VARIAX 657310Recalls & reported problems

Stryker GmbH · Locking Screw · Class 2 · Implant

All models of VARIAX (345)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 07613327086782), received between Sep 19, 2018 and Aug 4, 2025 — classified as 1 injury, 6 malfunctions.

The most frequently reported specific device problem is “Positioning Failure” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Positioning Failure5
  • Appropriate Device Problem Term/Code Not Available1
  • Biocompatibility1

Patient problems

  • No Known Impact Or Consequence To Patient6
  • No Clinical Signs, Symptoms or Conditions4
  • Erosion1
  • Foreign Body Reaction1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 4, 2025Biocompatibility; Biocompatibility

    AS REPORTED: "SURGEON ADVISED HE WAS REMOVING ONE OF OUR VARIAX FIBULA PLATES THAT HE HAD IMPLANTED 5 YEARS PRIOR & AT THE PLATE BONE INTERFACE AS PER ATTACHED PHOTO THERE WAS "GROSS METALLOSIS" AND "PRETTY BAD BONE EROSION" PLATE AND SCREWS HAVE BEEN SUCCESSFULLY REMOVED & WILL BE RETURNED ONCE DECONTAMINATED. SWABS HAVE BEEN SENT TO LOCAL PATHOLOGY."

    Patient problems: Erosion; Foreign Body Reaction

    MAUDE report 22699857 ↗
  • MalfunctionJul 15, 2020Positioning Failure; Positioning Failure

    IT WAS REPORTED: THERE WAS NO LOCKING IN THE PLATE WITH SEVERAL VARIAX2 LOCKING SCREWS 10MM 3.5 T10. THE PROCEDURE WAS SUCCESSFULLY COMPLETED BY REMOVING THE AFFECTED SCREWS AND USING VARIAX1 SCREWS INSTEAD. SURGICAL DELAY OF 20MIN. NOT LONGER.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10278902 ↗
  • MalfunctionJul 15, 2020Positioning Failure; Positioning Failure

    IT WAS REPORTED: THERE WAS NO LOCKING IN THE PLATE WITH SEVERAL VARIAX2 LOCKING SCREWS 10MM 3.5 T10. THE PROCEDURE WAS SUCCESSFULLY COMPLETED BY REMOVING THE AFFECTED SCREWS AND USING VARIAX1 SCREWS INSTEAD. SURGICAL DELAY OF 20MIN. NOT LONGER.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10278302 ↗
  • MalfunctionJul 15, 2020Positioning Failure; Positioning Failure

    IT WAS REPORTED: THERE WAS NO LOCKING IN THE PLATE WITH SEVERAL VARIAX2 LOCKING SCREWS 10MM 3.5 T10. THE PROCEDURE WAS SUCCESSFULLY COMPLETED BY REMOVING THE AFFECTED SCREWS AND USING VARIAX1 SCREWS INSTEAD. SURGICAL DELAY OF 20MIN. NOT LONGER.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10278275 ↗
  • MalfunctionJul 15, 2020Positioning Failure; Positioning Failure

    IT WAS REPORTED: THERE WAS NO LOCKING IN THE PLATE WITH SEVERAL VARIAX2 LOCKING SCREWS 10MM 3.5 T10. THE PROCEDURE WAS SUCCESSFULLY COMPLETED BY REMOVING THE AFFECTED SCREWS AND USING VARIAX1 SCREWS INSTEAD. SURGICAL DELAY OF 20MIN. NOT LONGER.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10278268 ↗
  • MalfunctionJul 15, 2020Positioning Failure; Positioning Failure

    IT WAS REPORTED: THERE WAS NO LOCKING IN THE PLATE WITH SEVERAL VARIAX2 LOCKING SCREWS 10MM 3.5 T10. THE PROCEDURE WAS SUCCESSFULLY COMPLETED BY REMOVING THE AFFECTED SCREWS AND USING VARIAX1 SCREWS INSTEAD. SURGICAL DELAY OF 20MIN. NOT LONGER.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10278222 ↗
  • MalfunctionSep 19, 2018Appropriate Device Problem Term/Code Not Available

    PRIMARY PROCEDURE, RIGHT CLAVICLE. IT WAS REPORTED THAT IN IMPLANTING THE VARIAX2 CORE 3.5 LOCKING SCREWS INTO THE CLAVICULAR PLATE, THE SCREWS WOULD NOT LOCK - THEY KEPT SPINNING WHEN THEY SHOULD HAVE LOCKED. REP OBTAINED LOCKING SCREWS FROM A VARIAX 1 ANKLE SOLUTION SET, WHICH LOCKED AS EXPECTED. THE PLATE IS IMPLANTED AND THE REPORTED SCREWS ARE AVAILABLE FOR RETURN. REP REPORTED THAT SURGEON WAS FOLLOWING THE SURGICAL TECHNIQUE THROUGHOUT THE PROCEDURE. THERE WAS A SURGICAL DELAY OF 5 MINUTES.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7891485 ↗

Other Screw, fixation, bone devices