VARIAX 657314SRecalls & reported problems
Stryker GmbH · Locking Screw · Class 2 · Implant
All models of VARIAX (345)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Stryker Variax locking screw is used to hold a bone plate in place during surgery. Most of the 31 reports describe the screw not locking into the plate, screws backing out or shifting after surgery, and metal fragments or thread damage appearing during insertion. Seven reports involve injuries and no deaths were reported. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 31
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 31 adverse event reports that list this exact device (DI 07613327086850), received between Sep 27, 2019 and Mar 25, 2026 — classified as 7 injuries, 24 malfunctions.
The most frequently reported specific device problem is “Positioning Failure” (12 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Positioning Failure12
- Material Fragmentation10
- Migration4
- Positioning Problem4
- Device Dislodged or Dislocated3
- Break2
- Difficult to Insert2
- Material Deformation2
Patient problems
- No Clinical Signs, Symptoms or Conditions19
- Insufficient Information5
- Failure of Implant3
- Foreign Body In Patient3
- Inflammation3
- No Consequences Or Impact To Patient3
- No Known Impact Or Consequence To Patient3
- Swelling/ Edema3
Recent adverse event reports
The 25 most recent of 31 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 25, 2026Positioning Failure
AS REPORTED: "PLATE AND LOCKING SCREW LOCKING FAILURE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24691599 ↗MalfunctionNov 10, 2025Positioning Failure; Positioning Failure
AS REPORTED: "VARIAX SUPERIOR LATERAL PLATE WAS USED AND THE LOCKING SCREW WAS NOT LOCKING IN THE DISTAL HOLE. (2 LOCATIONS)".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23509762 ↗InjuryJul 1, 2025Migration; Migration
AS REPORTED: "PRIMARY SURGERY WAS PERFORMED ON (B)(6) 2025. TWO MONTHS AFTER SURGERY, AN X-RAY REVEALED THAT THE 2 LOCKING SCREWS INSERTED ON THE OUTSIDE PLATE HAD BACKED OUT OF THE PLATE AND THE LOCK HAD COME LOOSE. THE SCREWS INSERTED INTO THE FIRST AND THIRD HOLES FROM THE OUTER SIDE OF THE PLATE BACKED OUT. IT IS UNCLEAR WHICH OF THE SCREWS ACTUALLY BACKED OUT. THE PATIENT IS ORIGINALLY A FARMER AND, ACCORDING TO REPORTS, RESUMED WORK THREE WEEKS AFTER SURGERY WITHOUT FOLLOWING THE DOCTOR'S INSTRUCTIONS. REVISION SURGERY IS SCHEDULED ON (B)(6) 2025"
Patient problems: Insufficient Information
MAUDE report 22378037 ↗MalfunctionFeb 21, 2025Material Fragmentation; Material Fragmentation
AS REPORTED: "WHEN THE LOCKING SCREW WAS INSERTED AS USUAL, METAL FRAGMENTS CAME OUT DURING LOCKING. JUST TO BE SURE, THE SCREW WAS REPLACED AND THE PROBLEM WAS RESOLVED".
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 21439700 ↗MalfunctionJan 9, 2025Positioning Failure; Material Fragmentation; Device Difficult to Setup or Prepare; Material Deformation; Positioning Failure; Material Fragmentation; Device Difficult to Setup or Prepare; Material Deformation
IT WAS REPORTED: DURING SCREW INSERTION, THE SCREW HEAD THREAD WAS SHAVED DUE TO AN INSERTION ANGLE PROBLEM, RESULTING IN POOR LOCKING, AND ANOTHER SCREW WAS OPENED AND USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21120459 ↗MalfunctionDec 4, 2024Positioning Failure; Positioning Failure
AS REPORTED: "THE 3.5 MM DIAMETER 14 MM LOCKING SCREW OF THE VARIAXS CLAVICLE REPORTED DID NOT LOCK, SO IT WAS REPLACED WITH ANOTHER SCREW AND NEW SCREW LOCKED WITHOUT PROBLEM. THE SUBJECT SCREW DID NOT LOCK IN THE OTHER HOLE THOUGH THE ANGLE INSERTED WAS WITHIN 15 DEGREES".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20849753 ↗MalfunctionOct 21, 2024Positioning Failure; Positioning Failure
THE LOCKING SCREW COULD NOT BE LOCKED TO THE PLATE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20494931 ↗MalfunctionSep 25, 2024Positioning Problem; Positioning Problem
IT WAS REPORTED, WHEN THE LOCKING SCREW WAS USED, THE LOCKING SCREW WAS NOT ENGAGED IN THE LIP OF THE PLATE AND COULD NOT BE LOCKED. A NEW SCREW WAS USED TO FIX THE PLATE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20306099 ↗MalfunctionApr 8, 2024Material Fragmentation; Unraveled Material; Material Fragmentation; Unraveled Material
AS REPORTED: "ORIF WAS PERFORMED FOR LEFT DISTAL FIBULA FRACTURE (PRIMARY SURGERY DATE IS UNKNOWN). DURING THE BARIATRIC FIBULA EXPLANTING SURGERY ON (B)(6) 2024, SOMETHING LIKE METAL SHAVINGS CAME OUT WHILE REMOVING THE BONE SURROUNDING THE SCREW HEAD." NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19068007 ↗MalfunctionFeb 20, 2024Positioning Problem; Positioning Problem
AS REPORTED: "LOCKING SCREW DID NOT LOCK WITH PLATE, OPENED AND USED A NEW DIFFERENT PRODUCT AND SUCCESSFULLY LOCKED."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18743742 ↗MalfunctionFeb 6, 2024Positioning Problem; Positioning Problem
AS REPORTED: "PLATE FIXATION USING VARIAX FIBULA FOR RIGHT FIBULA FRACTURE. DURING THE SURGERY, THE SCREW LOCKING FAILED. THE SCREW WAS REPLACED WITH ANOTHER ONE (DIFFERENT SIZE), AND THE NEW SCREW COULD BE USED WITHOUT PROBLEM."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18655476 ↗MalfunctionJan 2, 2024Material Fragmentation; Material Fragmentation
THE CUSTOMER REPORTED: "DURING THE INSTALLATION OF A LOCKED SCREW ON A PLATE, A STEEL WIRE CAME LOOSE. THE WIRE PRICKED THE SURGEON'S FINGER."
Patient problems: Needle Stick/Puncture
MAUDE report 18424445 ↗InjuryNov 28, 2023Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT ON (B)(6) 2021 THE PATIENT HAD PRIMARY SURGERY AND A FIBULAR PLATE AND SCREWS WERE INSERTED. THE PLATE AND SCREWS WERE EXTRACTED ON (B)(6) 2022 DUE TO COMPLETED BONE UNION. ON (B)(6) 2023 THE PATIENT WAS SEEN IN THE OUTPATIENT CLINIC BECAUSE OF THE APPEARANCE OF AN INFLAMMATORY REACTION. THE AFFECTED AREA HAD STARTED TO SWELL A LITTLE AROUND JANUARY AND THEN THE BULGE IN THE AFFECTED AREA GREW LARGER. A METAL SHARD-LIKE SUBSTANCE WAS REMOVED FROM THE SURGICAL WOUND THROUGH A SMALL INCISION.
Patient problems: Failure of Implant; Foreign Body In Patient; Inflammation; Swelling/ Edema
MAUDE report 18219260 ↗InjuryNov 28, 2023Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT ON (B)(6) 2021 THE PATIENT HAD PRIMARY SURGERY AND A FIBULAR PLATE AND SCREWS WERE INSERTED. THE PLATE AND SCREWS WERE EXTRACTED ON (B)(6) 2022 DUE TO COMPLETED BONE UNION. ON (B)(6) 2023 THE PATIENT WAS SEEN IN THE OUTPATIENT CLINIC BECAUSE OF THE APPEARANCE OF AN INFLAMMATORY REACTION. THE AFFECTED AREA HAD STARTED TO SWELL A LITTLE AROUND JANUARY AND THEN THE BULGE IN THE AFFECTED AREA GREW LARGER. A METAL SHARD-LIKE SUBSTANCE WAS REMOVED FROM THE SURGICAL WOUND THROUGH A SMALL INCISION.
Patient problems: Failure of Implant; Foreign Body In Patient; Inflammation; Swelling/ Edema
MAUDE report 18219248 ↗InjuryNov 28, 2023Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT ON 28 APR 2021 THE PATIENT HAD PRIMARY SURGERY AND A FIBULAR PLATE AND SCREWS WERE INSERTED. THE PLATE AND SCREWS WERE EXTRACTED ON 01 JUN 2022 DUE TO COMPLETED BONE UNION. ON (B)(6) 2023 THE PATIENT WAS SEEN IN THE OUTPATIENT CLINIC BECAUSE OF THE APPEARANCE OF AN INFLAMMATORY REACTION. THE AFFECTED AREA HAD STARTED TO SWELL A LITTLE AROUND JANUARY AND THEN THE BULGE IN THE AFFECTED AREA GREW LARGER. A METAL SHARD-LIKE SUBSTANCE WAS REMOVED FROM THE SURGICAL WOUND THROUGH A SMALL INCISION.
Patient problems: Failure of Implant; Foreign Body In Patient; Inflammation; Swelling/ Edema
MAUDE report 18219247 ↗MalfunctionAug 17, 2023Positioning Failure; Material Deformation; Positioning Failure; Material Deformation
AS REPORTED: "WHEN THE SCREW WAS INSERTED, THE DOCTOR FELT LIKE THE SCREW HEAD GOT STRIPPED. THUS, THE DOCTOR STOPPED INSERTING THE SCREW, AND THE PROCEDURE WAS CONTINUED TO REPLACE THE SCREW WITH NEW ONE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17564603 ↗MalfunctionJun 12, 2023Contamination /Decontamination Problem; Mechanical Jam; Contamination /Decontamination Problem; Mechanical Jam
AS REPORTED: "WHEN REMOVING THE SCREW FROM THE CASE WITH A FORCEPS, THE SCREW WAS STUCK TO THE CASE AND COULD NOT BE REMOVED EASILY. THE SCREW POPPED OUT IN REACTION TO BEING TAKEN OUT AND BECAME UNCLEAN." AN ADDITIONAL SCREW WAS USED TO COMPLETE THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17105569 ↗MalfunctionApr 3, 2023Material Fragmentation; Appropriate Device Problem Term/Code Not Available; Material Fragmentation; Appropriate Device Problem Term/Code Not Available
AS REPORTED, "ORIF WAS PERFORMED FOR ANKLE FRACTURE USING VARIAX 2 FIBULA PLATE AND ASNIS 4.0MM. X-RAY TAKEN AFTER THE PLATED WAS PLACED TO THE TARGET SITE REVEALED THAT THERE WAS WHAT APPEARED TO BE A PIECE OF METAL AT THE SITE. THE METAL PIECE WAS REMOVED FROM THE PATIENT AND THE PROCEDURE WAS COMPLETED."
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; No Clinical Signs, Symptoms or Conditions
MAUDE report 16667363 ↗MalfunctionSep 30, 2022Break; Positioning Failure; Fracture; Break; Positioning Failure; Fracture
IT WAS REPORTED THAT A SCREW DID NOT LOCK TO PLATE. WHEN IT WAS REMOVED, THE HEAD BROKE OFF.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15518707 ↗MalfunctionJul 22, 2022Positioning Failure; Positioning Failure
AS REPORTED: "THE SCREW DID NOT LOCK TO THE PLATE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15085510 ↗InjuryApr 29, 2022Device Dislodged or Dislocated; Migration; Device Dislodged or Dislocated; Migration
IT WAS REPORTED THAT THERE WAS SCREW LOOSENING AND POOR FIXATION WAS DISCOVERED. THEREFORE, ALL THE IMPLANTS WERE EXTRACTED.
Patient problems: Insufficient Information
MAUDE report 14251461 ↗InjuryApr 29, 2022Device Dislodged or Dislocated; Migration; Device Dislodged or Dislocated; Migration
IT WAS REPORTED THAT THERE WAS SCREW LOOSENING AND POOR FIXATION WAS DISCOVERED. THEREFORE, ALL THE IMPLANTS WERE EXTRACTED.
Patient problems: Insufficient Information
MAUDE report 14251460 ↗InjuryApr 29, 2022Device Dislodged or Dislocated; Migration; Device Dislodged or Dislocated; Migration
IT WAS REPORTED THAT THERE WAS SCREW LOOSENING AND POOR FIXATION WAS DISCOVERED. THEREFORE, ALL THE IMPLANTS WERE EXTRACTED.
Patient problems: Insufficient Information
MAUDE report 14251330 ↗MalfunctionDec 20, 2021Positioning Failure; Positioning Failure
AS REPORTED: "THE SUBJECT SCREW DID NOT LOCK IN THE MOST DISTAL HOLE. ATTEMPTS WERE MADE IN OTHER HOLES, BUT NO LOCKING OCCURRED."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13049454 ↗MalfunctionJul 12, 2021Positioning Failure; Positioning Failure
IT WAS REPORTED THE SCREW HAD A LOCKING FAILURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12150498 ↗
Other PLATE, FIXATION, BONE devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-PLF40Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-PRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page