NA 5407-120-900Recalls & reported problems
Stryker Corporation · MIS Hub · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Stryker MIS Hub is a powered surgical drill, burr and accessory hub. Most of the 64 reports describe the handpiece or burr overheating during testing or non-clinical procedures, with parts flaking off, cracking, fracturing or detaching, plus a few fluid leaks and minor burns. No deaths and no injuries were reported, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 64
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 64 adverse event reports that list this exact device (DI 07613327096439), received between Jul 24, 2018 and Apr 22, 2026 — classified as 64 malfunctions.
The most frequently reported specific device problem is “Overheating of Device” (38 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Overheating of Device38
- Flaked10
- Fracture9
- Break5
- Detachment of Device or Device Component5
- Fluid/Blood Leak2
- Material Disintegration1
Patient problems
- No Clinical Signs, Symptoms or Conditions43
- No Patient Involvement20
- No Consequences Or Impact To Patient7
- Insufficient Information5
- Superficial (First Degree) Burn4
- Burn(s)3
- No Known Impact Or Consequence To Patient3
- Burn, Thermal1
Recent adverse event reports
The 25 most recent of 64 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 22, 2026Detachment of Device or Device Component
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN (B)(6) 2026. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: DETACHMENT OF DEVICE OR DEVICE COMPONENT. 1 EVENT HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24954392 ↗MalfunctionJan 23, 2026Overheating of Device
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1 ¿ DECEMBER 31 2025. THIS REPORT SUMMARIZES 29 EVENTS FOR THE FAILURE: OVERHEATING OF DEVICE. - 22 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. - 7 EVENTS HAD PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24163084 ↗MalfunctionOct 28, 2025Overheating of Device; Detachment of Device or Device Component; Overheating of Device; Detachment of Device or Device Component
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURES: OVERHEATING OF DEVICE; DETACHMENT OF DEVICE OR DEVICE COMPONENT. 1 EVENT HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23399833 ↗MalfunctionOct 23, 2025Overheating of Device; Overheating of Device
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 11 EVENTS FOR THE FAILURE: OVERHEATING OF DEVICE - 2 EVENTS HAD PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT - 9 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23368938 ↗MalfunctionOct 23, 2025Fracture; Fracture
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: FRACTURE. - 1 EVENT HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23368915 ↗MalfunctionJul 24, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 3 EVENTS FOR THE FAILURE: DETACHMENT OF DEVICE OR DEVICE COMPONENT. - 3 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22587050 ↗MalfunctionJul 24, 2025Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 26 EVENTS FOR THE FAILURE: OVERHEATING OF DEVICE. 23 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. 3 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT.
Patient problems: Burn(s); No Clinical Signs, Symptoms or Conditions
MAUDE report 22586871 ↗MalfunctionApr 21, 2025Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 37 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. - 29 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. - 1 EVENT HAD THE PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT. - 7 EVENTS HAD PATIENT INVOLVEMENT: NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21876616 ↗MalfunctionJan 22, 2025Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 34 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. - 12 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. - 22 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21207340 ↗MalfunctionJan 21, 2025Flaked; Flaked
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD PAINT CHIPS MISSING. - 1 EVENT HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE. -THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21201038 ↗MalfunctionOct 21, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD A COMPONENT DETACH. - (B)(4) EVENT HAD THE PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20501114 ↗MalfunctionOct 21, 2024Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 24 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. 20 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. 4 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20495780 ↗MalfunctionJul 24, 2024Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 24 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. - 14 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. - 9 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT. - 1 EVENT HAD INSUFFICIENT INFORMATION RECEIVED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 19823100 ↗MalfunctionJul 24, 2024Fracture; Fracture
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. 1 EVENT HAD INSUFFICIENT INFORMATION RECEIVED.
Patient problems: Insufficient Information
MAUDE report 19817125 ↗MalfunctionApr 16, 2024Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 33 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. - 20 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. - 13 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19113435 ↗MalfunctionJan 22, 2024Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 28 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. - 21 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. - 7 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18557703 ↗MalfunctionOct 23, 2023Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 33 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. 25 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. 8 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17993110 ↗MalfunctionOct 23, 2023Fracture; Overheating of Device; Fracture; Overheating of Device
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE REPORTEDLY OVERHEATED AND CUTTING ACCESSORY FRACTURED. - 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17990695 ↗MalfunctionJul 26, 2023Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 29 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. - 22 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. - 7 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17401612 ↗MalfunctionApr 25, 2023Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 28 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. - 6 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. - 19 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT. - 3 EVENTS WHICH HAD A MILD INJURY, ILLNESS OR IMPAIRMENT WHICH CAN BE TREATED WITH MINIMAL OR NO INTERVENTION.
Patient problems: Burn(s); No Clinical Signs, Symptoms or Conditions
MAUDE report 16808017 ↗MalfunctionApr 25, 2023Fracture; Overheating of Device; Fracture; Overheating of Device
THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED AND CUTTING ACCESSORY FRACTURED. 3 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16807568 ↗MalfunctionApr 25, 2023Fracture; Fracture
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16806616 ↗MalfunctionJan 29, 2023Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 26 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. 14 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. 12 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16262042 ↗MalfunctionJan 29, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD A COMPONENT DETACH. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261930 ↗MalfunctionOct 25, 2022Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 29 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. 24 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. 5 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15666561 ↗
Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices
- NA EMAX2PLUSSynthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)2,623 reports
- MIDAS REX MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)952 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)416 reports
- MIDAS REX MR8 MR8-AS07Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)232 reports
- MIDAS REX MR8 MR8-AA10Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)192 reports
- Midas Rex MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)189 reports
- NA 5407-FA2-000Stryker Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)181 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)165 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page