Skip to content

TRUWAVE, VAMP PXVK0670Recalls & reported problems

Edwards Lifesciences LLC · PRESSURE MONITORING SET · Class 2

All models of TRUWAVE, VAMP (19)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07460691954748), received between May 7, 2024 and Jul 10, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

12 other recalls cover devices cleared under the same FDA submission (K142749, K896819); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • High Readings1
  • Incorrect Measurement1

Patient problems

  • Hemorrhage/Blood Loss/Bleeding1
  • Low Blood Pressure/ Hypotension1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 10, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    AS REPORTED, DURING USE IN PATIENT, IN THIS PRESSURE MONITORING SET WITH VAMP ADULT THE ARTERIAL LINE DETACHED FROM THE RESERVOIR CAUSING BLEEDING FOR THE PATIENT. APPARENTLY, THERE WAS NO TUGGING ON THE LINE. NO FURTHER INFORMATION AVAILABLE. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS AVAILABLE FOR EVALUATION. PATIENT DEMOGRAPHICS UNABLE TO BE OBTAINED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 22468644 ↗
  • MalfunctionMay 7, 2024Incorrect Measurement; High Readings; Incorrect Measurement; High Readings

    AS REPORTED, DURING USE IN PATIENT WITH THIS DISPOSABLE PRESSURE TRANSDUCER, THE ARTERIAL LINE WAS SHOWING A MEAN ARTERIAL PRESSURE (MAP) OF 100 MMHG. NO ALARMS OR ERROR MESSAGES WERE DISPLAYED. MANUAL BLOOD PRESSURE WAS DONE WITH A VERY CLEAR 70/40 MMHG WITH A MAP EQUALING 50. ARTERIAL LINE RE ZEROED, CHECKED PRESSURE BAG, CHECKED CANNULA FOR KINKS, ENSURED TRANSDUCER WAS IN CORRECT PLACE AND GOT A NEW CABLE. STILL SHOWING A BLOOD PRESSURE WITH A MAP OF 100. NEW TRANSDUCER LINE PRIMED AND CONNECTED, SHOWING AN ACCURATE BLOOD PRESSURE WITH A MAP OF APPROXIMATELY 50 MMHG. BLOOD PRESSURE WAS TREATED ACCORDINGLY. AS PER CUSTOMER OPINION, THE INACCURATE VALUE ISSUE RESULTED IN AN EXTENDED PERIOD OF PATIENT HYPOTENSION WHICH COULD CAUSE A PATIENT INJURY, THEREFORE THIS WAS REPORTED AS AN INJURY. PATIENT DEMOGRAPHICS UNABLE TO BE OBTAINED. THE DEVICE WAS AVAILABLE FOR EVALUATION.

    Patient problems: Low Blood Pressure/ Hypotension

    MAUDE report 19258525 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K142749, K896819. One submission often covers many models, so check the product and lot details in each record.

  • Z-0413-2024 · Oct 4, 2023 · Open, Classified
    TruWave Pressure monitoring set, REF: PX212: cc/12 in (30cm), PX260: (3 cc/60in (150cm), PX272: 3 cc/72in (180cm), PX284: 3 cc/84 in (210 cm), PX2X2: x2 (3cc), PX3X3284C: X3 (3cc),
  • Z-0414-2024 · Oct 4, 2023 · Open, Classified
    TruWave with Vamp System Pressure monitoring set, REF: PXAVMP(3cc)/VAMP 68 in (173cm) PXAVMP3: x3 (3cc)/VAMP PXVK0364:(3cc)/VAMP 96in (241cm); PXVK0853: x3 (3cc)/VAMP 76in (193cm)
  • Z-1966-2021 · May 26, 2021 · Terminated
    Edwards TruWave (3 cc)/VAMP Jr, REF: VMP406PX, Pressure Monitoring Set, Rx Only, CE UDI: (01)07460691958746
  • Z-1964-2021 · May 26, 2021 · Terminated
    Edwards TruWave (30 cc)/VAMP Jr Kit, REF: VMP306PX, Pressure Monitoring Set, Rx Only, CE, UDI: (01)57460691950370
  • Z-1960-2021 · May 26, 2021 · Terminated
    Edwards TruWave (30 cc)/VAMP Jr/52 in (132 cm) Kit, REF: PXVJ0711, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 07460691954489
  • Z-1963-2021 · May 26, 2021 · Terminated
    Edwards VAMP Jr. 6 in (15 cm) REF: VMP306, Blood Management System, Rx Only, Single Use, CE, Sterile EO UDI: (01)57460691944782
  • Z-1959-2021 · May 26, 2021 · Terminated
    Edwards TruWave (3 cc)/VAMP Jr 48 in (120 cm) Kit, REF: PXVJ356, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 57460691954507
  • Z-1961-2021 · May 26, 2021 · Terminated
    Edwards TruWave x 2 (3 cc)/VAMP Jr/54 in (136) Kit, REF: T443952B, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 57460691930174 ***N
  • Z-1965-2021 · May 26, 2021 · Terminated
    Edwards TruWave 3 cc/VAMP Jr Kit, REF: VMP426PX, Pressure Monitoring Set, Rx Only, CE UDI: (01)57460691950547
  • Z-1962-2021 · May 26, 2021 · Terminated
    Edwards REF: VJ0990, Blood Management System, Rx Only, Single Use, CE, Sterile EO UDI: (01)57460691951520
  • Z-2776-2015 · Jul 31, 2015 · Terminated
    Edwards Pressure Monitoring Kit with TruWave disposable pressure transducer. A sterile, single-use kit that monitors pressures when attached to pressure monitoring catheters. For
  • Z-1146-2013 · Jan 30, 2013 · Terminated
    Pressure Monitoring sets with VAMP Plus closed blood sampling system model # T001691A, PXVP0566, PXVP2260, PXVP0826, PXVP0827, PXVP2284AT3, PXVP23X3AT3, T398807D, T100605B, T001762

Other TRANSDUCER, PRESSURE, CATHETER TIP devices

All TRANSDUCER, PRESSURE, CATHETER TIP devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page