VARIAX 657322SRecalls & reported problems

Stryker GmbH · Locking Screw · Class 2 · Implant

All models of VARIAX (345)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 07613327086966), received between Feb 10, 2020 and Dec 22, 2023 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Material Fragmentation” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 3 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation3
  • Positioning Failure3
  • Difficult to Insert1
  • Migration1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Nonunion/Delayed-union Bone Fracture1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 22, 2023Material Fragmentation; Material Fragmentation

    AS REPORTED: "THREADED PIECES OF METAL APPEARED WHEN TIGHTENING A SCREW TO THE PLATE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18387221 ↗
  • MalfunctionDec 22, 2023Material Fragmentation; Material Fragmentation

    AS REPORTED: "THREADED PIECES OF METAL APPEARED WHEN TIGHTENING A SCREW TO THE PLATE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18387218 ↗
  • MalfunctionDec 22, 2023Material Fragmentation; Material Fragmentation

    AS REPORTED: "THREADED PIECES OF METAL APPEARED WHEN TIGHTENING A SCREW TO THE PLATE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18387216 ↗
  • MalfunctionNov 9, 2021Positioning Failure; Positioning Failure

    IT WAS REPORTED THERE WERE THREE SCREWS THAT DID NOT LOCK, AND EACH TIME THE SURGEON HAD THEM REPLACED WITH SCREWS OF THE SAME SIZE OR 2 MM SHORTER. THE REPLACED SCREWS LOCKED WITHOUT ANY PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12777517 ↗
  • MalfunctionOct 5, 2021Positioning Failure; Positioning Failure

    IT WAS REPORTED THAT AT THE TIME OF VARIAX ELBOW OPERATION, THE LOCKING SCREW WAS INSERTED, BUT IT WAS NOT LOCKED. IT WAS DIFFICULT TO REMOVE AND SPENT EXTRA OPERATING TIME. THE SURGEON WAS ABLE TO REMOVE IT AND REPLACE IT WITH A LOCKING SCREW OF A DIFFERENT LENGTH.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12577127 ↗
  • MalfunctionJan 5, 2021Positioning Failure; Difficult to Insert; Positioning Failure; Difficult to Insert

    AS REPORTED: "THE LOCKING SCREW COULD NOT BE COMPLETELY FASTENED. THE DOCTOR TRIED TO TURN IT WITH A SCREWDRIVER MANY TIMES, BUT I COULDN'T TURN THE SCREW ON THE WAY. DOCTOR'S VIEW: THE PATIENT'S BONES WERE STIFF AND THE SCREW COULD NOT ADVANCE. THE TIP OF THE SCREWDRIVER HAS COME OFF THE SCREW HEAD MANY TIMES. IT WOULD BE NICE TO HAVE A SCREW THAT CAN BE INSERTED EVEN IN PATIENTS WITH HARD BONES. I WANT THE TIP OF THE SCREWDRIVER TO BE SHAPED SO THAT IT DOES NOT COME OFF THE SCREW HEAD."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11121220 ↗
  • MalfunctionFeb 10, 2020Migration; Migration

    AS REPORTED: "IN 2006 MAYO PLATE WAS USED FOR NONUNION FRACTURE OF THE DISTAL HUMERUS. ON (B)(6) 2019, MAYO PLATE WAS REMOVED, AND MEDIAL AND LATERAL VARIAX ELBOW PLATE WAS IMPLANTED. TWO WEEKS AFTER REVISION SURGERY, BACK-OUT HAPPENED. (1 PROXIMAL LOCKING SCREW, 2 PROXIMAL CORTICAL SCREW, MOST DISTAL LOCKING SCREW OF LATERAL PLATE)". NO ADDITIONAL SURGERY IS PLANNED.

    Patient problems: Nonunion/Delayed-union Bone Fracture

    MAUDE report 9690090 ↗

Other PLATE, FIXATION, BONE devices

All PLATE, FIXATION, BONE devices →