ES2 SPINAL SYSTEM 482804645Recalls & reported problems

Stryker Corporation · LONG BLADE CANNULATED SCREW · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 07613327002188), received between Jun 25, 2018 and Nov 22, 2025 — classified as 8 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Fracture” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture5
  • Migration5
  • Biocompatibility2
  • Break2
  • Migration or Expulsion of Device2
  • Device Dislodged or Dislocated1
  • Device Operates Differently Than Expected1

Patient problems

  • Injury7
  • No Clinical Signs, Symptoms or Conditions4
  • Pain3
  • Neuropathy2
  • Numbness2
  • Failure of Implant1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 22, 2025Fracture; Fracture

    A COMPANY REPRESENTATIVE REPORTED THAT TWO ES2 INTEGRATED SCREWS FRACTURED POST-OPERATIVELY. NO ADVERSE CONSEQUENCES WERE REPORTED. THE FRACTURED SCREWS WERE OBSERVED IN A PLANNED REMOVAL SURGERY APPROXIMATELY THREE YEARS POST-OPERATIVELY. THE PROXIMAL PORTION OF THE SCREWS WERE EXPLANTED; HOWEVER THE DISTAL SHANK OF THE SCREWS REMAINED IN THE PATIENT. THIS REPORT CAPTURES THE SECOND OF TWO SCREWS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23625683 ↗
  • MalfunctionNov 22, 2025Fracture; Fracture

    A COMPANY REPRESENTATIVE REPORTED THAT TWO ES2 INTEGRATED SCREWS FRACTURED POST-OPERATIVELY. NO ADVERSE CONSEQUENCES WERE REPORTED. THE FRACTURED SCREWS WERE OBSERVED IN A PLANNED REMOVAL SURGERY APPROXIMATELY THREE YEARS POST-OPERATIVELY. THE PROXIMAL PORTION OF THE SCREWS WERE EXPLANTED, HOWEVER THE DISTAL SHANK OF THE SCREWS REMAINED IN THE PATIENT. THIS REPORT CAPTURES THE FIRST OF TWO SCREWS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23625679 ↗
  • InjuryJan 21, 2022Migration; Migration

    IT IS REPORTED THAT REVISION SURGERY WAS PERFORMED TO REMOVE AN ES2 SCREW THAT HAD MIGRATED AND WAS NOW CAUSING THE PATIENT PAIN. DURING THE REVISION SURGERY, WHILE THE SURGEON WAS REMOVING A DIFFERENT SCREW, THE SCREW APPEARED TO HAVE FRACTURED. UPON REVIEW OF THE PRE-OPERATIVE X-RAY, THE SURGEON NOTICED THAT THE OTHER ES2 SCREW HAD FRACTURED BEFORE THE REVISION SURGERY. BOTH SCREWS WERE REMOVED. THIS REPORT CAPTURES THE MIGRATED SCREW.

    Patient problems: Pain

    MAUDE report 13315323 ↗
  • InjurySep 29, 2020Migration; Migration

    A PHYSICIAN REPORTED THAT A ES2 INTEGRATED BLADE SCREW AT S1 FRACTURED AND THE CONTRALATERAL ES2 INTEGRATED BLADE SCREW AT WAS LOOSE IN THE BONE. THE PATIENT HAS BEEN REVISED. THIS REPORT REPRESENTS THE MIGRATED SCREW.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10602865 ↗
  • InjurySep 29, 2020Fracture; Fracture

    A PHYSICIAN REPORTED THAT A ES2 INTEGRATED BLADE SCREW AT S1 FRACTURED AND THE CONTRALATERAL ES2 INTEGRATED BLADE SCREW AT WAS LOOSE IN THE BONE. THE PATIENT HAS BEEN REVISED. THIS REPORT REPRESENTS THE FRACTURED SCREW.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10602740 ↗
  • InjuryJul 9, 2020Biocompatibility; Biocompatibility

    A SURGEON REPORTED THAT A PATIENT MAY HAVE EXPERIENCED METALLOSIS AT THE L4 LEVEL WHILST IMPLANTED WITH TWO ES2 SCREWS AND TWO ES2 HEX RODS. THE PATIENT WAS UNDERGOING A SCHEDULED REVISION SURGERY TO EXTEND THE CONSTRUCT TO S1 WHEN THE SURGEON NOTED, "METALLOSIS AROUND THE L4 SCREWS," AND, "SOME EVIDENCE OF DISCOLOURATION OF THE SURROUNDING TISSUE ADJACENT TO THE SCREWS." SURGERY WAS COMPLETED AS PLANNED; THE RODS WERE EXPLANTED AND REPLACED WITH LONGER RODS AND THE L4 SCREWS WERE REPLACED. THIS REPORT REPRESENTS THE FIRST OF THE TWO SCREWS.

    Patient problems: Injury

    MAUDE report 10254814 ↗
  • InjuryJul 9, 2020Biocompatibility; Biocompatibility

    A SURGEON REPORTED THAT A PATIENT MAY HAVE EXPERIENCED METALLOSIS AT THE L4 LEVEL WHILST IMPLANTED WITH TWO ES2 SCREWS AND TWO ES2 HEX RODS. THE PATIENT WAS UNDERGOING A SCHEDULED REVISION SURGERY TO EXTEND THE CONSTRUCT TO S1 WHEN THE SURGEON NOTED, "METALLOSIS AROUND THE L4 SCREWS," AND, "SOME EVIDENCE OF DISCOLOURATION OF THE SURROUNDING TISSUE ADJACENT TO THE SCREWS." SURGERY WAS COMPLETED AS PLANNED; THE RODS WERE EXPLANTED AND REPLACED WITH LONGER RODS AND THE L4 SCREWS WERE REPLACED. THIS REPORT REPRESENTS THE SECOND OF THE TWO SCREWS.

    Patient problems: Injury

    MAUDE report 10254714 ↗
  • MalfunctionNov 27, 2019Migration; Migration or Expulsion of Device; Migration

    COMPANY REPRESENTATIVE REPORTED THAT DURING AN EXTENSION OF FUSION, AN ES2 BLOCKER WAS FOUND TO MIGRATED/PULLED-OUT. THE DEVICE WAS REPLACED.

    Patient problems: Injury; No Consequences Or Impact To Patient

    MAUDE report 9387034 ↗
  • MalfunctionJun 18, 2019Fracture; Break; Fracture

    TWO SCREWS WERE REPORTED TO HAVE BROKEN POST-OPERATIVELY. ONE SCREW WAS NOTED BROKEN VIA X-RAY AND THE OTHER SCREW WAS NOTED BROKEN DURING THE REVISION SURGERY. IT WAS REPORTED THAT THE RIGHT-SIDE SCREW WAS REPLACED BUT THAT THE LEFT-SIDE SCREW WAS NOT REPLACED BECAUSE FUSION HAD OCCURRED. THIS RECORD REPRESENTS THE LEFT-SIDE SCREW.

    Patient problems: Injury

    MAUDE report 8707623 ↗
  • InjuryJun 18, 2019Fracture; Break; Fracture

    TWO SCREWS WERE REPORTED TO HAVE BROKEN POST-OPERATIVELY. ONE SCREW WAS NOTED BROKEN VIA X-RAY AND THE OTHER SCREW WAS NOTED BROKEN DURING THE REVISION SURGERY. IT WAS REPORTED THAT THE RIGHT-SIDE SCREW WAS REPLACED BUT THAT THE LEFT-SIDE SCREW WAS NOT REPLACED BECAUSE FUSION HAD OCCURRED. THIS RECORD REPRESENTS THE RIGHT-SIDE SCREW.

    Patient problems: Injury

    MAUDE report 8707400 ↗
  • InjuryJun 25, 2018Migration or Expulsion of Device; Device Dislodged or Dislocated; Migration; Migration

    THE PATIENT REPORTED THAT A CAGE WHICH WAS IMPLANTED ON (B)(6) 2014, HAD SLID OUT FROM L4/L5 AND HE WILL UNDERGO CORRECTIVE SURGERY (B)(6) 2018 DUE TO EXCRUCIATING PAIN. PATIENT REPORTED THE SAME TYPE OF PAIN RADIATING DOWN RIGHT LEG, NEUROPATHY, AND NUMBNESS IN RIGHT FOOT, WHICH GOT WORSE SINCE (B)(6) 2014, WHEN SYMPTOMS STARTED. DEVICE WILL BE REMOVED. IN ADDITION, PATIENT REPORTED THAT X-RAYS SHOW, CAGE MOVED ABOUT 3/4 INCH TOWARD POSTERIOR AND RIGHT AND PHYSICIAN STATED, "IT IS PARTIALLY FUSED IN A BAD PLACE. HE HAS TO CUT IT OUT." PATIENT ADDITIONALLY REPORTED THAT, "THE CT SCANS SHOW HALO AROUND SCREWS WHICH MIGHT INDICATE LOOSENING ACCORDING TO THE PHYSICIAN."

    Patient problems: Failure of Implant; Injury; Neuropathy; Numbness; Pain

    MAUDE report 7635291 ↗
  • InjuryJun 25, 2018Device Operates Differently Than Expected; Migration; Migration

    THE PATIENT REPORTED THAT A CAGE WHICH WAS IMPLANTED ON (B)(6) 2014, HAD SLID OUT FROM L4/L5 AND HE WILL UNDERGO CORRECTIVE SURGERY(B)(6) 2018 DUE TO EXCRUCIATING PAIN. PATIENT REPORTED THE SAME TYPE OF PAIN RADIATING DOWN RIGHT LEG, NEUROPATHY, AND NUMBNESS IN RIGHT FOOT, WHICH GOT WORSE SINCE (B)(6) 2014, WHEN SYMPTOMS STARTED. DEVICE WILL BE REMOVED. IN ADDITION, PATIENT REPORTED THAT X-RAYS SHOW, CAGE MOVED ABOUT 3/4 INCH TOWARD POSTERIOR AND RIGHT AND PHYSICIAN STATED, "IT IS PARTIALLY FUSED IN A BAD PLACE. HE HAS TO CUT IT OUT." PATIENT ADDITIONALLY REPORTED THAT, "THE CT SCANS SHOW HALO AROUND SCREWS WHICH MIGHT INDICATE LOOSENING ACCORDING TO THE PHYSICIAN."

    Patient problems: Injury; Neuropathy; Numbness; Pain

    MAUDE report 7635183 ↗

Other Orthosis, spondylolisthesis spinal fixation devices

All Orthosis, spondylolisthesis spinal fixation devices →