AXSOS 661440Recalls & reported problems
Stryker GmbH · Cortex Screw · Class 2 · Implant
All models of AXSOS (184)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07613252580416), received between Dec 2, 2019 and Sep 30, 2020 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Defective Device” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Defective Device2
- Break1
- Device Handling Problem1
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Device Embedded In Tissue or Plaque1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2020Defective Device; Defective Device
THE FOLLOWING EVENT WAS REPORTED THAT: "WE HAVE JUST RECEIVED A MATERIAL OVIGILANCE CONCERNING THE AXSOS CORTICAL SCREWS. ON THE RADIO, THE SCREWS WERE BENT, AND THE THREADS WERE TURNING. SCREWS CONCERNED: 2 SCREWS BENT DURING PLACEMENT: 1 CORTICAL 3.5 IN 32MM AND 1 CORTICAL 3.5 IN 40MM: AND 2 SCREWS WITH BAD THREAD: 2 CORTICAL 3.5 IN 40MM AND 2 LOCKED SCREWS IN 28MM. CONSEQUENCES: THE SURGEON REMOVED THE SCREWS AND PUT IN NEW ONES.¿
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10609916 ↗MalfunctionSep 30, 2020Defective Device; Defective Device
THE FOLLOWING EVENT WAS REPORTED THAT: "WE HAVE JUST RECEIVED A MATERIALOVIGILANCE CONCERNING THE AXSOS CORTICAL SCREWS. ON THE RADIO, THE SCREWS WERE BENT, AND THE THREADS WERE TURNING. SCREWS CONCERNED: 2 SCREWS BENT DURING PLACEMENT: 1 CORTICAL 3.5 IN 32MM AND 1 CORTICAL 3.5 IN 40MM: AND 2 SCREWS WITH BAD THREAD: 2 CORTICAL 3.5 IN 40MM AND 2 LOCKED SCREWS IN 28MM. CONSEQUENCES: THE SURGEON REMOVED THE SCREWS AND PUT IN NEW ONES.¿
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10609887 ↗MalfunctionDec 2, 2019Break; Material Twisted/Bent; Device Handling Problem; Break; Material Twisted/Bent; Device Handling Problem
PRIMARY LEFT TIBIA PROCEDURE. IT WAS REPORTED THAT THE 3.5 CORTICAL SCREW WAS BEING INSERTED OUTSIDE OF THE 20° CONE OF ANGULATION IN THE SHAFT AND WAS VISIBLY BENDING DURING INSERTION. IT IS NOT KNOWN IF THE SCREW WAS ALIGNED WITH THE DRILL HOLE IN THE FAR CORTEX. THE SCREW BROKE INSIDE THE PATIENT. AS THE HEAD WAS ENGAGED WITH THE PLATE AND WAS NOT SITTING PROUD, THE SURGEON MADE THE DECISION TO LEAVE THE ENTIRETY OF THE SCREW IN THE PATIENT. SURGERY WAS COMPLETED SUCCESSFULLY WITH NO DELAY. THE REP MAY BE ABLE TO OBTAIN X-RAYS, AND OTHERWISE CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Device Embedded In Tissue or Plaque; No Known Impact Or Consequence To Patient
MAUDE report 9394774 ↗
Other Plate, fixation, bone devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-PLF40Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-PRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page