VariAx 53-27216ERecalls & reported problems
Stryker GmbH · BONE SCREW, T7 · Class 2 · Implant
All models of VARIAX (345)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07613154913992), received between Sep 25, 2020 and Oct 27, 2020 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Material Twisted/Bent” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Twisted/Bent2
- Positioning Failure2
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 27, 2020Positioning Failure; Fracture; Material Twisted/Bent; Positioning Failure; Fracture; Material Twisted/Bent
AS REPORTED: "BY TIGHTENING ONE OF THE SCREWS, THIS GIVES UP WHAT THE SPECIALIST DOESN'T LIKE. / WHEN TAKING INTRAOPERATIVE X-RAY, TWO SCREWS MUST BE REMOVED TO RESET THEM, THE DOCTOR EXPRESSES THAT THE HEAD IS ALREADY ROLLED AND ARE ADDITIONALLY TORICIDED." ADDITIONAL INFORMATION: "1. WHEN ONE OF THE SCREWS (1ST SCREW) WAS BEING TIGHTENED IT LOST ITS HEAD, WHAT THE SPECIALIST DISLIKE. 2. WHEN TAKING INTRA-OPERATIVE RX, TWO SCREWS (2 ND AND 3 RD SCREWS )MUST BE REMOVED TO BE RE-ARRANGED ;WHEN REMOVING , THE DR EXPRESSED THAT ONE SCREW LOST ITS HEAD, IN ADDITION BOTH SCREWS WERE DISTORTED." "THERE WAS NOT HARD BONE" "OTHER SCREWS WERE USED TO COMPLETE THE PROCEDURE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10741635 ↗MalfunctionOct 27, 2020Material Twisted/Bent; Material Twisted/Bent
AS REPORTED: "BY TIGHTENING ONE OF THE SCREWS, THIS GIVES UP WHAT THE SPECIALIST DOESN'T LIKE. / WHEN TAKING INTRAOPERATIVE X-RAY, TWO SCREWS MUST BE REMOVED TO RESET THEM, THE DOCTOR EXPRESSES THAT THE HEAD IS ALREADY ROLLED AND ARE ADDITIONALLY TORICIDED." ADDITIONAL INFORMATION: "1. WHEN ONE OF THE SCREWS (1ST SCREW) WAS BEING TIGHTENED IT LOST ITS HEAD, WHAT THE SPECIALIST DISLIKE. 2. WHEN TAKING INTRA-OPERATIVE RX, TWO SCREWS (2 ND AND 3 RD SCREWS )MUST BE REMOVED TO BE RE-ARRANGED ;WHEN REMOVING , THE DR EXPRESSED THAT ONE SCREW LOST ITS HEAD, IN ADDITION BOTH SCREWS WERE DISTORTED." "THERE WAS NOT HARD BONE" "OTHER SCREWS WERE USED TO COMPLETE THE PROCEDURE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10741604 ↗MalfunctionSep 25, 2020Positioning Failure; Positioning Failure
AS REPORTED: "BY TIGHTENING ONE OF THE SCREWS, THIS GIVES UP WHAT THE SPECIALIST DOESN'T LIKE. / WHEN TAKING INTRAOPERATIVE X-RAY, TWO SCREWS MUST BE REMOVED TO RESET THEM, THE DOCTOR EXPRESSES THAT THE HEAD IS ALREADY ROLLED AND ARE ADDITIONALLY TORICIDED."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10589321 ↗
Other PLATE, FIXATION, BONE devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-PLF40Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-PRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗