Skip to content

INSTRUMENT 6496-9-054Recalls & reported problems

Howmedica Osteonics Corp. · GMRS TAPER SEPARATOR C CLAMP BRACKET · Class 1

All models of INSTRUMENT (131)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613327198430), received between Jan 3, 2019 and Aug 27, 2024 — classified as 2 injuries.

The most frequently reported specific device problem is “Difficult to Remove” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Remove1
  • Failure to Disconnect1
  • Insufficient Device Problem Information1

Patient problems

  • Ambulation or Postural Difficulties2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 27, 2024Difficult to Remove; Failure to Disconnect; Difficult to Remove; Failure to Disconnect

    WAS TRYING TO SEPARATE A GMRS STEM FROM A GMRS DISTAL FEMORAL GROWING PROSTHESIS. THE TAPER SEPARATOR FAILED TO RELEASE THE TAPER. SURGEON WAS UNABLE TO COMPLETE THE LENGTHENING OF THE PATIENTS FEMUR. DELAYED THE SURGERY OVER 1 HOUR WHILE SURGEON TRIED TO DISASSOCIATE THE COMPONENTS. FUTURE SURGERY WILL BE REQUIRED.

    Patient problems: Ambulation or Postural Difficulties; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20078203 ↗
  • InjuryJan 3, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT, FOR A PATIENT WHO WAS OPERATED SEVERAL TIMES BY THE SAME SURGEON, THE LAST SURGERY WAS THE SECOND FAILED ATTEMPT TO EXCHANGE THE GROWING PIECE OF HIS GMRS IMPLANT (HMRS/GMRS SPIGOT SEPARATOR). A LONGER GROWING PIECE IS NEEDED; GROWING PIECE THE PATIENT CURRENTLY HAS FAILED/LOST IT¿S ABILITY TO KEEP IT¿S FULL LENGTH. IT IS SUSPECTED THAT SEPARATION OF THE STEM/GROWING PIECE SPIGOT FAILED BECAUSE OF COLD FUSION. THE PATIENT ((B)(6) BOY) HAS 6CM DIFFERENCE IN LEG LENGTH; THE ORIGINAL GROWING COMPONENT IS SET TO ITS MAXIMUM LENGTH BUT "AS IT¿S" COLLAPSED ONCE THE PATIENT¿S PARENTS LOST THEIR TRUST IN THE DEVICE. IT IS NOT CLEAR WHY/HOW THE DEVICE LOST IT¿S INTERNAL BLOCKAGE, BUT THE SURGEON FOUND IT COLLAPSED (IN THE SHORTEST POSITION INSTEAD OF THE LAST SET ELONGATION) IN THE OPERATION WHEN THEY WANTED TO DO THE LAST ADJUSTMENT (THE LAST 5MM) BEFORE CHANGING TO THE LONGER PIECE. AFTER THEY WENT THROUGH THE ELONGATION STEPS AGAIN - 2-3 SMALL SURGERIES - THEY REACHED FULL LENGTH AGAIN. THE DEVICE SHOWED NO SIGNS OF FAILURE/INSTABILITY.

    Patient problems: Ambulation or Postural Difficulties

    MAUDE report 8214875 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page