HOFFMANN 4933-9-100Recalls & reported problems
Stryker GmbH · Wire Tensioner LRF · Class 1
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07613252611035), received between Apr 8, 2024 and Apr 8, 2024 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Use of Device Problem” (5 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- May 9, 2024Open, Classified
The adjustment ring on the device may become loose resulting in the wire tension not being able to be set correctly.
Z-2219-2024 · Root cause: Process change control · FDA record ↗
Most reported problems
Device problems
- Use of Device Problem5
- Break1
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 8, 2024Break; Fracture; Use of Device Problem; Break; Fracture; Use of Device Problem
AS REPORTED: " DURING THE CLEANING AND STERILIZATION , INSTRUMENT HAS BEEN BROKEN. DO NOT KNOW WHAT HAPPENED, BUT PARTS DO NOT FIT LIKE THEY HAVE DONE BEFORE. THEY HAVE USED CLEANING PROTOCOL ADVICE WE HAVE IN WRITING (IFU). NO HARM IN OPERATION WHEN USED, ONLY IN CLEANING PROCESS HAVE NOTICED BROKEN DEVICE. NO PATIENT HARM OR DELAY."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19067470 ↗MalfunctionApr 8, 2024Use of Device Problem; Use of Device Problem
IT IS REPORTED THAT THE PRODUCTS WOULD NOT SHOW HOW MUCH TRACTION WAS BEING PUT THROUGH (POSSIBLE MISSING PART).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19067342 ↗MalfunctionApr 8, 2024Use of Device Problem; Use of Device Problem
IT IS REPORTED THAT THE PRODUCTS WOULD NOT SHOW HOW MUCH TRACTION WAS BEING PUT THROUGH (POSSIBLE MISSING PART).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19067316 ↗MalfunctionApr 8, 2024Use of Device Problem; Use of Device Problem
IT IS REPORTED THAT THE PRODUCTS WOULD NOT SHOW HOW MUCH TRACTION WAS BEING PUT THROUGH (POSSIBLE MISSING PART).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19067118 ↗MalfunctionApr 8, 2024Use of Device Problem; Use of Device Problem
IT IS REPORTED THAT THE PRODUCTS WOULD NOT SHOW HOW MUCH TRACTION WAS BEING PUT THROUGH (POSSIBLE MISSING PART).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19066951 ↗
Other Orthopedic Manual Surgical Instrument devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.