Skip to content

TARGET XL M0036102050Recalls & reported problems

Stryker Corporation · 360 Standard Detachable Coil · Class 2 · Implant

All models of TARGET XL (34)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 07613327004731), received between Mar 14, 2017 and Aug 30, 2024 — classified as 3 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Premature Separation” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Premature Separation4
  • Adverse Event Without Identified Device or Use Problem1
  • Fracture1
  • Migration or Expulsion of Device1
  • Premature Activation1
  • Separation Failure1

Patient problems

  • No Consequences Or Impact To Patient3
  • Device Embedded In Tissue or Plaque2
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 30, 2024Separation Failure; Separation Failure

    IT WAS REPORTED THAT DURING THE PROCEDURE, THE PHYSICIAN ENCOUNTERED DIFFICULTY ADVANCING THE SUBJECT COIL WITHIN THE MICROCATHETER AND REQUIRED FORCE TO PUSH. ONCE THE SUBJECT COIL WAS VISIBLE THE DELIVERY WIRE OF THE SUBJECT COIL WAS PULLED BACK BUT THE SUBJECT COIL HAD DETACHED PREMATURELY IN THE MICROCATHETER. THE PHYSICIAN REPLACED IT WITH A NEW DEVICE AND CONTINUED THE PROCEDURE WITHOUT CLINICAL CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20110379 ↗
  • InjuryAug 31, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING AN ENDOVASCULAR PROCEDURE IN A PATIENT WITH INTERNAL CAROTID ARTERY ANEURYSM, THE SUBJECT COIL GOT STUCK IN THE MICROCATHETER, THE PHYSICIAN REMOVED THE SUBJECT COIL TOGETHER WITH THE MICROCATHETER OUT OF THE PATIENT VASCULATURE. THE PHYSICIAN DEPLOYED A FLOW DIVERTER STENT AND THEN ATTEMPTED TO DEPLOY COIL FROM POSTERIOR CIRCULATION. THERE WAS A SURGICAL DELAY OF 1 HOUR AND IT WAS REPORTED THAT THE PATIENT EXPERIENCED UNKNOWN ADVERSE CONSEQUENCES DUE TO THE SURGICAL DELAY. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 15332156 ↗
  • MalfunctionJul 24, 2020Premature Separation; Premature Separation

    ANALYSIS OF THE RETURNED DEVICE FOUND THAT THE COIL WAS PREMATURELY DETACHED/SEPARATED INSIDE PATIENT. THERE WERE NO CLINICAL CONSEQUENCES TO THE PATIENT AS A RESULT OF THIS EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10323796 ↗
  • MalfunctionFeb 5, 2020Premature Separation; Premature Separation

    IT WAS REPORTED THAT DURING THE PROCEDURE THE COIL (SUBJECT DEVICE) USED WAS RELEASED WITHIN THE MICROCATHETER PRIOR TO BEING IMPLANTED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AND NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9674708 ↗
  • MalfunctionSep 6, 2018Premature Separation; Premature Separation

    IT WAS REPORTED THAT THE COIL (SUBJECT DEVICE) PREMATURELY DETACHED AT THE JUNCTION WITHOUT ANY DETACHMENT ATTEMPT. HOWEVER, THE COIL WAS ENTIRELY PLACED INSIDE THE ANEURYSM. THERE WERE NO CLINICAL CONSEQUENCES TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7852605 ↗
  • InjuryAug 9, 2018Fracture; Premature Separation; Fracture; Premature Separation

    IT WAS REPORTED THAT DURING COIL EMBOLIZATION OF A BIG ANEURYSM, THE SUBJECT COIL DETACHED PREMATURELY AND A VERY TINY FILAMENT WAS OBSERVED. THE PHYSICIAN WAS NOT ABLE TO ADVANCE THE MICROCATHETER; THEREFORE, THE COIL WAS RETRIEVED WITH THE MICROCATHETER. THE FILAMENT WAS LEFT INSIDE THE PATIENT¿S ARTERY. NO OTHER PATIENT CONSEQUENCES REPORTED. NO OTHER ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 7768929 ↗
  • InjuryMar 14, 2017Migration or Expulsion of Device; Premature Activation; Migration or Expulsion of Device; Premature Activation

    DURING PLACEMENT OF THE FIRST COIL INTO A PULMONARY ARTERIOVENOUS FISTULA, AS THE FIRST COIL WAS PLACED, IT JAMMED AND STRETCHED. UNSUCCESSFULLY ATTEMPTS TO SNARE THE COIL WERE PERFORMED. THE STRETCHED PORTION OF THE FIRST COIL REMAINED IN THE PATIENT VESSEL. THE PHYSICIAN CONTINUED THE PROCEDURE BY PLACING A SECOND COIL (SUBJECT DEVICE) FOLLOWED BY UNSUCCESSFUL DETACHMENT ATTEMPTS. IT WAS REPORTED THAT WHEN SECOND COIL WAS BEING REMOVED, IT DETACHED PREMATURELY REMAINING UN-RETRIEVED INSIDE THE PULMONARY ARTERY. THE EVENT INDICATED THAT TO PREVENT THE COIL FROM MIGRATION IT WAS SECURED BY DEPLOYING 20 COILS TO OCCLUDE THE FISTULA WITHOUT ANY PROBLEM. PATIENT CURRENT CONDITION WAS REPORTED AS STABLE.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 6405002 ↗

Other DEVICE, NEUROVASCULAR EMBOLIZATION devices

All DEVICE, NEUROVASCULAR EMBOLIZATION devices →

Sources and updates