OMEGA 704005Recalls & reported problems
Stryker GmbH · Combination Reamer Assembly · Class 1
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 07613327079111), received between Aug 22, 2019 and May 17, 2021 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Mechanics Altered” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Mechanics Altered5
- Loose or Intermittent Connection2
- Defective Device1
- Use of Device Problem1
Patient problems
- No Known Impact Or Consequence To Patient7
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 17, 2021Mechanics Altered; Mechanics Altered
THE CUSTOMER REPORTED THAT THE TRIPLE BARRELED REAMER WHICH SHOULD FIX IN SITU BECAME LOSE WHILE REAMING AND THE WIRE IT WAS REAMING OVER BECAME STUCK AND HAD TO BE REMOVED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11836388 ↗MalfunctionMay 27, 2020Mechanics Altered; Mechanics Altered
THE CUSTOMER REPORTED THAT THE LOCKING MECHANISM THAT CONTROLS THE DEPTH OF THE DHS REAMER IS FAILING. NO ADVERSE EVENTS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10092904 ↗MalfunctionMay 27, 2020Mechanics Altered; Mechanics Altered
THE CUSTOMER REPORTED THAT THE LOCKING MECHANISM THAT CONTROLS THE DEPTH OF THE DHS REAMER IS FAILING. NO ADVERSE EVENTS WERE RECORDED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10092899 ↗MalfunctionMay 27, 2020Mechanics Altered; Mechanics Altered
THE CUSTOMER REPORTED THAT THE LOCKING MECHANISM THAT CONTROLS THE DEPTH OF THE DHS REAMER IS FAILING. NO ADVERSE EVENTS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10092891 ↗MalfunctionMay 27, 2020Mechanics Altered; Mechanics Altered
THE CUSTOMER REPORTED THAT THE LOCKING MECHANISM THAT CONTROLS THE DEPTH OF THE DHS REAMER IS FAILING.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10092889 ↗MalfunctionAug 23, 2019Loose or Intermittent Connection
THE CUSTOMER REPORTED THAT THEY SET THE DEVICE TO A SPECIFIC LENGTH AND WHEN THEY WE USING IT IN THE CASE IS CONTINUED TO ADVANCE. THEY STOPPED DRILLING. THERE WAS A DELAY OF 5 MINUTES WHILE THEY CHECKED IF THE DEVICE WAS PROPERLY LOCKED. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. NO ADVERSE EVENTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8927317 ↗MalfunctionAug 23, 2019Loose or Intermittent Connection
THE CUSTOMER REPORTED THAT THEY SET THE DEVICE TO A SPECIFIC LENGTH AND WHEN THEY WE USING IT IN THE CASE IS CONTINUED TO ADVANCE. THEY STOPPED DRILLING. THERE WAS A DELAY OF 5 MINUTES WHILE THEY CHECKED IF THE DEVICE WAS PROPERLY LOCKED. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. NO ADVERSE EVENTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8927314 ↗MalfunctionAug 22, 2019Defective Device
THE CUSTOMER REPORTED THAT THEY SET THE DEVICE TO A SPECIFIC LENGTH AND WHEN THEY WERE USING IT IN THE CASE IT CONTINUED TO ADVANCE. THEY STOPPED DRILLING. THERE WAS A DELAY OF 5 MINUTES WHILE THEY CHECKED IF THE DEVICE WAS PROPERLY LOCKED. NO ADVERSE EVENTS. THIS DEVICE WAS USED TO COMPLETE THE CASE. THE CUSTOMER COULD NOT REMEMBER WHICH SET WAS BEING USED AT THE TIME (2 POSSIBLE ITEMS).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8923006 ↗MalfunctionAug 22, 2019Use of Device Problem
THE CUSTOMER REPORTED THAT THEY SET THE DEVICE TO A SPECIFIC LENGTH AND WHEN THEY WERE USING IT IN THE CASE IT CONTINUED TO ADVANCE. THEY STOPPED DRILLING. THERE WAS A DELAY OF 5 MINUTES WHILE THEY CHECKED IF THE DEVICE WAS PROPERLY LOCKED. NO ADVERSE EVENTS. THIS DEVICE WAS USED TO COMPLETE THE CASE. THE CUSTOMER COULD NOT REMEMBER WHICH SET WAS BEING USED AT THE TIME (2 POSSIBLE ITEMS).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8922846 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page