Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07613327087017), received between Aug 5, 2020 and Jun 9, 2021 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Positioning Failure” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Positioning Failure3
- Material Fragmentation2
- Device Dislodged or Dislocated1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Failure of Implant1
- Foreign Body In Patient1
- Injury1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 9, 2021Positioning Failure; Positioning Failure
AS REPORTED: "WHEN LOCKING A TALUS-NAVICULAR PLATE FROM OUR VARIAX TRAY WITH A 3.5MM Ø AND 26MM LONG VARIAX2 LOCKING SCREW, THE FOLLOWING ISSUE OCCURRED: THE SCREW WAS SCREWED IN TWICE AND LOCKED IN THE PLATE - THE SECOND TIME IT JUST SPUN AND THE SURGEON COULD NOT LOCK IT."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11971181 ↗MalfunctionSep 29, 2020Positioning Failure; Material Fragmentation; Positioning Failure; Material Fragmentation
FOOT INJURY. EXTENDED SURGERY TIME. CHIP FORMATION. SCREW CHANGE 4 X. LENGTH OF SURGICAL DELAY IS NOT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10603164 ↗MalfunctionSep 25, 2020Positioning Failure; Material Fragmentation; Positioning Failure; Material Fragmentation
AS REPORTED: "ANGULAR STABILITY DID NOT WORK. SCREWS HAVE OVERRUN AND CHIPS HAVE COME LOOSE UNDER THE SCREW HEAD (UNFORTUNATELY, CHIPS COULD NOT BE SECURED). REPLACEMENT WAS AVAILABLE. PATIENT INVOLVEMENT WAS REPORTED BUT THE REPORTER IS NOT AWARE OF ANY IMPACT."
Patient problems: Foreign Body In Patient
MAUDE report 10586511 ↗InjuryAug 5, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated
THE CUSTOMER REPORTED THAT : INSTALLATION OF THE PLATE WITH 9 SCREWS (4 AT THE TOP AND 5 AT THE BOTTOM) IN THE PATIENT ON (B)(6) 2020 AFTER REMOVAL OF OLD EQUIPMENT FOR SEPTIC PSEUDOARTHROSIS (INITIAL MATERIAL IMPLANTED IN (B)(6) 2019 FOR FRACTURE LEFT HUMERUS) AND SUPPLY OF ILIAC GRAFT. IN POST-OPERATIVE RESIN IMMOBILIZATION ANTEBRACHIAL-PALMAR AND ANTIBIOTIC THERAPY PROBABILISTIC IV THEN RELAY PER OS ON ANTIBIOGRAM. CONTROL RADIOGRAPHY ON (B)(6) 2020 SHOWING A DISPLACEMENT OF THE SCREWS AND THE PLATE WITH PASSAGE OF 2 SCREWS THROUGH THE PLATE (UPPER PART) AND 1 SCREW OUT OF THE PLATE (UPPER PART). DECISION TO REMOVE THE MATERIAL IN THE OPERATING ROOM OPERATION ON (B)(6) 2020 AND THE INSTALLATION OF 2 NEW PLATES AND RESIN IMMOBILIZATION ANTEBRACHIAL-PALMAR. CONTINUED ANTIBIOTIC THERAPY. NEXT CHECK-UP SCHEDULED FOR (B)(6) 2020. SURGICAL REVISION WITH REMOVAL OF MATERIAL AND RETOOLING.
Patient problems: Failure of Implant; Injury
MAUDE report 10369308 ↗
Other Screw, fixation, bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-350Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-400Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-200Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page