VARIAX 657318SRecalls & reported problems

Stryker GmbH · Locking Screw · Class 2 · Implant

All models of VARIAX (345)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 07613327086904), received between Nov 19, 2019 and Jul 1, 2025 — classified as 2 injuries, 8 malfunctions.

The most frequently reported specific device problem is “Positioning Failure” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Positioning Failure6
  • Material Fragmentation3
  • Migration2
  • Device Dislodged or Dislocated1
  • Difficult to Insert1
  • Material Erosion1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Insufficient Information2
  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 1, 2025Migration; Migration

    AS REPORTED: "PRIMARY SURGERY WAS PERFORMED ON (B)(6) 2025. TWO MONTHS AFTER SURGERY, AN X-RAY REVEALED THAT THE 2 LOCKING SCREWS INSERTED ON THE OUTSIDE PLATE HAD BACKED OUT OF THE PLATE AND THE LOCK HAD COME LOOSE. THE SCREWS INSERTED INTO THE FIRST AND THIRD HOLES FROM THE OUTER SIDE OF THE PLATE BACKED OUT. IT IS UNCLEAR WHICH OF THE SCREWS ACTUALLY BACKED OUT. THE PATIENT IS ORIGINALLY A FARMER AND, ACCORDING TO REPORTS, RESUMED WORK THREE WEEKS AFTER SURGERY WITHOUT FOLLOWING THE DOCTOR'S INSTRUCTIONS. REVISION SURGERY IS SCHEDULED ON (B)(6) 2025".

    Patient problems: Insufficient Information

    MAUDE report 22378067 ↗
  • MalfunctionJul 9, 2024Positioning Failure; Positioning Failure

    AS REPORTED: "DID NOT LOCK INTO PLATE. IT WAS REPLACED BY ANOTHER SCREW."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19699850 ↗
  • MalfunctionApr 26, 2024Positioning Failure; Positioning Failure

    AS REPORTED: "ORIF WAS PERFORMED FOR A MALE PATIENT WITH FIBULA FRACTURE. DURING SURGERY, TWO LOCKING SCREW HAD TO BE REPLACED DUE TO LOCKING FAILURE DURING INSERTION. FINALLY IT WAS SUCCESSFULLY LOCKED BY USING THE THIRD LOCKING SCREW AND THE SURGERY WAS COMPLETED WITHOUT PROBLEM. NO ADVERSE EVENTS OCCURRED."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19197860 ↗
  • MalfunctionMar 2, 2023Positioning Failure; Material Fragmentation; Positioning Failure; Material Fragmentation

    AS REPORTED: "WIRE-LIKE SHAVINGS WERE PRODUCED DURING INSERTION OF THE SCREW. SINCE THE SCREW DID NOT LOCK, A NEW SCREW WAS REPLACED AND THE OPERATION WAS TERMINATED."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16474005 ↗
  • InjuryApr 29, 2022Device Dislodged or Dislocated; Migration; Device Dislodged or Dislocated; Migration

    IT WAS REPORTED THAT THERE WAS SCREW LOOSENING AND POOR FIXATION WAS DISCOVERED. THEREFORE, ALL THE IMPLANTS WERE EXTRACTED.

    Patient problems: Insufficient Information

    MAUDE report 14251331 ↗
  • MalfunctionApr 8, 2022Positioning Failure; Material Fragmentation; Positioning Failure; Material Fragmentation

    AS REPORTED: "THE LOCKING SCREW CONTINUED TO TURN WITHOUT LOCKING. ONCE THE SCREW WAS REMOVED, A DEBRIS WAS FOUND. OTHER SCREW WAS USED AFTER THAT AND THE SURGERY WAS FINISHED WITHOUT ANY OTHER PROBLEMS."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14060080 ↗
  • MalfunctionApr 16, 2021Positioning Failure; Positioning Failure

    AS REPORTED: "LOCKING SCREW DID NOT LOCK DURING THE OPERATION FOR CLAVICLE(RIGHT SIDE) . THE SURGEON'S VIEW FOR THE CAUSE WAS UNKNOWN. IT MAY BE TOO ANGLED, BUT THE REPORTING SR THOUGHT IT WAS WITHIN THE ACCEPTABLE RANGE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11678438 ↗
  • MalfunctionSep 28, 2020Positioning Failure; Positioning Failure

    AS REPORTED: "[EVENT DETAILS] WHEN THE CLAVICLE HOOK PLATE WAS OPERATED, THE LOCKING SCREW DID NOT LOCK IN THE LOCKING HOLE. PLATE USED: 628535S LOCATION OF DEFECTIVE LOCK: FRONT SIDE OF THE PLATE'S MOST DISTAL HOLE OCCURRENCE STATUS: - INSERT 3 SCREWS AFTER SETTING THE PLATE ((1) ELLIPTICAL HOLE: NON-LOCKING, (2) ROUND HOLE NEXT TO THE ELLIPSE DISTAL SIDE: ROCKING, (4) FAR-MOST REAR CIRCULAR HOLE: ROCKING) - A PHENOMENON THAT THE LOCKING SCREW DOES NOT LOCK WHEN THE LOCKING SCREW IS INSERTED 18MM INTO THE 4TH MOST DISTAL LOCKING HOLE - SET IT THERE AND INSERT THE LOCKING SCREW INTO THE MOST RECENT HOLE (LOCK WITHOUT ANY PROBLEM) - I TRIED THE FOLLOWING FOR HOLES THAT DID NOT LOCK AGAIN - CHECK THE DRILLING DIRECTION ¿ NO PROBLEM (THE GUIDE SLEEVE WAS WITHIN THE SWING RANGE) - CONFIRMATION OF PLATE LIP AND SCREW HEAD ¿ NO INTERSTITIAL TISSUE SUCH AS SOFT TISSUE CAN BE CONFIRMED ON THE LIP. AS FAR AS I COULD SEE THE SCREW HEAD, IT SEEMED TO BE NO PROBLEM. - CHANGE THE DRILL DIRECTION AND INSERT ANOTHER LOCKING SCREW 16 MM AGAIN, BUT IT DOES NOT LOCK - FINALLY, AT THE DISCRETION OF DR., IT ENDS WITHOUT INSERTING A SCREW INTO THE HOLE." "THE PLATE IS CURRENTLY INSIDE THE PATIENT'S BOD

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10597565 ↗
  • MalfunctionApr 8, 2020Difficult to Insert; Difficult to Insert

    AS REPORTED, THE LOCKING SCREW DID NOT LOCK WHEN THE LOCKING SCREW WAS INSERTED INTO THE PLATE. I CHECKED THE INSERTION ANGLE AND THE DEPTH OF THE DRILL, AND REINSERTED IT SEVERAL TIMES, BUT DID NOT LOCK. I CHANGED IT TO A SCREW OF THE SAME SIZE AND LOCKED IT WHEN I PUT IT BACK IN. DOCTOR'S VIEW: "IS THE SCREW DEFECTIVE, THE PLATE DEFECTIVE, OR THE SCREW TOO TIGHT?" (B)(4) COMMENT, "IT IS POSSIBLE THAT THE SCREW HAS CONTACTED HARD BONE AND CHANGED THE ANGLE."

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9936221 ↗
  • MalfunctionNov 19, 2019Material Erosion; Material Fragmentation; Material Erosion; Material Fragmentation

    VARIAX CLAVICLE OPERATION. WHEN THE LOCKING SCREW WAS INSERTED, A THREAD-LIKE METAL PIECE CAME OUT FROM BETWEEN THE PLATE AND THE SCREW. REPLACED WITH A NEW LOCKING SCREW OF THE SAME SIZE. THE OPERATION ENDED SUCCESSFULLY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9345730 ↗

Other PLATE, FIXATION, BONE devices

All PLATE, FIXATION, BONE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page