The Synchro Neuro Guidewires are a steerable guidewire… recallZ-1939-2024 · Stryker Neurovascular · initiated Apr 18, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Stryker Neurovascular or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Guidewires that may be caused by the practice of backloading the guidewire through the optional introducer accessory. This issue is limited to certain lots of the Synchro Guidewire that contain an older version of the introducer accessory. Due to variation in the manufacturing process of the supplier of the introducer accessory, certain lots of introducers have sharper than intended edges that can peel off the PTFE coating when physicians use a technique known as backloading. Users with impacted product in their inventory are cautioned not to use this backloading technique.
FDA root cause category: Component design/selection
What the company is doing
On 4/29/24, recall notices were mailed to Supply Chain Management/ Recall Coordinator/ Inventory Manager customers who were asked to do the following: 1) When using impacted guidewires, please do not backload the guidewire, follow the Instructions for Use (IFU) provided with the device, and do not use the guidewire if you see any visible signs of coating damage. 2) Circulate this customer notification internally to all interested/affected parties. 3) Inform firm if any of the subject devices have been distributed to other organizations. If yes, provide contact details so that the firm can inform the recipients appropriately. 4) Inform firm of any adverse events concerning the use of the subject devices. (US Customer Service: 1-855-91-NEURO (1-855-916-3876), Email: [email protected]) 5) Complete and return the customer response form via email to: [email protected]. 6) Contact the firm if you would like a replacement of the guidewire with a new optional introducer accessory.
Recall details
- Recall number
- Z-1939-2024
- Status
- Open, Classified
- Initiated by the company
- Apr 18, 2024
- Posted by the FDA
- May 31, 2024
- Recalling firm
- Stryker Neurovascular
- Quantity
- 105,286
- Product code
- DQX
- FDA submissions
- K032146
- Distribution
- Worldwide - US Nationwide distribution including in the states of TX, NY, MD, CA, KY, WI, OR, NE, SC, MA, MO, CT, ID, PA, TN, FL, NC, AL, IA, OK, VA, IL, OH, NH, MN, ND, KS, NJ, WV, IN, GA, HI, AR, UT, LA, MI, WA, NV, CO, AZ, DC, VT, ME, DE, AK, WY, MT, NM, SD, MS, RI, GU and the countries of ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BRAZIL, BULGARIA, CANADA, CHINA, COLOMBIA, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, HUNGARY, ICELAND, INDIA, IRAN, IRELAND, ISRAEL, ITALIA, JAPAN, JORDAN, KUWAIT, LITHUANIA, LUXEMBOURG, MEXICO, NETHERLANDS, NORTHERN IRELAND, NORWAY, PHILIPPINES, POLAND, PORTUGAL, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UAE, UK.
Product description
The Synchro Neuro Guidewires are a steerable guidewire family with a shapeable tip. The outside diameter of the Synchro 10 guidewire is 0.010in. The outside diameter of the Syncrho 14 guidewire is 0.014in. The guidewires are compatible with existing microcatheters used in common procedures such as those used in endovascular diagnosis and therapy of neurovascular disease. The distal portion of the guidewire tip is radiopaque. For lubricity, the distal portion of the device is coated with a hydrophilic polymer and the proximal portion of the guidewire is coated with polytetrafluoroethylene (PTFE). The torque device included with the guidewire attaches to the proximal end of the wire and functions as a steering guide. Rotation of this device facilitates guidewire placement into the appropriate vessel by precise directional manipulation of the guidewire tip. The introducer included with the guidewire is intended to aid insertion of the guidewire into the catheter hub and/or hemostasis valve.
Code information (models, lots, serial numbers)
Show the full code information (4k characters)
Catalog/UDI-DI/Lots: M00316310/07613252187158/0000131357, 0000132153, 0000132740, 0000135026, 0000138147, 0000138148, 0000138579, 0000142533, 0000144961, 0000145874, 0000154419, 0000156324, 0000156614, 0000158907, 0000158911, 0000164603, 0000173346, 0000173701, 0000173702, 0000173771, 0000173995; M00316330/07613252187165/0000142692, 0000158995; M00313010/07613252186922/0000129027, 0000129028, 0000129029, 0000129218, 0000129219, 0000129630, 0000129631, 0000129841, 0000129842, 0000129902, 0000130225, 0000130226, 0000130410, 0000130411, 0000130412, 0000130502, 0000130503, 0000130643, 0000130645, 0000130869, 0000130870, 0000131304, 0000131307, 0000131343, 0000131724, 0000131725, 0000131726, 0000132154, 0000132155, 0000132582, 0000132640, 0000132641, 0000133182, 0000133183, 0000133391, 0000133565, 0000133566, 0000133567, 0000133568, 0000133928, 0000133929, 0000134128, 0000134129, 0000134130, 0000134307, 0000134308, 0000135009, 0000135148, 0000135149, 0000135303, 0000135304, 0000135806, 0000135807, 0000136252, 0000136923, 0000137267, 0000137322, 0000137373, 0000137374, 0000137375, 0000138751, 0000139437, 0000141635, 0000142248, 0000142249, 0000142541, 0000142691, 0000143123, 0000143350, 0000143351, 0000143585, 0000143955, 0000144141, 0000144320, 0000144379, 0000144380, 0000144739, 0000144760, 0000144761, 0000144764, 0000144959, 0000144960, 0000145884, 0000145885, 0000146228, 0000146390, 0000146515, 0000146843, 0000146916, 0000147040, 0000147186, 0000147306, 0000147440, 0000147585, 0000147876, 0000148965, 0000149562, 0000150095, 0000150872, 0000150873, 0000151381, 0000151461, 0000151677, 0000151847, 0000151848, 0000152230, 0000152231, 0000152375, 0000152522, 0000152977, 0000153282, 0000153370, 0000153970, 0000154140, 0000154302, 0000154303, 0000154779, 0000154780, 0000154894, 0000154895, 0000155020, 0000155068, 0000156317, 0000156321, 0000156615, 0000156626, 0000156794, 0000156800, 0000157226, 0000157227, 0000157892, 0000157894, 0000158127, 0000158128, 0000158131, 0000158339, 0000158340, 0000159324, 0000159932, 0000159997, 0000159998, 0000160368, 0000160369, 0000160434, 0000160435, 0000160893, 0000160894, 0000160895, 0000161061, 0000161062, 0000161063, 0000162104, 0000162105, 0000162106, 0000162896, 0000162991, 0000162992, 0000163716, 0000163967, 0000163968, 0000164443, 0000164470, 0000165356, 0000165363, 0000165364, 0000166540, 0000166542, 0000166544, 0000170149, 0000170246, 0000170940, 0000171065, 0000171235, 0000172266, 0000172267, 0000172381, 0000172787, 0000172865, 0000172866, 0000173279, 0000174292, 0000174373, 0000174380, 0000174409, 0000174680, 0000174681, 0000175067, 0000175068, 0000175069, 0000175640, 0000175641, 0000175983, 0000175984, 0000175985, 0000175986, 0000176344, 0000176345, 0000176504, 0000176523, 0000176830, 0000176836, 0000177535, 0000178075, 0000178076, 0000178532, 0000178533, 0000178535, 0000178626, 0000179628, 0000179629, 0000180361, 0000180738, 0000180800, 0000193546; M00313020/07613252186939/0000159372, 0000172593; M00313310/07613252186946/0000135007, 0000135008, 0000136409, 0000136410,0000136679, 0000139937, 0000139938, 0000140307, 0000140689, 0000140690, 0000141634, 0000141833, 0000142246, 0000152799, 0000153160, 0000153453, 0000153968, 0000153969, 0000154301, 0000157222, 0000157223, 0000157224, 0000159320, 0000159370, 0000160740, 0000172380, 0000172594, 0000172619, 0000172661, 0000172662, 0000172748, 0000172786, 0000176476, 0000176840, 0000176841, 0000177540, 0000178232, 0000180657, 0000180737, 0000180795; M00313410/07613252186960/0000173997
Devices named in this recall
6 devices in the FDA device registry whose identifier appears in the recall's code information.
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- SYNCHRO 14 M00313310Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls64 reports
- SYNCHRO 10 M00316310Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls44 reports
- SYNCHRO 10 M00316330Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls4 reports
- SYNCHRO 14 M00313020Stryker Corporation · WIRE, GUIDE, CATHETER1 recall15 reports
- SYNCHRO 14 S M00313410Stryker Corporation · WIRE, GUIDE, CATHETER1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 94543).
- Z-1940-2024 · Open, ClassifiedThe Synchro SELECT Guidewires are a steerable guidewire family with a shapeable tip and are available in straight and pre-shaped versions. The outside diameters of the guidewires a…
- Z-1941-2024 · Open, ClassifiedThe Synchro2 Guidewires are a steerable guidewire family with a shapeable tip and are available in straight and pre-shaped versions. The outside diameters of the guidewires are 0.0…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page