Skip to content

BIXCUT 0225-5070Recalls & reported problems

Stryker GmbH · IM Reamer, AO Fitting · Class 1

All models of BIXCUT (34)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07613153174714), received between Feb 11, 2025 and Feb 11, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Failure to Clean Adequately” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Clean Adequately1

Patient problems

  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 11, 2025Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS REPORTED THAT "''MOST OF THE REAMERS, EVEN AFTER ALL THE MEASURES PUT IN PLACE TO CLEAN THEM, SUCH AS ULTRASONIC CLEANING, SWABBING, AND CONNECTING THEM TO A WASHER-DISINFECTOR TO WASH THE CAVITIES, STILL HAVE TRACES ON THE INSIDE THAT ARE SOMETIMES DIFFICULT TO IDENTIFY, SUCH AS BLOOD OR RED METAL. THESE REAMERS WERE REMOVED AND REPLACED WITH NEW ONES. DESPITE THE VIGILANCE THAT HAS BEEN PUT IN PLACE, IT STILL HAPPENS THAT A DM IS DELIVERED TO THE OR WITH INSUFFICIENT CLEANING, WHICH CAUSES THE DIRT TO BE DILUTED WHEN THE STEAM COMES INTO CONTACT WITH IT, SO THAT THE TRAY CANNOT BE USED - DIRTY REAMERS DETECTED WHEN THE TRAY IS OPENED IN THE OR). AFTER ALL THESE INCIDENTS AND THE NUMBER OF REAMERS REMOVED BECAUSE OF NON-COMPLIANT CLEANING, AND OUR COMMITMENT TO PROVIDING A CLEAN AND FUNCTIONAL DMX, OUR REQUEST IS THAT THE SUPPLIER/MANUFACTURER TAKE THE POSITION OF MAKING THESE DMX SINGLE USE.'' FOR 2024: WE HAD TO ORDER 51 REAMERS, ALL REFERENCES COMBINED, TO REPLACE THE NON-COMPLIANT ONES.''

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21355470 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page