Skip to content

Precision Ideal Eyes 0502503030Recalls & reported problems

Stryker Corporation · HD Laparoscope, Autoclavable, 30 DEG View Direction · Class 2

All models of Precision Ideal Eyes (10)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Stryker Precision Ideal Eyes HD laparoscope is a reusable scope for general and plastic surgery. The 56 reports mostly describe image problems — blurry or fogged views during surgery — plus a few cases where parts of the scope broke or chipped, sometimes leaving a small fragment that was retrieved. Reports describe no deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
59
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 59 adverse event reports that list this exact device (DI 07613327059144), received between Sep 11, 2019 and Sep 30, 2026 — classified as 59 malfunctions.

The most frequently reported specific device problem is “Poor Quality Image” (33 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 1 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Poor Quality Image33
  • Optical Problem19
  • Fogging11
  • Break5
  • Nonstandard Device5
  • Material Fragmentation4
  • Moisture or Humidity Problem3
  • Material Integrity Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions52
  • Insufficient Information22
  • No Known Impact Or Consequence To Patient7
  • Abrasion1
  • Burn(s)1

Recent adverse event reports

The 25 most recent of 59 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Fogging

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50952911 ↗
  • MalfunctionSep 29, 2026Fogging

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50940525 ↗
  • MalfunctionAug 26, 2026Fogging

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50462049 ↗
  • MalfunctionJul 15, 2026Fogging

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25833916 ↗
  • MalfunctionJan 8, 2026Material Fragmentation

    IT WAS REPORTED THAT PIECE OF THE SCOPE IS BROKEN OFF. THEY REFERENCED THE DEVICE AS BROKEN AND ONLY A VERY SMALL PORTION OF THE DISTAL TIP WAS CHIPPED AND DETACHED INTO THE TROCAR. THEY BELIEVE THAT THE SCOPE MAY HAVE BEEN DROPPED DURING TRANSPORT. THIS ISSUE WAS IDENTIFIED AT ENTRY. THE DETACHED PIECE WAS SUCCESSFULLY RETRIEVED, AND A REPLACEMENT SCOPE WAS USED TO COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24004908 ↗
  • MalfunctionNov 21, 2025Fogging; Fogging

    IT WAS REPORTED THAT THERE WAS FOGGING ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23619843 ↗
  • MalfunctionAug 20, 2025Fogging; Optical Problem; Fogging; Optical Problem

    IT WAS REPORTED THAT THERE WAS FOGGY IMAGE DURING SURGERY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22854880 ↗
  • MalfunctionApr 22, 2025Fogging; Moisture or Humidity Problem; Fogging; Moisture or Humidity Problem

    IT WAS REPORTED THAT THERE WAS FOGGY IMAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21886356 ↗
  • MalfunctionNov 8, 2024Fogging; Moisture or Humidity Problem; Fogging; Moisture or Humidity Problem

    IT WAS REPORTED THAT THERE WAS FOGGY IMAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20645586 ↗
  • MalfunctionNov 8, 2024Fogging; Moisture or Humidity Problem; Fogging; Moisture or Humidity Problem

    IT WAS REPORTED THAT THERE WAS FOGGY IMAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20644931 ↗
  • MalfunctionOct 24, 2024Break; Material Fragmentation; Break; Material Fragmentation

    IT WAS REPORTED THAT THE DEVICE BROKE. PLEASE NOTE THAT THE BROKEN PIECE WAS SUCCESSFULLY REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20522426 ↗
  • MalfunctionJan 17, 2024Fracture; Detachment of Device or Device Component; Fracture; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE DEVICE BROKE. PLEASE NOTE THAT THE BROKEN PIECE WAS SUCCESSFULLY REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18531188 ↗
  • MalfunctionOct 12, 2023Fogging; Fogging

    IT WAS REPORTED THAT THERE WAS FOGGY IMAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17926211 ↗
  • MalfunctionSep 6, 2023Material Fragmentation; Material Integrity Problem; Material Fragmentation; Material Integrity Problem

    IT WAS REPORTED THAT THE TIP OF THE SCOPE BROKE. THE BROKEN PIECE WAS SUCCESSFULLY REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17694973 ↗
  • MalfunctionSep 6, 2023Material Fragmentation; Material Integrity Problem; Material Fragmentation; Material Integrity Problem

    IT WAS REPORTED THAT THE TIP OF THE SCOPE BROKE. THE BROKEN PIECE WAS SUCCESSFULLY REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17694972 ↗
  • MalfunctionAug 15, 2023Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THAT THERE WAS BLURRY IMAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17553149 ↗
  • MalfunctionAug 9, 2023Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THAT THERE WAS BLURRY IMAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17493739 ↗
  • MalfunctionMay 25, 2023Optical Problem; Optical Problem

    IT WAS REPORTED THAT THERE WAS BLURRY IMAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17007932 ↗
  • MalfunctionApr 27, 2023Sharp Edges; Sharp Edges

    IT WAS REPORTED THAT THE PATIENT WAS SCRATCHED BY THE SHARP DISTAL TIP.

    Patient problems: Abrasion

    MAUDE report 16827490 ↗
  • MalfunctionApr 21, 2023Optical Problem; Optical Problem

    IT WAS REPORTED THAT THERE WAS BLURRY IMAGE DURING PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16792209 ↗
  • MalfunctionMar 13, 2023Poor Quality Image

    IT WAS REPORTED THAT THERE WAS BLURRY IMAGE DURING PROCEDURE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16527416 ↗
  • MalfunctionJan 30, 2023Fogging; Poor Quality Image; Fogging

    IT WAS REPORTED THAT THERE WAS FOGGY IMAGE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16264711 ↗
  • MalfunctionJan 5, 2023Poor Quality Image

    IT WAS REPORTED THAT THERE WAS BLURRY IMAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16104489 ↗
  • MalfunctionDec 15, 2022Optical Problem; Optical Problem

    IT WAS REPORTED THAT THERE WAS BLURRY IMAGE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 15989080 ↗
  • MalfunctionNov 23, 2022Poor Quality Image

    IT WAS REPORTED THAT THERE WAS BLURRY IMAGE DURING PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15858930 ↗

Other Laparoscope, general & plastic surgery devices

All Laparoscope, general & plastic surgery devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page