XIA 03807028Recalls & reported problems
Stryker Corporation · TORQUE WRENCH · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07613327262292), received between Sep 25, 2017 and Jun 18, 2021 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Fracture” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Fracture4
- Break3
Patient problems
- No Known Impact Or Consequence To Patient3
- No Consequences Or Impact To Patient2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 18, 2021Fracture; Fracture
IT WAS REPORTED THAT WHILE USING THE XIA LL TORQUE WRENCH TO LOCK THE NUT, THE TIP OF THE DEVICE FRACTURED. NO ADVERSE CONSEQUENCES, SURGICAL DELAY OR MEDICAL INTERVENTION WERE REPORTED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING AN ALTERNATIVELY AVAILABLE INSTRUMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12026116 ↗MalfunctionJul 25, 2019Fracture
IT WAS REPORTED THAT WHEN THE SURGEON WENT TO LOCK THE NUT, THE TIP OF THE XIA II TITANIUM TORQUE WRENCH FRACTURED. THIS EVENT RESULTED IN A 15 MINUTE SURGICAL DELAY TO REMOVE THE BROKEN PART. THE DEVICE HAS BEEN IN USE APPROXIMATELY "7 YEARS AND USED AN ESTIMATED 700 TIMES." NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8826448 ↗MalfunctionAug 29, 2018Break; Fracture; Fracture
IT WAS REPORTED THAT THE TIP OF THE INSTRUMENT BROKE OFF DURING SURGERY. NO ADVERSE CONSEQUENCES TO THE PATIENT IS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7831430 ↗MalfunctionFeb 7, 2018Break
IT WAS REPORTED THAT THE INNER SHAFT OF THE TORQUE WRENCH BROKE DURING THE FINAL TIGHTENING OF PEDICLE SCREWS AT THE L4-L5 LEVEL. THERE WAS A REPORTED DELAY OF MORE THAN 20 MINUTES. THE SURGERY WAS SUCCESSFULLY COMPLETED WITH AN ALTERNATE TORQUE WRENCH. THERE ARE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7249952 ↗MalfunctionSep 25, 2017Break; Fracture; Fracture
IT WAS REPORTED THAT; DURING TIGHTENING OF THE BLOCKER IN THE SCREW DURING A DORSALE STABILISATION THE TIP OF THE HEX DRIVERBROKE PRIOR REACHING THE 12 NM MARK.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 6891826 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page