TRUWAVE, VAMP PXVK0670Recalls & reported problems
Edwards Lifesciences LLC · PRESSURE MONITORING SET · Class 2
All models of TRUWAVE, VAMP (19)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07460691954748), received between May 7, 2024 and Jul 10, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
12 other recalls cover devices cleared under the same FDA submission (K142749, K896819); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component1
- High Readings1
- Incorrect Measurement1
Patient problems
- Hemorrhage/Blood Loss/Bleeding1
- Low Blood Pressure/ Hypotension1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 10, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, DURING USE IN PATIENT, IN THIS PRESSURE MONITORING SET WITH VAMP ADULT THE ARTERIAL LINE DETACHED FROM THE RESERVOIR CAUSING BLEEDING FOR THE PATIENT. APPARENTLY, THERE WAS NO TUGGING ON THE LINE. NO FURTHER INFORMATION AVAILABLE. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS AVAILABLE FOR EVALUATION. PATIENT DEMOGRAPHICS UNABLE TO BE OBTAINED.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 22468644 ↗MalfunctionMay 7, 2024Incorrect Measurement; High Readings; Incorrect Measurement; High Readings
AS REPORTED, DURING USE IN PATIENT WITH THIS DISPOSABLE PRESSURE TRANSDUCER, THE ARTERIAL LINE WAS SHOWING A MEAN ARTERIAL PRESSURE (MAP) OF 100 MMHG. NO ALARMS OR ERROR MESSAGES WERE DISPLAYED. MANUAL BLOOD PRESSURE WAS DONE WITH A VERY CLEAR 70/40 MMHG WITH A MAP EQUALING 50. ARTERIAL LINE RE ZEROED, CHECKED PRESSURE BAG, CHECKED CANNULA FOR KINKS, ENSURED TRANSDUCER WAS IN CORRECT PLACE AND GOT A NEW CABLE. STILL SHOWING A BLOOD PRESSURE WITH A MAP OF 100. NEW TRANSDUCER LINE PRIMED AND CONNECTED, SHOWING AN ACCURATE BLOOD PRESSURE WITH A MAP OF APPROXIMATELY 50 MMHG. BLOOD PRESSURE WAS TREATED ACCORDINGLY. AS PER CUSTOMER OPINION, THE INACCURATE VALUE ISSUE RESULTED IN AN EXTENDED PERIOD OF PATIENT HYPOTENSION WHICH COULD CAUSE A PATIENT INJURY, THEREFORE THIS WAS REPORTED AS AN INJURY. PATIENT DEMOGRAPHICS UNABLE TO BE OBTAINED. THE DEVICE WAS AVAILABLE FOR EVALUATION.
Patient problems: Low Blood Pressure/ Hypotension
MAUDE report 19258525 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K142749, K896819. One submission often covers many models, so check the product and lot details in each record.
- Z-0413-2024 · Oct 4, 2023 · Open, ClassifiedTruWave Pressure monitoring set, REF: PX212: cc/12 in (30cm), PX260: (3 cc/60in (150cm), PX272: 3 cc/72in (180cm), PX284: 3 cc/84 in (210 cm), PX2X2: x2 (3cc), PX3X3284C: X3 (3cc),
- Z-0414-2024 · Oct 4, 2023 · Open, ClassifiedTruWave with Vamp System Pressure monitoring set, REF: PXAVMP(3cc)/VAMP 68 in (173cm) PXAVMP3: x3 (3cc)/VAMP PXVK0364:(3cc)/VAMP 96in (241cm); PXVK0853: x3 (3cc)/VAMP 76in (193cm)
- Z-1966-2021 · May 26, 2021 · TerminatedEdwards TruWave (3 cc)/VAMP Jr, REF: VMP406PX, Pressure Monitoring Set, Rx Only, CE UDI: (01)07460691958746
- Z-1964-2021 · May 26, 2021 · TerminatedEdwards TruWave (30 cc)/VAMP Jr Kit, REF: VMP306PX, Pressure Monitoring Set, Rx Only, CE, UDI: (01)57460691950370
- Z-1960-2021 · May 26, 2021 · TerminatedEdwards TruWave (30 cc)/VAMP Jr/52 in (132 cm) Kit, REF: PXVJ0711, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 07460691954489
- Z-1963-2021 · May 26, 2021 · TerminatedEdwards VAMP Jr. 6 in (15 cm) REF: VMP306, Blood Management System, Rx Only, Single Use, CE, Sterile EO UDI: (01)57460691944782
- Z-1959-2021 · May 26, 2021 · TerminatedEdwards TruWave (3 cc)/VAMP Jr 48 in (120 cm) Kit, REF: PXVJ356, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 57460691954507
- Z-1961-2021 · May 26, 2021 · TerminatedEdwards TruWave x 2 (3 cc)/VAMP Jr/54 in (136) Kit, REF: T443952B, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 57460691930174 ***N
- Z-1965-2021 · May 26, 2021 · TerminatedEdwards TruWave 3 cc/VAMP Jr Kit, REF: VMP426PX, Pressure Monitoring Set, Rx Only, CE UDI: (01)57460691950547
- Z-1962-2021 · May 26, 2021 · TerminatedEdwards REF: VJ0990, Blood Management System, Rx Only, Single Use, CE, Sterile EO UDI: (01)57460691951520
- Z-2776-2015 · Jul 31, 2015 · TerminatedEdwards Pressure Monitoring Kit with TruWave disposable pressure transducer. A sterile, single-use kit that monitors pressures when attached to pressure monitoring catheters. For
- Z-1146-2013 · Jan 30, 2013 · TerminatedPressure Monitoring sets with VAMP Plus closed blood sampling system model # T001691A, PXVP0566, PXVP2260, PXVP0826, PXVP0827, PXVP2284AT3, PXVP23X3AT3, T398807D, T100605B, T001762
Other TRANSDUCER, PRESSURE, CATHETER TIP devices
- TRUWAVE PX260Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall29 reports
- TRUWAVE, VAMP PLUS PXVP2284AT3Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall9 reports
- TruWave, VAMP T001744MEdwards Lifesciences LLC · Transducer, Pressure, Catheter Tip1 recall6 reports
- Comet H7493932430Boston Scientific Corporation · TRANSDUCER, PRESSURE, CATHETER TIP56 reports
- TRUWAVE PXMK2041Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall4 reports
- TRUWAVE PX284Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall3 reports
- TRUWAVE, VAMP PLUS PXVP2260Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall2 reports
- TRUWAVE, VAMP PLUS PXVP23X3Edwards Lifesciences LLC · TRANSDUCER, PRESSURE, CATHETER TIP1 recall2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page