AXSOS 371536Recalls & reported problems

Stryker GmbH · Locking Screw · Class 2 · Implant

All models of AXSOS (184)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613154864263), received between Aug 5, 2019 and Oct 3, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2
  • Break1
  • Difficult to Remove1
  • Material Deformation1
  • Positioning Problem1

Patient problems

  • Device Embedded In Tissue or Plaque1
  • Foreign Body In Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 3, 2024Break; Fracture; Material Deformation; Positioning Problem; Break; Fracture; Material Deformation; Positioning Problem

    IT WAS REPORTED, THE THREADS ON THE TWO LOCKING GUIDES, REFERENCE (B)(4) , ARE IN POOR CONDITION. THIS ISSUE CAUSES INCORRECT SCREW ALIGNMENT BECAUSE THE SYSTEM IS UNIAXIAL AND ALTERS THE DIRECTION OF THE LOCKING SCREW. AS A RESULT, WHEN ATTEMPTING TO INSERT A 4.0X36MM LOCKING SCREW, REFERENCE (B)(4) , THE FORCE AND THE CHANGE IN THE ALIGNMENT OF THE LOCKING GUIDE CAUSED THE SCREW HEAD TO BREAK OFF, LEAVING IT INSIDE THE PATIENT. A SURGICAL DELAY OF 15-20 MINUTES WAS REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 20362947 ↗
  • MalfunctionAug 5, 2019Fracture; Difficult to Remove

    AS REPORTED: "THE HEAD OF THE SCREW WAS SPLIT WHEN TRYING TO PUT IT TO THE END, IT WAS HAPPENING MANUALLY, IN THE SAME DIRECTION AS THE GUIDE. THE SCREW IS BLOCKED DURING THE SAME PROCEDURE, ONE OF THE SCREWS DID NOT PASS PROPERLY THROUGH THE PLATE, WAS CHANGED BY ANOTHER AND THIS DID NOT ENTER IN ITS TOTALIDADE." ADDITIONALLY: "REMAINS A SCREW WITHOUT HEAD INSIDE THE PATIENT." PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT ADVERSE CONSEQUENCES.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 8862856 ↗