Skip to content

SYNCHRO 14 M00313020Recalls & reported problems

Stryker Corporation · Neuro Guidewire with Hydrophilic Coating · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
15
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 07613252186939), received between Aug 9, 2018 and Oct 16, 2024 — classified as 15 malfunctions.

The most frequently reported specific device problem is “Peeled/Delaminated” (12 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 18, 2024Open, Classified

    Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Guidewires that may be caused by the practice of backloading the guidewire through the optional introducer accessory. This issue is limited to certain lots of the Synchro Guidewire that contain an older version of the introducer accessory. Due to variation in the manufacturing process of the supplier of the introducer accessory, certain lots of introducers have sharper than intended edges that can peel off the PTFE coating when physicians use a technique known as backloading. Users with impacted product in their inventory are cautioned not to use this backloading technique.

    Z-1939-2024 · Root cause: Component design/selection · FDA record ↗

Most reported problems

Device problems

  • Peeled/Delaminated12
  • Fracture4

Patient problems

  • No Clinical Signs, Symptoms or Conditions10
  • No Consequences Or Impact To Patient7
  • Device Embedded In Tissue or Plaque2

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 16, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE SUBJECT GUIDEWIRE TIP BROKE DURING THE PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20463332 ↗
  • MalfunctionOct 25, 2023Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT THE SUBJECT GUIDEWIRE COATING WAS PEELING OFF FROM THE PROXIMAL END. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18002530 ↗
  • MalfunctionOct 11, 2023Fracture; Fracture

    IT WAS REPORTED THAT DURING THE PROCEDURE, THE GUIDEWIRE (SUBJECT DEVICE) FRACTURED DURING USE AND THE UN-RETRIEVED PIECE OF GUIDEWIRE REMAINED IN THE PATIENT. THE PHYSICIAN REPLACED IT WITH A NEW DEVICE AND CONTINUED THE PROCEDURE WITHOUT CLINICAL CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION RECEIVED FROM COMPLAINT INTAKE CENTER ON 23-OCT-2023, CLARIFIED THAT SUBJECT GUIDEWIRE HAD FRACTURED OUTSIDE THE PATIENT.

    Patient problems: Device Embedded In Tissue or Plaque; No Clinical Signs, Symptoms or Conditions

    MAUDE report 17912179 ↗
  • MalfunctionOct 11, 2023Fracture; Fracture

    IT WAS REPORTED THAT DURING THE PROCEDURE, THE GUIDEWIRE (SUBJECT DEVICE) FRACTURED DURING USE AND THE UN-RETRIEVED PIECE OF GUIDEWIRE REMAINED IN THE PATIENT. THE PHYSICIAN REPLACED IT WITH A NEW DEVICE AND CONTINUED THE PROCEDURE WITHOUT CLINICAL CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION RECEIVED FROM COMPLAINT INTAKE CENTER ON 23-OCT-2023, CLARIFIED THAT SUBJECT GUIDEWIRE HAD FRACTURED OUTSIDE THE PATIENT.

    Patient problems: Device Embedded In Tissue or Plaque; No Clinical Signs, Symptoms or Conditions

    MAUDE report 17911797 ↗
  • MalfunctionJun 23, 2023Peeled/Delaminated; Peeled/Delaminated

    THE SUBJECT DEVICE WAS RETURNED FOR ANALYSIS AND IT WAS DISCOVERED THAT THE GUIDEWIRE SUBJECT DEVICE PTFE (POLYTETRAFLUOROETHYLENE) COATING WAS NOTED TO BE PEELING. THE PHYSICIAN REPLACED IT WITH A NEW DEVICE AND CONTINUED THE PROCEDURE WITHOUT CLINICAL CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17192061 ↗
  • MalfunctionJan 25, 2022Peeled/Delaminated; Peeled/Delaminated

    ANALYSIS OF THE RETURNED DEVICE FOUND THE SUBJECT GUIDEWIRE (PTFE) POLYTETRAFLUOROETHYLENE COATING WAS PEELING/PEELED. THERE WERE NO CLINICAL CONSEQUENCES TO THE PATIENT REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13344865 ↗
  • MalfunctionDec 2, 2021Peeled/Delaminated; Peeled/Delaminated

    ANALYSIS OF THE RETURNED DEVICE FOUND THE SUBJECT GUIDEWIRE PTFE COATING WAS PEELING/PEELED. THERE WERE NO CLINICAL CONSEQUENCES TO THE PATIENT REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12921885 ↗
  • MalfunctionJun 28, 2021Fracture; Peeled/Delaminated; Fracture; Peeled/Delaminated

    THE DEVICE WAS RETURNED FOR ANALYSIS AND THE INVESTIGATION OF THE DEVICE REVEALED THAT THE GUIDEWIRE (SUBJECT DEVICE) DISTAL TIP WAS BROKEN DURING USE AND THE PTFE COATING WAS PEELING OR HAD PEELED AT THE PROXIMAL END. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12081703 ↗
  • MalfunctionSep 8, 2020Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT DURING THE PROCEDURE WHILE THE PHYSICIAN REMOVED THE SUBJECT GUIDE WIRE FROM THE MICROCATHETER ITS PTFE COATING PEELED OFF FROM THE DISTAL END.THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 10505761 ↗
  • MalfunctionJun 23, 2020Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT DURING THE PROCEDURE WHILE USING THE SUBJECT GUIDE WIRE, THE PHYSICIAN ENCOUNTERED RESISTANCE WHILE INSERTING THROUGH A BALLOON CATHETER. THE COATING ON THE DISTAL SUBJECT GUIDE WIRE WAS ALSO COMING OFF. ACCORDING TO THE PHYSICIAN, NO COATING WAS LEFT INSIDE THE PATIENT. THE PHYSICIAN REPLACED THE SUBJECT GUIDE WIRE WITH A NEW DEVICE AND CONTINUED THE PROCEDURE WITHOUT CLINICAL CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 10184786 ↗
  • MalfunctionFeb 13, 2020Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT DURING PROCEDURE, WHEN THE PHYSICIAN TRIED TO REINSERT THE GUIDEWIRE (SUBJECT DEVICE) INTO THE MICROCATHETER, THE COATING OF THE GUIDEWIRE PEELED OFF. THERE WERE NO CLINICAL CONSEQUENCES TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9709169 ↗
  • MalfunctionJul 29, 2019Peeled/Delaminated; Peeled/Delaminated

    DURING THE PROCEDURE, IT WAS REPORTED THAT THE GUIDEWIRE PTFE (POLYTETRAFLUOROETHYLENE) COATING WAS PEELED OFF. HENCE THE GUIDEWIRE WAS CHANGED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY AFTER. NO CLINICAL CONSEQUENCES REPORTED TO PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8834830 ↗
  • MalfunctionFeb 1, 2019Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT DURING PREPARATION FOR THE PROCEDURE, THE POLYTETRAFLUOROETHYLENE (PTFE) COATING ON THE GUIDEWIRE HAD PEELED. THE PHYSICIAN REPLACED IT WITH A NEW GUIDEWIRE AND CONTINUED THE PROCEDURE WITHOUT CLINICAL CONSEQUENCES TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8299062 ↗
  • MalfunctionSep 8, 2018Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT WHILE TRYING TO INSERT THE SUBJECT DEVICE (GUIDEWIRE) TO MICROCATHETER, THE HYDROPHILIC COATING WAS REMOVED/PEELED OFF. THE PROCEDURE WAS COMPLETED WITH OTHER WIRE. THERE WERE NO REPORTED CLINICAL CONSEQUENCES TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7858442 ↗
  • MalfunctionAug 9, 2018Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT FLAKING OF COATING OF THE GUIDEWIRE WAS OBSERVED BEFORE THE WIRE WAS INTRODUCED IN THE PATIENT. NO CONSEQUENCES TO THE PATIENT WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7770533 ↗

Other WIRE, GUIDE, CATHETER devices

All WIRE, GUIDE, CATHETER devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.