Skip to content

C-Taper Head 06-3299Recalls & reported problems

Howmedica Osteonics Corp. · Low Friction Ion Treatment · Class 2 · Implant

All models of C-Taper Head (18)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07613327016338), received between May 23, 2025 and May 23, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Difficult to Insert” (1 report).

3 other recalls cover devices cleared under the same FDA submission (K910988, K993601); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Insert1
  • Osseointegration Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 23, 2025Difficult to Insert; Osseointegration Problem; Difficult to Insert; Osseointegration Problem

    PRIMARY PROCEDURE, LEFT HIP. IT WAS REPORTED THAT AFTER IMPLANTATION, THE FEMORAL HEAD WAS LOOSE ON THE EON STEM (TO THE DEGREE WHERE VIRTUALLY NO FORCE WAS NEEDED TO MANUALLY REMOVE THE HEAD). HEAD WAS TESTED ON ALL THE NECK TRIALS IN THE TRAY, AND BEHAVED THE SAME WAY. A CERAMIC HEAD WAS OPENED AND LOCKED ON THE STEM WITH NO ISSUE. PROCEDURE WAS COMPLETE SUCCESSFULLY WITH A DELAY OF APPROXIMATELY 2-3 MINUTES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22078118 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K910988, K993601. One submission often covers many models, so check the product and lot details in each record.

  • Z-0390-2010 · Sep 4, 2009 · Terminated
    Stryker Howmedica Osteonics, Skirted 5 40' Taper V40 Femoral Head (26 mm +8mm offset), hip prosthesis component.
  • Z-0391-2010 · Sep 4, 2009 · Terminated
    Stryker Howmedica Osteonics, Skirtless 5 40' LFIT V40 Femoral Head (32 mm +0 mm offset); hip prosthesis component.
  • Z-1822-2009 · Jun 18, 2009 · Terminated
    V 40 Femoral Head, Sterile, 22 mm; Catalog number 6260-4-122, Catalog number 6260-5-226; Howmedica Osteonics Corp., Stryker Ireland, Carrigtwohill County Cork, Ireland; Intended f

Other Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented devices

All Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page