ANCHORAGE 705172Recalls & reported problems
Stryker GmbH · Reamer for Cross-Plates · Class 2
All models of ANCHORAGE (52)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 07613327131055), received between Sep 17, 2024 and Sep 10, 2026 — classified as 2 injuries, 14 malfunctions.
The most frequently reported specific device problem is “Dull, Blunt” (16 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Dull, Blunt16
Patient problems
- No Clinical Signs, Symptoms or Conditions14
- Necrosis2
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 10, 2026Dull, Blunt
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50716214 ↗MalfunctionSep 7, 2026Dull, Blunt
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50652062 ↗MalfunctionApr 20, 2026Dull, Blunt
AS REPORTED: "USED BOTH REAMERS AND THEY WERE NOT REAMING. VERY DULL."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24922339 ↗MalfunctionApr 20, 2026Dull, Blunt
AS REPORTED: "USED BOTH REAMERS AND THEY WERE NOT REAMING. VERY DULL."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24922337 ↗MalfunctionDec 29, 2025Dull, Blunt; Dull, Blunt
AS REPORTED: "ANCHORAGE 2 CP REAMERS HAVE BEEN SHIPPING DULL. THEY DO NOT CUT INTO THE BONE. OUR SURGEONS HAVE BEEN USING ANCHORAGE 1 CP REAMERS FOR ANCHORAGE 2 CASES AS A BACKUP BUT THAT ISN'T IDEAL AS THEY ARE NOT THE SAME SIZE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23913033 ↗MalfunctionDec 29, 2025Dull, Blunt; Dull, Blunt
AS REPORTED: "ANCHORAGE 2 CP REAMERS HAVE BEEN SHIPPING DULL. THEY DO NOT CUT INTO THE BONE. OUR SURGEONS HAVE BEEN USING ANCHORAGE 1 CP REAMERS FOR ANCHORAGE 2 CASES AS A BACKUP¿BUT THAT ISN¿T IDEAL AS THEY ARE NOT THE SAME SIZE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23913030 ↗MalfunctionDec 29, 2025Dull, Blunt; Dull, Blunt
AS REPORTED: "ANCHORAGE 2 CP REAMERS HAVE BEEN SHIPPING DULL. THEY DO NOT CUT INTO THE BONE. OUR SURGEONS HAVE BEEN USING ANCHORAGE 1 CP REAMERS FOR ANCHORAGE 2 CASES AS A BACKUP¿BUT THAT ISN'T IDEAL AS THEY ARE NOT THE SAME SIZE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23912057 ↗MalfunctionDec 29, 2025Dull, Blunt; Dull, Blunt
AS REPORTED: "ANCHORAGE 2 CP REAMERS HAVE BEEN SHIPPING DULL. THEY DO NOT CUT INTO THE BONE. OUR SURGEONS HAVE BEEN USING ANCHORAGE 1 CP REAMERS FOR ANCHORAGE 2 CASES AS A BACKUP¿BUT THAT ISN¿T IDEAL AS THEY ARE NOT THE SAME SIZE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23912056 ↗MalfunctionDec 29, 2025Dull, Blunt; Dull, Blunt
AS REPORTED: "ANCHORAGE 2 CP REAMERS HAVE BEEN SHIPPING DULL. THEY DO NOT CUT INTO THE BONE. OUR SURGEONS HAVE BEEN USING ANCHORAGE 1 CP REAMERS FOR ANCHORAGE 2 CASES AS A BACKUP¿BUT THAT ISN¿T IDEAL AS THEY ARE NOT THE SAME SIZE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23912055 ↗MalfunctionDec 29, 2025Dull, Blunt; Dull, Blunt
AS REPORTED: "ANCHORAGE 2 CP REAMERS HAVE BEEN SHIPPING DULL. THEY DO NOT CUT INTO THE BONE. OUR SURGEONS HAVE BEEN USING ANCHORAGE 1 CP REAMERS FOR ANCHORAGE 2 CASES AS A BACKUP¿BUT THAT ISN¿T IDEAL AS THEY ARE NOT THE SAME SIZE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23912053 ↗MalfunctionDec 29, 2025Dull, Blunt; Dull, Blunt
AS REPORTED: "ANCHORAGE 2 CP REAMERS HAVE BEEN SHIPPING DULL. THEY DO NOT CUT INTO THE BONE. OUR SURGEONS HAVE BEEN USING ANCHORAGE 1 CP REAMERS FOR ANCHORAGE 2 CASES AS A BACKUP¿BUT THAT ISN¿T IDEAL AS THEY ARE NOT THE SAME SIZE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23912052 ↗MalfunctionDec 29, 2025Dull, Blunt; Dull, Blunt
AS REPORTED: "ANCHORAGE 2 CP REAMERS HAVE BEEN SHIPPING DULL. THEY DO NOT CUT INTO THE BONE. OUR SURGEONS HAVE BEEN USING ANCHORAGE 1 CP REAMERS FOR ANCHORAGE 2 CASES AS A BACKUP¿BUT THAT ISN¿T IDEAL AS THEY ARE NOT THE SAME SIZE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23912051 ↗MalfunctionDec 29, 2025Dull, Blunt; Dull, Blunt
AS REPORTED: "ANCHORAGE 2 CP REAMERS HAVE BEEN SHIPPING DULL. THEY DO NOT CUT INTO THE BONE. OUR SURGEONS HAVE BEEN USING ANCHORAGE 1 CP REAMERS FOR ANCHORAGE 2 CASES AS A BACKUP¿BUT THAT ISN¿T IDEAL AS THEY ARE NOT THE SAME SIZE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23912047 ↗MalfunctionDec 29, 2025Dull, Blunt; Dull, Blunt
AS REPORTED: "ANCHORAGE 2 CP REAMERS HAVE BEEN SHIPPING DULL. THEY DO NOT CUT INTO THE BONE. OUR SURGEONS HAVE BEEN USING ANCHORAGE 1 CP REAMERS FOR ANCHORAGE 2 CASES AS A BACKUP¿BUT THAT ISN¿T IDEAL AS THEY ARE NOT THE SAME SIZE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23912046 ↗InjurySep 17, 2024Dull, Blunt; Dull, Blunt
IT WAS REPORTED THAT THE CP REAMER FROM ANCHORAGE 2CC WAS TOO DULL TO FUNCTION, FRUSTRATING THE SURGEON AS NEITHER REAMER IN THE TRAY WORKED. THIS LOT CODE HAS KNOWN ISSUES, RAISING CONCERNS ABOUT WHY IT¿S STILL IN CIRCULATION. THE SURGEON APPLIED EXCESSIVE PRESSURE, CAUSING SMOKE AND POTENTIAL THERMAL NECROSIS, BUT PROCEEDED WITH THE PROCEDURE. THE REAMERS, PART OF A LOANER KIT, WERE DISCARDED AFTER USE.
Patient problems: Necrosis
MAUDE report 20236553 ↗InjurySep 17, 2024Dull, Blunt; Dull, Blunt
IT WAS REPORTED THAT THE CP REAMER FROM ANCHORAGE 2CC WAS TOO DULL TO FUNCTION, FRUSTRATING THE SURGEON AS NEITHER REAMER IN THE TRAY WORKED. THIS LOT CODE HAS KNOWN ISSUES, RAISING CONCERNS ABOUT WHY IT¿S STILL IN CIRCULATION. THE SURGEON APPLIED EXCESSIVE PRESSURE, CAUSING SMOKE AND POTENTIAL THERMAL NECROSIS, BUT PROCEEDED WITH THE PROCEDURE. THE REAMERS, PART OF A LOANER KIT, WERE DISCARDED AFTER USE.
Patient problems: Necrosis
MAUDE report 20236546 ↗
Other Plate, fixation, bone devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-PLF40Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-PRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page