AXSOS 371536Recalls & reported problems
Stryker GmbH · Locking Screw · Class 2 · Implant
All models of AXSOS (184)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613154864263), received between Aug 5, 2019 and Oct 3, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Fracture” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture2
- Break1
- Difficult to Remove1
- Material Deformation1
- Positioning Problem1
Patient problems
- Device Embedded In Tissue or Plaque1
- Foreign Body In Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 3, 2024Break; Fracture; Material Deformation; Positioning Problem; Break; Fracture; Material Deformation; Positioning Problem
IT WAS REPORTED, THE THREADS ON THE TWO LOCKING GUIDES, REFERENCE (B)(4) , ARE IN POOR CONDITION. THIS ISSUE CAUSES INCORRECT SCREW ALIGNMENT BECAUSE THE SYSTEM IS UNIAXIAL AND ALTERS THE DIRECTION OF THE LOCKING SCREW. AS A RESULT, WHEN ATTEMPTING TO INSERT A 4.0X36MM LOCKING SCREW, REFERENCE (B)(4) , THE FORCE AND THE CHANGE IN THE ALIGNMENT OF THE LOCKING GUIDE CAUSED THE SCREW HEAD TO BREAK OFF, LEAVING IT INSIDE THE PATIENT. A SURGICAL DELAY OF 15-20 MINUTES WAS REPORTED.
Patient problems: Foreign Body In Patient
MAUDE report 20362947 ↗MalfunctionAug 5, 2019Fracture; Difficult to Remove
AS REPORTED: "THE HEAD OF THE SCREW WAS SPLIT WHEN TRYING TO PUT IT TO THE END, IT WAS HAPPENING MANUALLY, IN THE SAME DIRECTION AS THE GUIDE. THE SCREW IS BLOCKED DURING THE SAME PROCEDURE, ONE OF THE SCREWS DID NOT PASS PROPERLY THROUGH THE PLATE, WAS CHANGED BY ANOTHER AND THIS DID NOT ENTER IN ITS TOTALIDADE." ADDITIONALLY: "REMAINS A SCREW WITHOUT HEAD INSIDE THE PATIENT." PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT ADVERSE CONSEQUENCES.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 8862856 ↗
Other PLATE, FIXATION, BONE devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-PLF40Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-PRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page