Neptune 0702001000Recalls & reported problems
Stryker Corporation · 2 Rover-ULTRA 120V 60Hz 12A · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
This Stryker Neptune portable suction unit has 40 adverse event reports, listing 1 death and no injuries, mostly malfunctions such as smoke evacuator failures, sparking or burning power cords, and filtration problems. The reports mostly describe no impact on patients. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 40
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 40 adverse event reports that list this exact device (DI 07613154522309), received between Jun 4, 2018 and Oct 5, 2020 — classified as 1 death, 39 malfunctions.
The most frequently reported specific device problem is “Filtration Problem” (14 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Filtration Problem14
- Sparking9
- Smoking4
- Circuit Failure2
- Electrical /Electronic Property Problem2
- Inadequate Filtration Process2
- Increase in Suction2
- Charred1
Patient problems
- No Consequences Or Impact To Patient35
- Death1
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
- No Patient Involvement1
- Shock1
Recent adverse event reports
The 25 most recent of 40 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 5, 2020Filtration Problem; Filtration Problem
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE AT THE USER FACILITY THE DEVICE'S SMOKE EVACUATOR FAILED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10630611 ↗MalfunctionJun 22, 2020Thermal Decomposition of Device; Thermal Decomposition of Device
IT WAS REPORTED BY THE CUSTOMER THAT DURING A PROCEDURE, THE DEVICE'S POWER CORD BURNED. THE PROCEDURE WAS COMPLETED WITHOUT FURTHER USE OF THE DEVICE. NO SIGNIFICANT DELAY, NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10178370 ↗MalfunctionFeb 18, 2020Inadequate Filtration Process; Inadequate Filtration Process
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE AT THE USER FACILITY THE DEVICE'S SMOKE EVACUATOR FAILED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9722268 ↗MalfunctionFeb 18, 2020Inadequate Filtration Process; Inadequate Filtration Process
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE AT THE USER FACILITY THE DEVICE'S SMOKE EVACUATOR FAILED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9722100 ↗MalfunctionDec 13, 2019Sparking; Sparking
IT WAS REPORTED THAT AT THE USER FACILITY THE DEVICE'S POWER CORD WAS SPARKING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9469852 ↗MalfunctionJul 30, 2019Circuit Failure; Circuit Failure
IT WAS REPORTED BY THE CUSTOMER THAT DURING SET UP FOR A PROCEDURE, THE DEVICE'S PLUG SPARKED. BACK-UP EQUIPMENT WAS USED TO COMPLETE THE PROCEDURE. NO CLINICALLY SIGNIFICANT DELAY, NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8841776 ↗MalfunctionJul 22, 2019Inappropriate/Inadequate Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation
IT WAS REPORTED THAT DURING SETUP FOR A PROCEDURE AT THE USER FACILITY THE NEPTUNE ROVER SHOCKED AN EMPLOYEE. THE EMPLOYEE WAS NOT HARMED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A DIFFERENT DEVICE, WITH NO SURGICAL DELAY, MEDICAL INTERVENTION, OR ADVERSE CONSEQUENCES TO THE PATIENT OR EMPLOYEE.
Patient problems: Shock
MAUDE report 8813010 ↗MalfunctionJul 16, 2019Circuit Failure; Circuit Failure
IT WAS REPORTED BY THE CUSTOMER THAT DURING SET UP FOR A PROCEDURE, THE DEVICE'S PLUG WAS FOUND CHARRED. BACK-UP EQUIPMENT WAS USED TO COMPLETE THE PROCEDURE. NO CLINICALLY SIGNIFICANT DELAY, NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8793741 ↗MalfunctionJul 1, 2019Increase in Suction; Increase in Suction
IT WAS REPORTED BY THE CUSTOMER THAT THE NEPTUNE INCREASED IN SUCTION WITHOUT BEING PROMPTED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH THE SAME DEVICE WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8749350 ↗MalfunctionMay 9, 2019Sparking; Sparking
IT WAS REPORTED BY THE CUSTOMER THAT THE DEVICE WAS SPARKING DURING SET UP. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING SET UP AT THE USER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8596339 ↗MalfunctionApr 12, 2019Filtration Problem; Filtration Problem
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE AT THE USER FACILITY THE SMOKE EVACUATOR STOPPED FUNCTIONING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8511373 ↗MalfunctionApr 9, 2019Filtration Problem; Filtration Problem
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE AT THE USER FACILITY THE DEVICE'S SMOKE EVACUATOR FAILED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8496179 ↗MalfunctionApr 1, 2019Filtration Problem; Filtration Problem
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE AT THE USER FACILITY THE DEVICE'S SMOKE EVACUATOR FAILED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8470425 ↗MalfunctionMar 8, 2019Smoking; Smoking
IT WAS REPORTED THAT THE DEVICE WAS SMOKING FROM THE POWER CORD DURING SET UP BEFORE A PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8405522 ↗MalfunctionFeb 22, 2019Filtration Problem; Filtration Problem
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE AT THE USER FACILITY THE SMOKE EVACUATOR STOPPED FUNCTIONING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8362688 ↗MalfunctionFeb 1, 2019Sparking; Sparking
IT WAS REPORTED BY THE CUSTOMER THAT THE DEVICE WAS SPARKING AND SMOKING DURING SET UP. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING SET UP AT THE USER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8297898 ↗MalfunctionJan 2, 2019Sparking; Sparking
IT WAS REPORTED BY THE CUSTOMER THAT THE DEVICE WAS SPARKING. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE USER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8208289 ↗DeathDec 20, 2018Off-Label Use; Off-Label Use
IT WAS REPORTED BY THE CUSTOMER THAT DURING AN OPEN ABDOMEN PROCEDURE, A PATIENT EXPERIENCED CARDIOPULMONARY ARREST. PER THE CUSTOMER, AT THE TIME, THE NEPTUNE WAS USED AS A "WOUND VAC" (VACUUM ASSISTED CLOSURE) IN THE PROCEDURE, THE PATIENT NO LONGER SHOWED VITAL SIGNS. THE CUSTOMER FURTHER REPORTED THAT A CUSTOMER EMPLOYEE ATTEMPTED TO USE THE NEPTUNE DEVICE TO CREATE A "WOUND VAC" TO THE AREA WHERE BLEEDING WAS PROFUSE. THE PATIENT WAS REPORTED TO HAVE DIED DURING THE PROCEDURE. PER THE ORIGINAL REPORTER FROM THE CUSTOMER, THE NEPTUNE IN NO WAY CAUSED OR CONTRIBUTED TO THE DEMISE OF THE PATIENT. WHILE IT WAS REPORTED THAT THE NEPTUNE DID NOT CAUSE OR CONTRIBUTE TO THE PATIENT¿S CONDITION PER THE ORIGINAL REPORTER, USING THE NEPTUNE IN CONJUNCTION WITH CLOSED WOUND DRAINS IS A CONTRAINDICATION OR USAGE AND COULD RESULT IN A SERIOUS INJURY.
Patient problems: Death
MAUDE report 8185131 ↗MalfunctionDec 20, 2018Fire; Fire
IT WAS REPORTED THAT WHILE DOCKING THE ROVER AT THE USER FACILITY THE DEVICE LEAKED AND HAD AN INTERNAL ELECTRICAL FIRE. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING DOCKING OF THE ROVER AT THE USER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8184509 ↗MalfunctionDec 7, 2018Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
IT WAS REPORTED THAT AFTER A MAMMOPLASTY PROCEDURE AT THE USER FACILITY THE DEVICE WAS USED INCORRECTLY AS A SOURCE OF SUCTION IN THE CHEST AREA. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8141835 ↗MalfunctionNov 13, 2018Sparking; Sparking
IT WAS REPORTED BY THE CUSTOMER THAT THE DEVICE WAS SPARKING DURING A PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8064453 ↗MalfunctionNov 7, 2018Smoking; Sparking; Smoking; Sparking
IT WAS REPORTED BY THE CUSTOMER THAT THE DEVICE WAS SPARKING AND SMOKING FROM THE POWER CORD DURING TESTING BEFORE A PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8047786 ↗MalfunctionNov 5, 2018Sparking; Sparking
IT WAS REPORTED BY THE CUSTOMER THAT THE DEVICE WAS SPARKING DURING TESTING BEFORE A PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8039411 ↗MalfunctionOct 29, 2018Filtration Problem; Filtration Problem
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE AT THE USER FACILITY THE NEPTUNE 2 ROVER ULTRA'S SMOKE EVACUATOR FAILED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8015900 ↗MalfunctionOct 23, 2018Sparking; Sparking
IT WAS REPORTED BY THE CUSTOMER THAT THE DEVICE WAS SPARKING DURING TESTING BEFORE A PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7994964 ↗
Other APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED devices
- Penumbra System PMX110Penumbra, Inc. · APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED166 reports
- IntelliCart™ 00514010400Zimmer Orthopaedic Surgical Products, Inc. · APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED112 reports
- Penumbra System PMX220Penumbra, Inc. · APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED102 reports
- DeVilbiss 7314P-DDevilbiss Healthcare LLC · Apparatus, Suction, Ward Use, Portable, Ac-Powered69 reports
- PIRANHA 2208011Richard Wolf GmbH · APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED54 reports
- LCSU 4, 300 ml, RTCA Complete Unit 880062Laerdal Medical AS · Apparatus, Suction, Ward Use, Portable, Ac-Powered1 recall0 reports
- LCSU 4, 800 ml, RTCA Complete Unit 880052Laerdal Medical AS · Apparatus, Suction, Ward Use, Portable, Ac-Powered1 recall0 reports
- Neptune 0770100000Stryker Corporation · APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗