Pneumosure 0620040600Recalls & reported problems
Stryker Corporation · High Flow Insufflator · Class 2
All models of Pneumosure (5)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 07613327063219), received between Feb 8, 2019 and Jun 5, 2025 — classified as 2 injuries, 14 malfunctions.
The most frequently reported specific device problem is “Pressure Problem” (9 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Pressure Problem9
- Increase in Pressure5
- Therapeutic or Diagnostic Output Failure4
- Inflation Problem3
- Mechanical Problem2
- No Flow2
- Output Problem2
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions13
- Insufficient Information7
- No Known Impact Or Consequence To Patient2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Chemical Exposure1
- Electrolyte Imbalance1
- Low Oxygen Saturation1
- Peritoneal Laceration(s)1
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 5, 2025Image Display Error/Artifact; Decrease in Pressure; Increase in Pressure; Image Display Error/Artifact; Decrease in Pressure; Increase in Pressure
IT WAS REPORTED THAT DURING THE PROCEDURE THE PNEUMOSURE PRESSURE ROSE ABOVE THE SET PRESSURE BUT INSTEAD OF THE INSUFFLATOR GIVING AN ¿OVERPRESSURE¿ WARNING MESSAGE OR MAKING ANY AUDITORY BEEPING OR ANY SOUND, THE INSUFFLATION STOPPED. AT THAT POINT PRESSURE BEGAN DECREASING AND FLOW WAS AT 0. TO BEGIN INSUFFLATING AGAIN THE ¿START¿ BUTTON HAD TO BE PRESSED ON THE INSUFFLATOR ITSELF. THEREFORE, THERE WAS OVERPRESSURE AND LOSS OF INSUFFLATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22152888 ↗MalfunctionApr 17, 2025Increase in Pressure; Pressure Problem; Increase in Pressure; Pressure Problem
IT WAS REPORTED THAT DURING A CASE THE INSUFFLATOR WAS NOT GIVING AN ACCURATE PRESSURE OR FLOW READING THAT CAUSED OVER DISTENSION OF THE PATIENT. THEY HAD TO SHUT THE INSUFFLATOR OFF AND START AN OPEN CASE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21860549 ↗MalfunctionMar 7, 2024No Flow; Output Problem; No Flow; Output Problem
IT WAS REPORTED THERE WAS LOSS OF INSUFFLATION DURING PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18862106 ↗InjuryApr 21, 2023No Flow; Pressure Problem; No Flow; Pressure Problem
IT WAS REPORTED THAT THE CASE WAS DELAYED A FEW DAYS DUE TO LOSS OF INSUFFLATION DURING PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions; Peritoneal Laceration(s)
MAUDE report 16792208 ↗MalfunctionJun 30, 2022No Pressure; Output Problem; No Pressure
IT WAS REPORTED THAT THERE WAS LOSS OF INSUFFLATION DURING PROCEDURE.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 14878685 ↗MalfunctionMay 27, 2022Pressure Problem; Therapeutic or Diagnostic Output Failure; Pressure Problem
IT WAS REPORTED THAT THERE WAS LOSS OF INSUFFLATION DURING PROCEDURE.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 14522242 ↗MalfunctionMay 27, 2022Pressure Problem; Therapeutic or Diagnostic Output Failure; Pressure Problem
IT WAS REPORTED THAT THERE WAS LOSS OF INSUFFLATION DURING PROCEDURE.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 14522241 ↗MalfunctionMay 27, 2022Pressure Problem; Therapeutic or Diagnostic Output Failure; Pressure Problem
IT WAS REPORTED THAT THERE WAS LOSS OF INSUFFLATION DURING PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14522240 ↗MalfunctionMay 26, 2022Pressure Problem; Therapeutic or Diagnostic Output Failure; Pressure Problem
IT WAS REPORTED THAT THERE WAS LOSS OF INSUFFLATION DURING PROCEDURE.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 14516438 ↗MalfunctionApr 28, 2022Increase in Pressure; Pressure Problem; Pressure Problem
IT WAS REPORTED THAT PATIENT HAD SUFFERED A COLLAPSED LUNG DUE TO OVERPRESSURE.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Pneumothorax
MAUDE report 14240411 ↗MalfunctionJan 18, 2022Pressure Problem; Increase in Pressure; Pressure Problem
IT WAS REPORTED THAT THERE WAS OVERPRESSURE DURING PROCEDURE.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 13282742 ↗InjuryOct 14, 2021Increase in Pressure; Increase in Pressure; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED EXCESSIVE CO2 LEVELS.
Patient problems: Chemical Exposure; Electrolyte Imbalance; Low Oxygen Saturation
MAUDE report 12636594 ↗MalfunctionMay 26, 2021Inflation Problem; Inflation Problem; Mechanical Problem
IT WAS REPORTED THAT THERE WAS LOSS OF INSUFFLATION DURING THE PROCEDURE.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 11882364 ↗MalfunctionMay 25, 2021Inflation Problem; Inflation Problem; Mechanical Problem
IT WAS REPORTED THAT THERE WAS LOSS OF INSUFFLATION DURING THE PROCEDURE.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 11882319 ↗MalfunctionOct 13, 2020Calibration Problem; Calibration Problem; Pressure Problem
IT WAS REPORTED THAT THERE WAS INLET GAS PRESSURE ISSUE DURING THE PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10673365 ↗MalfunctionFeb 8, 2019Device Alarm System; Inflation Problem
IT WAS REPORTED THAT THE DEVICE WAS OVER INFLATING WITH NO ALARMS. PLEASE NOTE, THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8321864 ↗
Other INSUFFLATOR, LAPAROSCOPIC devices
- HIGH FLOW INSUFFLATION UNIT UHI-3Olympus Medical Systems Corp. · INSUFFLATOR, LAPAROSCOPIC1 recall162 reports
- HIGH FLOW INSUFFLATION UNIT UHI-3Olympus Medical Systems Corp. · INSUFFLATOR, LAPAROSCOPIC1 recall48 reports
- AIRSEAL AS-iFS1Conmed Corporation · Insufflator, laparoscopic4 deaths93 reports
- Pneumosure 0620040610Stryker Corporation · INSUFFLATOR, LAPAROSCOPIC1 death79 reports
- Pneumosure 0620040690Stryker Corporation · INSUFFLATOR, LAPAROSCOPIC1 recall22 reports
- Pneumosure 0620040660Stryker Corporation · INSUFFLATOR, LAPAROSCOPIC1 recall18 reports
- CYLINDER HOSE (PIN) FOR UHI-3 MAJ-1080Olympus Medical Systems Corp. · INSUFFLATOR, LAPAROSCOPIC58 reports
- Pneumosure 0620040680Stryker Corporation · INSUFFLATOR, LAPAROSCOPIC1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page