AXSOS 705015Recalls & reported problems
Stryker GmbH · Screwdriver Bit T15, AO Fitting · Class 1
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 07613327091649), received between May 2, 2018 and Sep 21, 2026 — classified as 1 injury, 7 malfunctions.
The most frequently reported specific device problem is “Fracture” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture8
- Break5
- Detachment Of Device Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Foreign Body In Patient2
- Insufficient Information1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 21, 2026Fracture
The primary surgery was performed on (b)(6) 2024.The tip of the screwdriver broke while removing a locking screw. The plate and screw head were broken up using a carbide drill, and the remaining screw was extracted using the hospital¿s broken-screw extractor.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50840847 ↗MalfunctionMar 17, 2025Break; Fracture; Break; Fracture
AS REPORTED: "TORX15 SCREWDRIVER OR SCREWDRIVER BIT TWISTED OFF OR BROKEN OFF WHEN INSERTING THE SCREWS, A DIFFERENT SCREWDRIVER WAS THEN USED. BOTH THE BIT AND THE SCREWDRIVER ARE 3 YEARS OLD."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21621802 ↗MalfunctionNov 28, 2024Break; Fracture; Break; Fracture
MalfunctionOct 15, 2024Break; Fracture; Break; Fracture
IT WAS REPORTED THAT DURING A PROCEDURE AN ISSUE WAS OBSERVED WITH A PANGEA T15 SHORT SCREWDRIVER SHAFT WHILE USING A TORQUE LIMITER TO PLACE LOCKING SCREWS IN A PROXIMAL HUMERUS PLATE. AFTER SUCCESSFULLY PLACING 3-5 SCREWS, THE TIP OF THE SCREWDRIVER SHAFT TWISTED, LEAVING SMALL METAL FRAGMENTS IN THE SCREW HEAD. DESPITE A BRIEF 3-5 MINUTE SURGICAL DELAY TO ATTEMPT REMOVAL OF THE FRAGMENTS USING A CURETTE AND BONE HOOK, SOME DEBRIS REMAINED IN THE PATIENT. THE SURGEON EXPRESSED CONCERN THAT A PIECE OF THE DRIVER HEAD MAY HAVE COLD-WELDED TO THE SCREW, POTENTIALLY COMPLICATING FUTURE REMOVAL OF THE PLATE. LONG-TERM COMPLICATIONS MAY ARISE IF THE SCREW HEAD IS DEFORMED, WHICH WOULD REQUIRE MORE SPECIALIZED EQUIPMENT, SUCH AS A CARBIDE DRILL BIT OR MIDAS REX, FOR REMOVAL.
Patient problems: Foreign Body In Patient
MAUDE report 20456223 ↗InjuryJul 24, 2024Break; Fracture; Break; Fracture
A PLATE AND A SCREW WERE COMPLETELY LOCKED AND THE SCREWDRIVER WAS BROKEN DURING REMOVAL PROCEDURE. THUS, THE SURGEON REMOVED THE SCREW HEAD USING A CARBIDE DRILL, AND THE TIP OF THE SCREW WAS REMOVED AFTER THE PLATE WAS REMOVED TO COMPLETE THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19820419 ↗MalfunctionMar 26, 2021Fracture; Fracture
AS REPORTED: "SCREW BREAKAGE IN AXSOS 3 DISTAL MEDIAL PLATE DESCRIBED BY CUSTOMER AS COLD WELDING. PROXIMAL LOCKING SCREW MEASURED TOO LONG, WHEN GOING TO CHANGE THE SCREW OUT THE SCREW HEAD BROKE. THE LOCKING TOWER WAS USED CORRECTLY, THE TORQUE WRENCH WAS USED TO TIGHTEN THE SCREW INITIALLY. INITIAL REMOVAL WAS ATTEMPTED ON HAND SCREWDRIVER. A BROKEN SCREW REMOVAL SET WAS USED TO REMOVE THE BROKEN SCREW FROM THE LATERAL SIDE, TO DO THIS A BURR WAS USED TO REMOVE A PORTION OF A FIBULA PLATE WHICH HAD BEEN PLACED IN THE SAME OPERATION. 2 SCREW DRIVER SHAFTS WERE BROKEN IN THE PROCEDURE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11577487 ↗MalfunctionMar 26, 2021Fracture; Fracture
AS REPORTED: "SCREW BREAKAGE IN AXSOS 3 DISTAL MEDIAL PLATE DESCRIBED BY CUSTOMER AS COLD WELDING. PROXIMAL LOCKING SCREW MEASURED TOO LONG, WHEN GOING TO CHANGE THE SCREW OUT THE SCREW HEAD BROKE. THE LOCKING TOWER WAS USED CORRECTLY, THE TORQUE WRENCH WAS USED TO TIGHTEN THE SCREW INITIALLY. INITIAL REMOVAL WAS ATTEMPTED ON HAND SCREWDRIVER. A BROKEN SCREW REMOVAL SET WAS USED TO REMOVE THE BROKEN SCREW FROM THE LATERAL SIDE, TO DO THIS A BURR WAS USED TO REMOVE A PORTION OF A FIBULA PLATE WHICH HAD BEEN PLACED IN THE SAME OPERATION. 2 SCREW DRIVER SHAFTS WERE BROKEN IN THE PROCEDURE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11577436 ↗MalfunctionMay 2, 2018Fracture; Break; Detachment Of Device Component; Fracture
IT WAS REPORTED THAT DURING AN AXSOS PROXIMAL HUMERUS CASE, THE SCREWDRIVER BIT WHICH IS ATTACHED TO THE TORQUE LIMITING ADAPTOR WAS USED TO INSERT THE SCREW BY HAND. PRIOR TO ACHIEVING 2.5NM TORQUE, THE TIP OF THE SCREWDRIVER BROKE OFF. IT WAS REPORTED THAT THE BROKEN TIP REMAINED INSIDE THE SCREW-HEAD. IT WAS REPORTED THAT THE SCREW WAS LEFT IN THE PATIENT.
Patient problems: Foreign Body In Patient
MAUDE report 7481055 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page