Skip to content

ABG 4845-0205Recalls & reported problems

Howmedica Osteonics Corp. · Femoral Stem - Left · Class 2 · Implant

All models of ABG (10)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07613327051032), received between Nov 23, 2020 and Mar 30, 2025 — classified as 4 injuries.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2
  • Corroded1
  • Device Dislodged or Dislocated1
  • Malposition of Device1

Patient problems

  • Ambulation or Postural Difficulties2
  • Fall1
  • Foreign Body Reaction1
  • Insufficient Information1
  • Pain1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 30, 2025Corroded; Device Dislodged or Dislocated; Corroded; Device Dislodged or Dislocated

    IT WAS REPORTED THROUGH THE FILING OF A LAWSUIT THAT ALLEGEDLY ON OR ABOUT (B)(6) 2011, THE PATIENT WAS IMPLANTED WITH AN LFIT ANATOMIC COCR V40 FEMORAL HEAD ON HIS LEFT HIP AND WAS REVISED ON (B)(6) 2024. IT IS FURTHER ALLEGED THAT THE TAPER HAD COMPLETELY CORRODED AND THE FEMORAL HEAD HAD FALLEN OFF INTO THE ACETABULUM.

    Patient problems: Ambulation or Postural Difficulties

    MAUDE report 21728707 ↗
  • InjuryOct 4, 2021Fracture; Fracture

    IT HAS BEEN REPORTED THAT: "(B)(6) PATIENT WITH A LEFT TOTAL HIP PROSTHESIS IN 2008. RUPTURE OF THE LEFT TOTAL HIP PROSTHESIS AT THE NECK OF THE STEM, FOLLOWING AN IMPACT WITH A DOOR AND A FALL WITH LANDING ON THE LEFT HIP. PAIN IN THE LEFT HIP WITH FUNCTIONAL IMPOTENCE MIG. ACTION TAKEN: ORTHOPAEDIC SURGERY TO FIT NEW PROSTHETIC MATERIAL."

    Patient problems: Fall; Pain

    MAUDE report 12569339 ↗
  • InjuryJun 24, 2021Malposition of Device; Malposition of Device

    AS REPORTED IN MEDWATCH MW5101277: "I AM THE PATIENT RELATED TO THE FDA REPORT: STRYKER ORTHOPAEDICS-(B)(4) ABG II NO. 5 CEMENTLESS LEFT V40 CATALOG NUMBER 4845-0205 EVENT DATE (B)(6) 2013 MDR REPORT KEY (B)(4). THIS IS THE PROSTHESIS THAT WAS PUT ON ME: STRYKER ORTHOPEDICS-(B)(4) ABG II NO 5 CEMENTLESS LEFT V40. AFTERWARDS, TWO CT SCANS, ONE PERFORMED ON (B)(6) 2013 AND THE OTHER - RESOLUTIVE - ON (B)(6) 2013, SHOWED THAT THE PROSTHESIS WAS BADLY POSITIONED AND FOR THIS REASON AN IMPINGEMENT HAD BEEN CAUSED BETWEEN THE STEM AND THE ACETABULAR CUP (INCLUDING POLYETHYLENE INSERT). THE ACETABULAR CUP HAD BEEN IMPLANTED WITH POOR ANTEVERSION: THE ANTEVERSION ANGLE OF 20° WAS NOT RESPECTED AND IT WAS ONLY 7°. THIS FACT HAD INJURED THE ILIOPSOAS TENDON, WHICH HAS NEVER RECOVERED, EVEN AFTER THE REMOVAL AND REPLACEMENT OF THE PROSTHESIS. THE TESTS TO WHICH STRYKER REFERS IN THE REPORT OF THE CASE, WERE ONLY NICKEL AND CHROMIUM BECAUSE STRYKER ON (B)(6) 2013 HAD NOT YET PROVIDED THE DOCTOR WHO HAD MADE THE REPORT, THE TECHNICAL DATA SHEET OF THE PROSTHESIS, WHERE ALL THE METALS TO BE SEARCHED WERE INDICATED. ACTUALLY, I HAVE THE EMAIL IN WHICH STRYKER CONFIRMS THE OPENING OF THE ACCIDENT

    Patient problems: Foreign Body Reaction

    MAUDE report 12061525 ↗
  • InjuryNov 23, 2020Fracture; Fracture

    DR. REPORT: "I WOULD LIKE TO INFORM YOU OF THE WEAR (BREAKAGE) OF THE FEMORAL STEM NECK OF THE ABG PROSTHESIS. CODE: HOW / OST 4845-0205 ABG II FEMORAL STEM LEFT. THE PATIENT HAD TO UNDERGO A PROSTHETIC REPLACEMENT SURGERY - ON (B)(6) 2020"

    Patient problems: Ambulation or Postural Difficulties; Insufficient Information

    MAUDE report 10883613 ↗

Other Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented devices

All Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented devices →