This Stryker MIS Curved Nose Tube is a small attachment for powered surgical shaver tools. Its 85 reports mostly describe the tube overheating, flaking material, or cracking, usually noticed outside the body with no patient involved; there are no deaths and no injuries. No recalls name this device. Recent reports fell to 2, down from 10 the prior year.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 85
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 85 adverse event reports that list this exact device (DI 07613327096521), received between Jul 24, 2018 and Jan 23, 2026 — classified as 85 malfunctions.
The most frequently reported specific device problem is “Overheating of Device” (35 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Overheating of Device35
- Flaked34
- Fracture12
- Break4
- Fluid/Blood Leak1
- Material Disintegration1
- Vibration1
Patient problems
- No Clinical Signs, Symptoms or Conditions59
- No Patient Involvement24
- No Consequences Or Impact To Patient7
- No Known Impact Or Consequence To Patient4
- Superficial (First Degree) Burn1
Recent adverse event reports
The 25 most recent of 85 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 23, 2026Flaked
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1 ¿ DECEMBER 31 2025. THIS REPORT SUMMARIZES 2 EVENTS FOR THE FAILURE: FLAKED. - 2 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24164262 ↗MalfunctionJan 23, 2026Overheating of Device
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1 ¿ DECEMBER 31 2025. THIS REPORT SUMMARIZES 10 EVENTS FOR THE FAILURE: OVERHEATING OF DEVICE. 7 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. 3 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24163088 ↗MalfunctionOct 23, 2025Overheating of Device; Overheating of Device
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN (B)(6) - (B)(6) 2025. THIS REPORT SUMMARIZES 19 EVENTS FOR THE FAILURE: OVERHEATING OF DEVICE. 19 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23369136 ↗MalfunctionOct 23, 2025Flaked; Flaked
MalfunctionOct 23, 2025Fracture; Fracture
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: FRACTURE. - 1 EVENT HAD NO HEALTH CONSEQUENCES OR IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23369119 ↗MalfunctionJul 24, 2025Fracture; Fracture
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN APRIL 1 ¿ JUNE 30 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: FRACTURE. - 1 EVENT HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22587175 ↗MalfunctionJul 24, 2025Flaked; Flaked
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN APRIL 1 ¿ JUNE 30 2025. THIS REPORT SUMMARIZES 5 EVENTS FOR THE FAILURE: FLAKED. - 5 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22586858 ↗MalfunctionJul 24, 2025Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 8 EVENTS FOR THE FAILURE: OVERHEATING OF DEVICE. 6 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22586849 ↗MalfunctionApr 21, 2025Flaked; Flaked
THIS REPORT SUMMARIZES 5 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD PAINT CHIPS MISSING. 5 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21876817 ↗MalfunctionApr 21, 2025Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 28 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. 28 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21876624 ↗MalfunctionJan 22, 2025Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 21 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. 20 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. 1 EVENT HAD PATIENT INVOLVEMENT: NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21207513 ↗MalfunctionJan 21, 2025Flaked; Flaked
THIS REPORT SUMMARIZES 16 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD PAINT CHIPS MISSING. - 15 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. - 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21201017 ↗MalfunctionOct 21, 2024Flaked; Flaked
THIS REPORT SUMMARIZES 6 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD PAINT CHIPS MISSING. 6 EVENTS HAD NO PATIENT INVOLVEMENT: NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20500900 ↗MalfunctionOct 21, 2024Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 26 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. 24 EVENTS HAD NO PATIENT INVOLVEMENT: NO PATIENT IMPACT. 2 EVENTS HAD PATIENT INVOLVEMENT: NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20497826 ↗MalfunctionJul 24, 2024Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 12 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. - 12 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19823104 ↗MalfunctionJul 24, 2024Flaked; Flaked
THIS REPORT SUMMARIZES 9 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD PAINT CHIPS MISSING. - 9 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19822819 ↗MalfunctionApr 16, 2024Flaked; Flaked
THIS REPORT SUMMARIZES 14 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD PAINT CHIPS MISSING. 14 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19113509 ↗MalfunctionApr 16, 2024Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 32 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. 30 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. 2 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19113508 ↗MalfunctionJan 22, 2024Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 17 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. - 16 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. - 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18557718 ↗MalfunctionJan 22, 2024Fracture; Fracture
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. 2 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18557325 ↗MalfunctionJan 16, 2024Flaked; Flaked
THIS REPORT SUMMARIZES 10 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD PAINT CHIPS MISSING. 10 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18517874 ↗MalfunctionOct 23, 2023Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 37 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. 35 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. 2 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17993088 ↗MalfunctionOct 23, 2023Flaked; Flaked
THIS REPORT SUMMARIZES 13 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD PAINT CHIPS MISSING. 13 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17992912 ↗MalfunctionJul 26, 2023Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 12 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. - 11 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. - 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17401594 ↗MalfunctionJul 26, 2023Flaked; Flaked
THIS REPORT SUMMARIZES 7 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD PAINT CHIPS MISSING. 7 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17398268 ↗
Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices
- NA EMAX2PLUSSynthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)2,623 reports
- MIDAS REX MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)952 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)416 reports
- MIDAS REX MR8 MR8-AS07Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)232 reports
- MIDAS REX MR8 MR8-AA10Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)192 reports
- Midas Rex MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)189 reports
- NA 5407-FA2-000Stryker Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)181 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)165 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.