Skip to content

TARGET XL M0036116200Recalls & reported problems

Stryker Corporation · Helical Detachable Coil · Class 2 · Implant

All models of TARGET XL (34)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613327004588), received between Mar 6, 2020 and Jan 25, 2022 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Fracture1
  • Premature Separation1

Patient problems

  • Device Embedded In Tissue or Plaque1
  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 25, 2022Premature Separation; Premature Separation

    THE SUBJECT COIL WAS RETURNED FOR ANALYSIS AND IT WAS DISCOVERED THAT THE SUBJECT COIL WAS PREMATURELY DETACHED/SEPARATED DURING USE. THERE WERE NO CLINICAL CONSEQUENCES TO THE PATIENT REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13344023 ↗
  • MalfunctionMar 6, 2020Fracture; Fracture

    IT WAS REPORTED THAT DURING THE CASE OF A CAROTID BLOWOUT FROM PRIOR RADIATION OF A NECK TUMOR, THE CAROTID WAS SUCCESSFULLY SHUTDOWN WITH A TWO CATHETER TECHNIQUE USING THE MICROCATHETERS AND COILS, AND THE SUBJECT COIL DETACHED PREMATURELY FROM THE MIDDLE SECTION INSIDE THE PATIENT¿S ANATOMY DURING PROCEDURE. THE COIL PUSHER AND A PORTION OF THE COIL WERE SUCCESSFULLY RETRACTED BY THE PHYSICIAN. A PORTION OF THE DETACHED COIL STAYED INSIDE THE PATIENT¿S ANATOMY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: Device Embedded In Tissue or Plaque; No Consequences Or Impact To Patient

    MAUDE report 9803277 ↗

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page