Skip to content

VARIAX 628516SRecalls & reported problems

Stryker GmbH · Hook Plate · Class 2 · Implant

All models of VARIAX (345)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613327085525), received between Jun 4, 2019 and Sep 30, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Defective Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Defective Device1
  • Device Dislodged or Dislocated1

Patient problems

  • Erythema1
  • No Known Impact Or Consequence To Patient1
  • Skin Inflammation/ Irritation1
  • Swelling/ Edema1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjurySep 30, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    AS REPORTED: "STIMULATION OF SOFT TISSUE DUE TO LOCKING SCREW BACKOUT. THE LOCKING SCREW WAS REMOVED." 09/06/2025 - ADDITIONAL INFORMATION RECEIVED: "THE SCREW AT THE MOST DISTAL END OF THE HOOK PLATE WAS ALSO LOOSE, BUT THAT WAS COMPLETELY LOCKED WHEN IT WAS RE-TIGHTENED, SO IT WAS RE-TIGHTENED. THE BACKED-OUT SCREW CAUSED IRRITATION, AND THE SKIN ON THE CLAVICLE SURFACE WAS RED AND SWOLLEN."

    Patient problems: Erythema; Skin Inflammation/ Irritation; Swelling/ Edema

    MAUDE report 23184987 ↗
  • MalfunctionJun 4, 2019Defective Device

    IT WAS REPORTED THAT VARIAX CLAVICLE SURGERY WAS PERFORMED ON 5/7/2019. WHEN THE SURGEON TRIED TO LOCK THE LOCKING SCREW TO THE SECOND HOLE FROM THE DISTAL END OF THE PLATE, IT DID NOT LOCK. THE SURGEON DECIDED TO LEAVE THE SCREW IN A STATE THAT IS NOT LOCKED TO THE PLATE. THE PROCEDURE WAS OTHERWISE COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY OR ADVERSE CONSEQUENCES REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8668134 ↗

Other PLATE, FIXATION, BONE devices

All PLATE, FIXATION, BONE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page