Skip to content

TRUWAVE, VAMP PLUS PXVP0566Recalls & reported problems

Edwards Lifesciences LLC · PRESSURE MONITORING KIT - STERILE · Class 2

All models of TRUWAVE, VAMP PLUS (17)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07460691955172), received between Aug 12, 2024 and Feb 24, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

12 other recalls cover devices cleared under the same FDA submission (K142749, K181684, K896819); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Detachment of Device or Device Component1

Patient problems

  • Full thickness (Third Degree) Burn1
  • No Clinical Signs, Symptoms or Conditions1
  • Partial thickness (Second Degree) Burn1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ADDITIONAL INFORMATION WAS RECEIVED 7 MARCH 2025: PER CUSTOMER, THE STRENGTH OF THE MRI WAS 1.5 TESLA, THE PATIENT WAS UNDERGOING A LENGTHY MRI FOR CERVICAL, THORACIC, AND LUMBAR SPINES. THE SCAN TIME WAS 1 HOUR 46 MINUTES ON/OFF FOR CARE AS NEEDED. THE PATIENT WAS LAYING FACE UP IN SPINAL PRECAUTIONS. THE PATIENT WAS A TRAUMATIC QUADRIPLEGIA, AND HE WAS INTUBATED AND SEDATED. THE PATIENT REQUIRED LEFT SHOULDER EXCISIONAL DEBRIDEMENT AND APPLICATION OF ARTIFICIAL SKIN GRAFT. PRODUCT QUALITY CLINICIAN SPOKE DIRECTLY WITH CUSTOMER TO INFORM THEM THAT THIS DEVICE IS NOT TO BE USED IN THE BORE OF THE MRI. THE CUSTOMER STATED THAT SHE WILL PROVIDE EDUCATION TO HOSPITAL STAFF.

    Patient problems: Full thickness (Third Degree) Burn; Partial thickness (Second Degree) Burn

    MAUDE report 21447086 ↗
  • MalfunctionAug 12, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE PXVP05066 TUBING WAS NOT GLUED PROPERLY AND COMING APART. THERE WAS NO ALLEGATION OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19968737 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K142749, K181684, K896819. One submission often covers many models, so check the product and lot details in each record.

  • Z-0413-2024 · Oct 4, 2023 · Open, Classified
    TruWave Pressure monitoring set, REF: PX212: cc/12 in (30cm), PX260: (3 cc/60in (150cm), PX272: 3 cc/72in (180cm), PX284: 3 cc/84 in (210 cm), PX2X2: x2 (3cc), PX3X3284C: X3 (3cc),
  • Z-0414-2024 · Oct 4, 2023 · Open, Classified
    TruWave with Vamp System Pressure monitoring set, REF: PXAVMP(3cc)/VAMP 68 in (173cm) PXAVMP3: x3 (3cc)/VAMP PXVK0364:(3cc)/VAMP 96in (241cm); PXVK0853: x3 (3cc)/VAMP 76in (193cm)
  • Z-1963-2021 · May 26, 2021 · Terminated
    Edwards VAMP Jr. 6 in (15 cm) REF: VMP306, Blood Management System, Rx Only, Single Use, CE, Sterile EO UDI: (01)57460691944782
  • Z-1960-2021 · May 26, 2021 · Terminated
    Edwards TruWave (30 cc)/VAMP Jr/52 in (132 cm) Kit, REF: PXVJ0711, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 07460691954489
  • Z-1965-2021 · May 26, 2021 · Terminated
    Edwards TruWave 3 cc/VAMP Jr Kit, REF: VMP426PX, Pressure Monitoring Set, Rx Only, CE UDI: (01)57460691950547
  • Z-1962-2021 · May 26, 2021 · Terminated
    Edwards REF: VJ0990, Blood Management System, Rx Only, Single Use, CE, Sterile EO UDI: (01)57460691951520
  • Z-1961-2021 · May 26, 2021 · Terminated
    Edwards TruWave x 2 (3 cc)/VAMP Jr/54 in (136) Kit, REF: T443952B, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 57460691930174 ***N
  • Z-1966-2021 · May 26, 2021 · Terminated
    Edwards TruWave (3 cc)/VAMP Jr, REF: VMP406PX, Pressure Monitoring Set, Rx Only, CE UDI: (01)07460691958746
  • Z-1964-2021 · May 26, 2021 · Terminated
    Edwards TruWave (30 cc)/VAMP Jr Kit, REF: VMP306PX, Pressure Monitoring Set, Rx Only, CE, UDI: (01)57460691950370
  • Z-1959-2021 · May 26, 2021 · Terminated
    Edwards TruWave (3 cc)/VAMP Jr 48 in (120 cm) Kit, REF: PXVJ356, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 57460691954507
  • Z-2776-2015 · Jul 31, 2015 · Terminated
    Edwards Pressure Monitoring Kit with TruWave disposable pressure transducer. A sterile, single-use kit that monitors pressures when attached to pressure monitoring catheters. For
  • Z-1146-2013 · Jan 30, 2013 · Terminated
    Pressure Monitoring sets with VAMP Plus closed blood sampling system model # T001691A, PXVP0566, PXVP2260, PXVP0826, PXVP0827, PXVP2284AT3, PXVP23X3AT3, T398807D, T100605B, T001762

Other TRANSDUCER, PRESSURE, CATHETER TIP devices

All TRANSDUCER, PRESSURE, CATHETER TIP devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page