Skip to content

VARIAX 657310SRecalls & reported problems

Stryker GmbH · Locking Screw · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 07613327086799), received between Jul 22, 2020 and Feb 19, 2025 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Positioning Failure” (10 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Positioning Failure10
  • Material Fragmentation2

Patient problems

  • No Clinical Signs, Symptoms or Conditions10
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 19, 2025Positioning Failure; Positioning Failure

    AS REPORTED: "THE LOCKING SCREW ON THE VARIAX FIBULA PLATE FAILED TO LOCK. AFTER OPENING ANOTHER SCREW, THE SURGERY WAS COMPLETED WITHOUT ANY PROBLEMS".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21417486 ↗
  • MalfunctionJul 17, 2024Positioning Failure; Positioning Failure

    AS REPORTED: "LOCKING SCREW DID NOT LOCK." 7/4/2024 - ADDITIONAL INFORMATION RECEIVED. 3 SCREWS WERE SUCCESSFULLY IMPLANTED. LOCKING BAD SCREW IS NOT PUT IN, AND IT IS FINISHED AS IT IS. THE SURGEON IS RELATIVELY SATISFIED (BECAUSE IT IS USED IN CONJUNCTION WITH A TENSION BAND).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19767452 ↗
  • MalfunctionMar 21, 2023Positioning Failure; Material Fragmentation; Positioning Failure; Material Fragmentation

    IT WAS REPORTED: "THE LOCKING SCREW DID NOT LOCK TO THE PLATE. A THIN METAL PIECE WAS FOUND DURING THE SCREW INSERTION. THE SCREW WAS REPLACED AND THE NEW SCREW WAS ABLE TO LOCK TO THE PLATE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16587936 ↗
  • MalfunctionMar 20, 2023Positioning Failure; Positioning Failure

    AS REPORTED: "THE LOCKING SCREW DID NOT LOCK TO THE PLATE. THE SCREW AND THE PLATE WERE REPLACED".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16577564 ↗
  • MalfunctionNov 16, 2022Positioning Failure; Positioning Failure

    AS REPORTED: "THE LOCKING SCREW DID NOT LOCK TO THE PLATE AFTER DRILLING WITH 2.6MM DRILL. ANOTHER LOT NUMBER SCREW WITH THE SAME SIZE LOCKED AGAINST THE SAME HOLE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15804089 ↗
  • MalfunctionJun 28, 2022Positioning Failure; Positioning Failure

    AS REPORTED: "THE LOCKING SCREW DID NOT LOCK TO THE PLATE. SPARE SCREW WAS ABLE TO LOCK."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14840540 ↗
  • MalfunctionJan 17, 2022Positioning Failure; Positioning Failure

    AS REPORTED: "THE SUBJECT SCREW DID NOT LOCK TO THE FIBULA PLATE. THE SURGEON EXCHANGED IT TO A NEW ONE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13266548 ↗
  • MalfunctionJan 4, 2022Positioning Failure; Positioning Failure

    IT WAS REPORTED THAT INTRAOPERATIVELY, THE LOCKING DID NOT LOCK, SO IT WAS REPLACED WITH A NEW SCREW.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13158084 ↗
  • MalfunctionFeb 24, 2021Positioning Failure; Positioning Failure

    IT WAS REPORTED: "DURING THE INITIAL ORIF PROCEDURE FOR FIBULA (RIGHT SIDE), WHEN THE SURGEON ATTEMPTED TO INSERT THE 10MM LOCKING SCREW, IT DIDN'T FEEL LIKE IT WAS SPINNING OR TIGHTENING. THE REPORTING SR CONFIRMED THAT ANGLE OF THE DRILLING WAS LESS THAN 15 DEGREE, AND TRIED TO DRILL IT AGAIN. BUT IT DID NOT LOCK. THEREAFTER, WHEN THE SURGEON EXCHANGED WITH A NEW SCREW, IT LOCKED WITHOUT PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11376332 ↗
  • MalfunctionJul 22, 2020Positioning Failure; Material Fragmentation; Positioning Failure; Material Fragmentation

    IT WAS REPORTED: WHEN THE LOCKING SCREW WAS INSERTED INTO THE PLATE, THE SCREW THREAD WAS SCRAPED AND IT COULD NOT BE LOCKED. COLLECT THE SCREW AND DEBRIS. THE SURGEON INSERTED A NEW SCREW AND FINISHED THE OPERATION SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10311168 ↗

Other PLATE, FIXATION, BONE devices