Skip to content

AVS TL 48370124Recalls & reported problems

Stryker Corporation · VERTEBRAL SPACER · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07613153046462), received between Nov 28, 2016 and Nov 19, 2024 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Detachment Of Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment Of Device Component1
  • Detachment of Device or Device Component1
  • Fracture1
  • Loss of or Failure to Bond1
  • Material Deformation1
  • Material Integrity Problem1
  • Mechanical Problem1

Patient problems

  • Injury1
  • No Clinical Signs, Symptoms or Conditions1
  • No Code Available1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 19, 2024Fracture; Fracture

    A COMPANY REPRESENTATIVE REPORTED THAT AN AVS TL VERTEBRAL SPACER FRACTURED INTRA-OPERATIVELY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY AND NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20719079 ↗
  • MalfunctionJan 9, 2017Loss of or Failure to Bond; Material Deformation; Material Integrity Problem; Material Deformation; Material Integrity Problem

    IT WAS REPORTED THAT AN IMPLANT WAS WARPED ON THE AREA WHERE THE INSERTER IS SUPPOSE TO ATTACH TO THE IMPLANT. BECAUSE OF THAT THE IMPLANT WOULDN'T STAY ATTACHED TO THE INSERTER IT KEEP FALLING OFF.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6234200 ↗
  • InjuryNov 28, 2016Mechanical Problem; Detachment of Device or Device Component; Detachment Of Device Component; Mechanical Problem; Detachment of Device or Device Component

    IT WAS REPORTED THAT; CAGE BECAME DIS-ATTACHED FROM THE INSERTER DURING IMPLANTATION. WHILE TRYING TO FURTHER IMPACT THE CAGE AND THEN ATTEMPTING TO REMOVE IT APPLIED PRESSURE ON THE SPINAL CORD CAUSING A DECREASE ON THE PATIENTS SENSORY DURING SSEP MONITORING. WE EVENTUALLY REMOVED THAT CAGE AND INSERTED A SECOND ONE SUCCESSFULLY. THE SECOND CAGE WAS EVENTUALLY INSERTED CORRECTLY AND THE SAME INSERTER WAS USED TO COMPLETE THE SURGERY.

    Patient problems: Injury; No Code Available

    MAUDE report 6130519 ↗

Other Intervertebral fusion device with integrated fixation, lumbar devices

All Intervertebral fusion device with integrated fixation, lumbar devices →

Sources and updates