NA 5407-120-482Recalls & reported problems
Stryker Corporation · Elite Angled Attachment · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Stryker Elite angled attachment, used with powered surgical tools, has 132 reports, many describing overheating and flaking parts, often found during testing before use on patients. The reports mention no deaths or injuries, and no recalls name this device. Reports fell to 7 in the last 12 months from 18 in the previous 12.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 133
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 133 adverse event reports that list this exact device (DI 07613327096590), received between Jul 24, 2018 and Jul 22, 2026 — classified as 133 malfunctions.
The most frequently reported specific device problem is “Overheating of Device” (60 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Overheating of Device60
- Flaked49
- Detachment of Device or Device Component28
- Fluid/Blood Leak7
- Fracture6
- Break4
- Material Disintegration1
Patient problems
- No Clinical Signs, Symptoms or Conditions97
- No Patient Involvement33
- No Consequences Or Impact To Patient16
- No Known Impact Or Consequence To Patient11
- Insufficient Information9
- Superficial (First Degree) Burn3
- Burn(s)1
- Partial thickness (Second Degree) Burn1
Recent adverse event reports
The 25 most recent of 133 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 22, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25907821 ↗MalfunctionApr 22, 2026Flaked
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JANUARY 1 ¿ MARCH 31 2026. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: FLAKED. - 1 EVENT HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24954404 ↗MalfunctionApr 22, 2026Overheating of Device
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN (B)(6) 2026. THIS REPORT SUMMARIZES 73 EVENTS FOR THE FAILURE: OVERHEATING OF DEVICE. 3 EVENTS HAD PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT. 63 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. 7 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24953592 ↗MalfunctionJan 23, 2026Flaked; Overheating of Device
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN (B)(6) 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURES: FLAKED; OVERHEATING OF DEVICE. - 1 EVENT HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24163987 ↗MalfunctionJan 23, 2026Flaked
MalfunctionJan 23, 2026Fluid/Blood Leak
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1 ¿ DECEMBER 31 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: FLUID/BLOOD LEAK. 1 EVENT HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24163429 ↗MalfunctionJan 23, 2026Flaked; Detachment of Device or Device Component
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1 ¿ DECEMBER 31, 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURES: FLAKED; DETACHMENT OF DEVICE OR DEVICE COMPONENT. - 1 EVENT HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24163422 ↗MalfunctionJan 23, 2026Overheating of Device
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1 ¿ DECEMBER 31, 2025. THIS REPORT SUMMARIZES 111 EVENTS FOR THE FAILURE: OVERHEATING OF DEVICE. - 95 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. - 12 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT. - 4 EVENTS HAD PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24163073 ↗MalfunctionOct 28, 2025Fluid/Blood Leak; Fluid/Blood Leak
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30 2025. THIS REPORT SUMMARIZES 4 EVENTS FOR THE FAILURE: FLUID/BLOOD LEAK. 1 EVENT HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. 3 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23399860 ↗MalfunctionOct 28, 2025Overheating of Device; Overheating of Device
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30 2025. THIS REPORT SUMMARIZES 131 EVENTS FOR THE FAILURE: OVERHEATING OF DEVICE. 110 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. -21 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions; Partial thickness (Second Degree) Burn
MAUDE report 23399844 ↗MalfunctionOct 23, 2025Fracture; Fracture
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30 2025. THIS REPORT SUMMARIZES 2 EVENTS FOR THE FAILURE: FRACTURE. - 2 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23368932 ↗MalfunctionOct 23, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 4 EVENTS FOR THE FAILURE: DETACHMENT OF DEVICE OR DEVICE COMPONENT - 3 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT - 1 EVENT HAD INSUFFICIENT INFORMATION.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 23368930 ↗MalfunctionOct 23, 2025Flaked; Flaked
MalfunctionOct 23, 2025Overheating of Device; Detachment of Device or Device Component; Overheating of Device; Detachment of Device or Device Component
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURES: OVERHEATING OF DEVICE; DETACHMENT OF DEVICE OR DEVICE COMPONENT. - 1 EVENT HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23368909 ↗MalfunctionJul 29, 2025Overheating of Device; Overheating of Device
NO CHANGE.
Patient problems: Burn(s); Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 22649373 ↗MalfunctionJul 29, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 3 EVENTS FOR THE FAILURE: DETACHMENT OF DEVICE OR DEVICE COMPONENT. 2 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. 1 EVENT HAD NO HEALTH CONSEQUENCES OR IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22648116 ↗MalfunctionJul 24, 2025Flaked; Flaked
THIS REPORT SUMMARIZES 21 EVENTS FOR THE FAILURE: FLAKED. 21 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22586829 ↗MalfunctionMay 30, 2025Overheating of Device; Overheating of Device
IT WAS REPORTED THAT DURING A PROCEDURE, THE DEVICE OVERHEATED. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES, NO MEDICAL INTERVENTION AND NO DELAY AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22115413 ↗MalfunctionApr 21, 2025Flaked; Flaked
THIS REPORT SUMMARIZES 30 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD PAINT CHIPS MISSING. - 30 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21876808 ↗MalfunctionApr 21, 2025Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 132 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. 121 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. 4 EVENTS HAD THE PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT. 7 EVENTS HAD PATIENT INVOLVEMENT: NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21876672 ↗MalfunctionApr 21, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 6 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A COMPONENT DETACH. - 5 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. - 1 EVENT HAD THE PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21876272 ↗MalfunctionApr 21, 2025Overheating of Device; Detachment of Device or Device Component; Overheating of Device; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE REPORTEDLY OVERHEATED AND DISASSEMBLED. - 1 EVENT HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21875710 ↗MalfunctionApr 21, 2025Fluid/Blood Leak; Overheating of Device; Fluid/Blood Leak; Overheating of Device
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE REPORTEDLY OVERHEATED AND WAS LEAKING. 1 EVENT HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21875703 ↗MalfunctionJan 22, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 5 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A COMPONENT DETACH. - 4 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. - 1 EVENT HAD THE PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21207666 ↗MalfunctionJan 22, 2025Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 171 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. 149 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. 3 EVENTS HAD THE PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT. 19 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21207337 ↗
Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices
- NA EMAX2PLUSSynthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)2,623 reports
- MIDAS REX MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)952 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)416 reports
- MIDAS REX MR8 MR8-AS07Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)232 reports
- MIDAS REX MR8 MR8-AA10Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)192 reports
- Midas Rex MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)189 reports
- NA 5407-FA2-000Stryker Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)181 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)165 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗