ANCHORAGE PLP11342Recalls & reported problems
Stryker GmbH · Plate MTP, right · Class 2 · Implant
All models of ANCHORAGE (52)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 07613252260516), received between Feb 20, 2019 and Mar 3, 2026 — classified as 3 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Device Damaged by Another Device” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Device Damaged by Another Device1
- Fitting Problem1
- Material Fragmentation1
- Material Protrusion/Extrusion1
- Positioning Failure1
- Structural Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Insufficient Information1
- No Known Impact Or Consequence To Patient1
- Nonunion/Delayed-union Bone Fracture1
- Pain1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryMar 3, 2026Structural Problem
THE MANUFACTURER RECEIVED RWE REPORT NAMED "SAFETY AND PERFORMANCE OF THE ANCHORAGE AND ANCHORAGE 2 CP IMPLANTS UTILIZING THE PREMIER HEALTHCARE DATABASE" THAT CONTAINS COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF ANCHORAGE AND ANCHORAGE 2 CP. THE REPORT DETAILS ANALYSIS PROVIDED FOR PROCEDURES PERFORMED BETWEEN 1 JAN 2016 ¿ 30 MAR 2025 DURING THE REVIEW OF THE REPORT, IT WAS NOT POSSIBLE TO ESTABLISH A SPECIFIC DEVICE DETAIL, PATIENT INFORMATION, AND CURRENTLY NO ADDITIONAL DEVICE INFORMATION IS AVAILABLE; HOWEVER, PATIENT 138 HAD STRUCTURAL PROBLEM.
Patient problems: Insufficient Information
MAUDE report 24497649 ↗InjuryMar 3, 2026Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED RWE REPORT NAMED "SAFETY AND PERFORMANCE OF THE ANCHORAGE AND ANCHORAGE 2 CP IMPLANTS UTILIZING THE PREMIER HEALTHCARE DATABASE" THAT CONTAINS COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF ANCHORAGE AND ANCHORAGE 2 CP. THE REPORT DETAILS ANALYSIS PROVIDED FOR PROCEDURES PERFORMED BETWEEN 1 JAN 2016 ¿ 30 MAR 2025 DURING THE REVIEW OF THE REPORT, IT WAS NOT POSSIBLE TO ESTABLISH A SPECIFIC DEVICE DETAIL, PATIENT INFORMATION, AND CURRENTLY NO ADDITIONAL DEVICE INFORMATION IS AVAILABLE; HOWEVER, PATIENT 146 HAD GENERALIZED PAIN.
Patient problems: Pain
MAUDE report 24495333 ↗InjuryMar 3, 2026Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED RWE REPORT NAMED "SAFETY AND PERFORMANCE OF THE ANCHORAGE AND ANCHORAGE 2 CP IMPLANTS UTILIZING THE PREMIER HEALTHCARE DATABASE" THAT CONTAINS COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF ANCHORAGE AND ANCHORAGE 2 CP. THE REPORT DETAILS ANALYSIS PROVIDED FOR PROCEDURES PERFORMED BETWEEN 1 JAN 2016 ¿ 30 MAR 2025 DURING THE REVIEW OF THE REPORT, IT WAS NOT POSSIBLE TO ESTABLISH A SPECIFIC DEVICE DETAIL, PATIENT INFORMATION, AND CURRENTLY NO ADDITIONAL DEVICE INFORMATION IS AVAILABLE; HOWEVER, PATIENT 149 HAD NONUNION.
Patient problems: Nonunion/Delayed-union Bone Fracture
MAUDE report 24495054 ↗MalfunctionJan 26, 2021Material Fragmentation; Device Damaged by Another Device; Material Fragmentation; Device Damaged by Another Device
AS REPORTED: "PRIMARY PROCEDURE, RIGHT FIRST MTP. IT WAS REPORTED THAT WHEN THE DRILL GUIDE WAS SEATED INTO A HOLE OF THE PLATE AND AFTER DRILLING WAS DONE, A THREAD-LIKE PIECE OF METAL WAS SEEN IN THE PATIENT. THE PIECE WAS REMOVED BY THE SURGEON. SURGEON REPORTED THAT IT FELT AS THOUGH THE DRILL GUIDE DID NOT SEEM TO BE SEATING AS DEEPLY IN THE HOLE AS IT HAD IN THE PREVIOUS HOLES, AND BELIEVES THAT THE THREAD OF METAL WAS SHAVED OFF OF THE PLATE WHILE DRILLING. PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11229114 ↗MalfunctionDec 22, 2020Positioning Failure; Positioning Failure
IT WAS REPORTED THAT, "THE SCREWS COULD NOT BE SCREWED ON THE PLATE DELAY OF INTERVENTION; NO PATIENT CONSEQUENCES."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11060307 ↗MalfunctionFeb 20, 2019Fitting Problem; Material Protrusion/Extrusion
THE FOLLOWING EVENT WAS REPORTED BY CLINIC: DESPITE THE USE OF THE BARREL ON THE PLATE ANCHORAGE MTP RIGHT (REF PLP11342), THE SCREW DIAMETER 3 LOCKING DOES NOT FIT CORRECTLY IN THE HOLE. THREADING PROBLEM WAS OBSERVED AND 1 WIRE CAME OUT OF THE SCREW. THE SCREW PROTRUDES FROM THE PLATE AND A METAL WIRE COMES OUT OF THE SCREW. THE SCREW WAS REMOVED AND ANOTHER SCREW REINSERTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8356235 ↗
Other PLATE, FIXATION, BONE devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-PLF40Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-PRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.