Skip to content

NON-DEHP, SINGLE DPT LINE WITH VAMP JR RESERVOIR AND LASS PXVJ050MRecalls & reported problems

Edwards Lifesciences LLC · PRESSURE MONITORING SET · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07460691960572), received between Jun 17, 2024 and Feb 6, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

11 other recalls cover devices cleared under the same FDA submission (K142749); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available1
  • Detachment of Device or Device Component1
  • Leak/Splash1

Patient problems

  • Fluid Discharge1
  • Hemorrhage/Blood Loss/Bleeding1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 6, 2026Leak/Splash

    AS REPORTED, WHEN RETURNING BLOOD FROM PREVIOUSLY EXTRACTED BLOOD SAMPLE TO AN ICU PATIENT, THIS DISPOSABLE PRESSURE TRANSDUCER (DPT) WITH VAMP JUNIOR EXPERIENCED RESISTANCE, AS CONSEQUENCE, BLOOD SPRAYED FROM SAMPLE PORT TO STAFF MEMBER FACE AND TRUNK. THE DISTAL STOPCOCK WAS OPEN DURING PROCESS. A NEW TRANSDUCER WAS USED FROM THE SAME LOT, HOWEVER, IT ALSO EXPERIENCED RESISTANCE PUSHING THE PLUNGER RESULTING IN SALINE SPRAYING FROM SAMPLE PORT. THE NURSE WORE GLOVES BUT NO GOGGLES NOR FACEMASK. SEROLOGY TESTS WERE PERFORMED TO THE PATIENT AND NURSE AND CAME BACK NEGATIVE. THERE WAS NO ALLEGATION OF PATIENT INJURY.

    Patient problems: Fluid Discharge

    MAUDE report 24278251 ↗
  • MalfunctionJun 17, 2024Detachment of Device or Device Component; Appropriate Device Problem Term/Code Not Available; Detachment of Device or Device Component; Appropriate Device Problem Term/Code Not Available

    AS REPORTED, DURING USE IN PATIENT WITH THIS DISPOSABLE PRESSURE TRANSDUCER WITH VAMP JR, THE PRESSURE TUBING BECAME DISCONNECTED RIGHT BESIDE THE SAMPLING SYRINGE. WHILST TAKING A SAMPLE OF BLOOD FROM THE PATIENT THE NURSE NOTICED A SMALL PATCH OF BLOOD ON THE SHEETS WHICH SHE FOUND TO BE AS A RESULT OF THE BONDED TUBE CONNECTION BREAKING OFF. THE LINE WAS IMMEDIATELY CLAMPED, AND THE ARTERIAL LINE PRESSURE TRANSDUCER KIT WAS CHANGED. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS NOT AVAILABLE FOR EVALUATION.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 19547053 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K142749. One submission often covers many models, so check the product and lot details in each record.

  • Z-0413-2024 · Oct 4, 2023 · Open, Classified
    TruWave Pressure monitoring set, REF: PX212: cc/12 in (30cm), PX260: (3 cc/60in (150cm), PX272: 3 cc/72in (180cm), PX284: 3 cc/84 in (210 cm), PX2X2: x2 (3cc), PX3X3284C: X3 (3cc),
  • Z-0414-2024 · Oct 4, 2023 · Open, Classified
    TruWave with Vamp System Pressure monitoring set, REF: PXAVMP(3cc)/VAMP 68 in (173cm) PXAVMP3: x3 (3cc)/VAMP PXVK0364:(3cc)/VAMP 96in (241cm); PXVK0853: x3 (3cc)/VAMP 76in (193cm)
  • Z-1966-2021 · May 26, 2021 · Terminated
    Edwards TruWave (3 cc)/VAMP Jr, REF: VMP406PX, Pressure Monitoring Set, Rx Only, CE UDI: (01)07460691958746
  • Z-1960-2021 · May 26, 2021 · Terminated
    Edwards TruWave (30 cc)/VAMP Jr/52 in (132 cm) Kit, REF: PXVJ0711, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 07460691954489
  • Z-1963-2021 · May 26, 2021 · Terminated
    Edwards VAMP Jr. 6 in (15 cm) REF: VMP306, Blood Management System, Rx Only, Single Use, CE, Sterile EO UDI: (01)57460691944782
  • Z-1959-2021 · May 26, 2021 · Terminated
    Edwards TruWave (3 cc)/VAMP Jr 48 in (120 cm) Kit, REF: PXVJ356, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 57460691954507
  • Z-1965-2021 · May 26, 2021 · Terminated
    Edwards TruWave 3 cc/VAMP Jr Kit, REF: VMP426PX, Pressure Monitoring Set, Rx Only, CE UDI: (01)57460691950547
  • Z-1961-2021 · May 26, 2021 · Terminated
    Edwards TruWave x 2 (3 cc)/VAMP Jr/54 in (136) Kit, REF: T443952B, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 57460691930174 ***N
  • Z-1962-2021 · May 26, 2021 · Terminated
    Edwards REF: VJ0990, Blood Management System, Rx Only, Single Use, CE, Sterile EO UDI: (01)57460691951520
  • Z-1964-2021 · May 26, 2021 · Terminated
    Edwards TruWave (30 cc)/VAMP Jr Kit, REF: VMP306PX, Pressure Monitoring Set, Rx Only, CE, UDI: (01)57460691950370
  • Z-2776-2015 · Jul 31, 2015 · Terminated
    Edwards Pressure Monitoring Kit with TruWave disposable pressure transducer. A sterile, single-use kit that monitors pressures when attached to pressure monitoring catheters. For

Other Transducer, Pressure, Catheter Tip devices

All Transducer, Pressure, Catheter Tip devices →

Sources and updates