATHETER TIP devices
- TRUWAVE PX260Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall29 reports
- TRUWAVE, VAMP PLUS PXVP2284AT3Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall9 reports
- Comet H7493932430Boston Scientific Corporation · TRANSDUCER, PRESSURE, CATHETER TIP56 reports
- TruWave, VAMP T001744MEdwards Lifesciences LLC · Transducer, Pressure, Catheter Tip1 recall6 reports
- TRUWAVE PXMK2041Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall4 reports
- TRUWAVE PX284Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall3 reports
- TRUWAVE, VAMP PLUS PXVP23X3Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall2 reports
- TRUWAVE, VAMP PLUS PXVP2260Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall2 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.