Skip to content

POLARCUP 75000649Recalls & reported problems

Smith & Nephew, Inc. · POLARCUP HANDLE ADAPT. 22MM TRIAL INSERT · Class 1

All models of POLARCUP (61)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07611996116632), received between Oct 14, 2020 and Nov 28, 2024 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Fracture3
  • Crack1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 28, 2024Fracture; Fracture

    IT WAS REPORTED, THAT DURING IMPACTION OF THE POLARCUP SHELL IN A THR SURGERY. ONE (1) POLARCUP HANDLE ADAPT. 22MM TRIAL INSERT FRACTURED. NO PIECES FELL INSIDE THE WOUND. THE PROCEDURE WAS RESUMED WITHOUT DELAY. USING THE SAME DEVICE. AND NO INJURIES WERE REPORTED, AS A RESULT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20805824 ↗
  • MalfunctionApr 18, 2024Break; Crack; Fracture; Break; Crack; Fracture

    IT WAS REPORTED THAT, DURING A THR SURGERY, A POLARCUP HANDLE ADAPT. 22MM TRIAL INSERT CRACKED AND BROKE DURING IMPACTION. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, WITH THE SAME DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19130509 ↗
  • MalfunctionAug 9, 2023Break; Break

    IT WAS REPORTED THAT, DURING THR SURGERY, IT WAS NOTICED THAT ONE (1) POLARCUP HANDLE ADAPT. 22MM TRIAL INSERT IS BROKEN. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17487147 ↗
  • InjuryApr 19, 2021Fracture; Fracture

    IT WAS REPORTED THAT THE ADAPTER BROKE DURING IMPLANT IMPACTION. A CHANGE IN SURGICAL TECHNIQUE WAS REQUIRED IN ORDER TO COMPLETE THE PROCEDURE. A DELAY OF 30 MIN OR LESS WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11684523 ↗
  • InjuryOct 14, 2020Break; Material Deformation; Break; Material Deformation

    IT WAS REPORTED THAT DURING A THR PROCEDURE THE POLARCUP HANDLE ADAPT. 22MM TRIAL INSERT BENT AND THEN BROKE INSIDE OF THE PATIENT. SURGERY WAS DELAYED FOR LESS THAN 30 MIN. AN S+N BACKUP DEVICE WAS AVAILABLE AND USED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10678861 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page