Skip to content

AXSOS 627204SRecalls & reported problems

Stryker GmbH · Proximal Lateral Humerus Plate, LEFT · Class 2 · Implant

All models of AXSOS (184)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07613327091106), received between Jul 29, 2020 and Jul 29, 2020 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Fragmentation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Material Fragmentation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 29, 2020Material Fragmentation; Material Fragmentation

    AS REPORTED: "EVENT DETAILS: WHEN FIXING THE SCREW TO THE PLATE, IT WAS CONFIRMED THAT THERE WAS METAL LINT. PHYSICIAN'S POINT OF VIEW: THE SCREWS WERE INSERTED AT A DIFFICULT ANGLE, SO THE METALS RUBBED AGAINST EACH OTHER AND WERE SCRAPED."

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10339750 ↗

Other Plate, fixation, bone devices

All Plate, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page