Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613327072365), received between Jul 24, 2019 and Jul 20, 2020 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Device Difficult to Setup or Prepare” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Device Difficult to Setup or Prepare1
  • Unstable1

Patient problems

  • Injury1
  • Joint Dislocation1
  • No Consequences Or Impact To Patient1
  • Pain1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 20, 2020Unstable; Unstable

    IT WAS REPORTED THAT THE PATIENT'S RIGHT TOTAL HIP WAS REVISED DUE TO INSTABILITY.

    Patient problems: Injury; Joint Dislocation; Pain

    MAUDE report 10294566 ↗
  • MalfunctionJul 24, 2019Device Difficult to Setup or Prepare; Device Difficult to Setup or Prepare

    AS REPORTED: "SURGEON TRIED TO IMPLANT STEM AND IMPACT HEAD ONTO STEM. C-TAPER METAL HEAD DID NOT ENGAGE ONTO THE TRUNNION OF THE C-TAPER HFX STEM. HEAD JUST SPUN ON TRUNNION". A SURGICAL DELAY OF 5 MINUTES WAS REPORTED. A USAGE SHEET PROVIDED BY THE REP SHOWS A 36 +0 C-TAPER HEAD WITH "NO CHARGE", AND A 36 +2.5 C-TAPER HEAD WITH NO NOTE PLACED AFTERWARD.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8821679 ↗

Other Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented devices

All Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented devices →