NA 5820107530Recalls & reported problems
Stryker Corporation · 3.0mm Precision Neuro (Match Head) · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 07613154028177), received between Oct 24, 2018 and Oct 28, 2025 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Jun 28, 2023Open, Classified
There is potential the device inside the package may not match the product information on the outer label (e.g., label states 5820-107-430 but package contains a 5820-107-530 device or vice versa).
Z-0982-2025 · Root cause: Process control · FDA record ↗
Most reported problems
Device problems
- Break1
- Device Dislodged or Dislocated1
- Fitting Problem1
- Flaked1
- Fracture1
- Material Disintegration1
- Vibration1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Consequences Or Impact To Patient1
- No Patient Involvement1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 28, 2025Flaked; Flaked
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: FLAKED. 1 EVENT HAD NO HEALTH CONSEQUENCES OR IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23399875 ↗MalfunctionOct 2, 2025Material Disintegration; Material Disintegration
IT HAS BEEN DETERMINED THAT THIS RECORD IS NOT REPORTABLE. IT IS A DUPLICATE OF AN EVENT REPORTED IN VMSR SUBMISSION (MFR REPORT # 3015967359-2025-99384).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23198696 ↗MalfunctionAug 19, 2022Fitting Problem; Device Dislodged or Dislocated; Fitting Problem; Device Dislodged or Dislocated
IT WAS REPORTED THAT PRIOR TO A PROCEDURE, THE BUR WOULD NOT FIT THE ATTACHMENT, IT FELL OUT. IT WAS ALSO REPORTED THAT THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE AND THERE WAS NO DELAY OR ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15257764 ↗MalfunctionMay 24, 2021Vibration; Vibration
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE BUR WAS CHATTERING. IT WAS ALSO REPORTED THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11869955 ↗MalfunctionApr 27, 2020Fracture; Fracture
IT WAS REPORTED THAT DURING TESTING THE BUR BROKE IN THE ATTACHMENT. IT WAS ALSO REPORTED THAT THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE, AND THERE WAS NO DELAY OR ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 10000853 ↗MalfunctionOct 24, 2018Break; Break
THIS REPORT SUMMARIZES <NOE> 1 </NOE> MALFUNCTION EVENT IN WHICH THE DEVICE BROKE DURING A POSTERIOR SPINAL FUSION REVISION SURGICAL PROCEDURE. THERE WAS PATIENT INVOLVEMENT. THERE WAS NO PATIENT IMPACT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8001717 ↗
Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices
- NA EMAX2PLUSSynthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)2,623 reports
- MIDAS REX MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)952 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)416 reports
- MIDAS REX MR8 MR8-AS07Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)232 reports
- MIDAS REX MR8 MR8-AA10Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)192 reports
- Midas Rex MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)189 reports
- NA 5407-FA2-000Stryker Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)181 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)165 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.